Operating instructions
Signum 1-2-3
Totalizing Application s
73
Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select the Setup menu: Press the
k key repeatedly until appl is
displayed
§ Confirm appl: Press the ) key
§ Select the Totalizing application:
Press the k key repeatedly and
confirm with the ) key
Totalizing Application Parameters
3. 6. Minimum load for autosave
3.6.1* 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
3.16. Autosave mode
3.16.1* Off
3.16.2 On
3.17. Individual/Component printout
when saved
3.17.1 Automatic printing off
3.17.2* Print the entire standard
print configuration
every time with the O
key
3.22. Source of data for autosave
3.22.1* Application 1
3.22.2 Application 2
3.23. Saved value
3.23.1* Net
3.23.2 Calculated
3.23.3 Net and Calculated
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times.
Minimum load
The minimum load required for
automatic taring of the container
weight on the platform (first weight)
is configured in Setup under:
Appl:
s: 3.5.
Signum
®
3:
Appl: m.weigh: 3.5.
The minimum amount that a
component must weigh before it
can be saved in totalizing memory
is configured in Setup under:
Appl:
s: 3.6.
Signum
®
3:
appl 3
:
totaliz: 3.6.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals
= 1000 digits) on the weighing platform
for autotaring (only with the “Autotare
first weight” option selected).
Protocol
In Setup, under:
Appl:
s: 3.17.
Signum
®
3:
appl 3:
totaliz: 3.17.
You can configure whether a printout
is generated manually, by pressing p,
or automatically when a weight value is
stored in the totalizing memory.
If you select the 3.17.1 menu item,
printouts can only be generated
manually by pressing the p key
(single printout).
If the 3.17.2. menu item is selected
(Print one component), the component
record is printed.
The total data record is printed when
you clear the totalizing memory
(by pressing the c key).