Product data

111
Signum 3 Ex
Data output options
Data output options
Page
6
of
5
36953-751-40
Data Transfer
2013-02-02
Dr. D. Klausgrete
Drawing number
Title
Revision 00
72583
RS485 connection with YDI05-Z
YDI05-Z
Options
A24, M50,
M55
3)
12-contact
YDI05-Z
Signal_GND
RxD-TxD-P
RxD-TxD-N
J
K
Signal_GND
Junction Box
e.g.
72583
RxD-TxD-P
RxD-TxD-N
RxD-TxD-P
RxD-TxD-N
Signal_GND
M
Connector 69Y03166
1)
connector
cable gland
14-contact
1
4
1
LV2
LV3
LV1
1
4
5
12-contact
C
B
E
screened cable
1)
(e.g. YCC485-X)
up to 1000m
LV1
6
1
1
4
LV2
CIXS3
2)
14
LV3
Mainboard
der CIXS3
If necessary, a terminating resistor 470R, 1%, 0,6W
may be soldered between Pin J and K of the connector
Junction Box
e.g.
72583
LV4
2
3
1
LV4
1
10
Cable gland
SIWXS…-3-..-…
or
ISX…-..-…
Option A13
RS485
14-contact
M
J
K
Non-hazardous areaHazardous area
12-contact female
connector on YDI05-Z..
14-contact female
connector on
SIWXS.../ISX...
Important note!
Max. voltage V
m
of any electrical apparatus connected to the interface converter type YDI05-Z.. may not exceed 250V. The internal
Zener voltage V
z
is 12.6V.
Note:
1: not in the scope of delivery
2: up to 8 devices; see „Verification of Intrinsic safety“ (in the manual of these devices)
3: A cable can be rooted diretly into the housing via cable gland YAS04CIS instead of using the 12-contact connector (Option M58).
B
E
C