Product data

36
Signum 3 Ex
Operation
Example:
Numeric input of tare weight,
print the results
e
1) Switch on the scale
The automatic self-test runs
When the weight readout is
shown, the scale is ready to
operate and automatically set
to zero With no load on the
platform, you can zero the
weighing platform at any time
by pressing (
250
2) Enter the tare weight in the
current weight unit using
the keypad (in this example,
250 g)
)
3) Save the tare weight
4) Place the sample (in this
example, 2 kg) in its container
on the scale
Read the result
L
5) Toggle the display from net
to gross weight values
The display shows the gross
weight (in this example, 250 g
for the container plus 2000 g
for the sample)
L 6) Toggle to the previous display
p 7) Print the results
--------------------
G# + 2.250 kg
T + 0.000 kg
PT2 + 0.250 kg
N + 2.000 kg
--------------------
Example:
Tare the scale by placing a container on the weighing platform
e
1) Switch on the scale
The automatic self-test runs
When the weight readout is
shown, the scale is ready to
operate and automatically set
to zero With no load on the
platform, you can zero the
weighing platform at any time
by pressing (
2) Place empty container on the
platform
)
3) Tare the scale
Note: If the automatic tare
function is enabled, you do
not need to press the )
key The tare weight is saved
automatically when you place
the container on the platform
Wait until a zero value is
displayed together with the
NET symbol (net weight)
4) Place sample on the platform
Wait until the weight unit
symbol is displayed (indicating
stability) and then read off the
weight value
Operation