Product data

38
Signum 3 Ex
Operation
Calibration and Adjustment
Purpose
Perform calibration to determine the
difference between the weight value
displayed and the actual weight on
the platform Calibration does not
entail making any changes within the
weighing equipment
During adjustment, the difference
between the measured value displayed
and the true weight of a sample is
corrected, or is reduced to an allowable
level within maximum permissible error
limits
Configuration for Use in Legal
Metrology
Configuration of the weighing
instrument for use in legal metrology
is set by a switch The switch is located
on the back of the weighing platform
and covered by a protective cap
Using a Verified Scale in Legal
Metrology in the EU:
The Type-Approval Certificate for
verified scales is only valid for non-
automatic weighing instruments For
automatic operation with or without
additional, integrated equipment,
please follow the applicable national
regulations for the installation location
§ Before use in legal metrology, the
scale should be adjusted via the
internal adjustment equipment at the
installation location: see the “Internal
Adjustment" section in this chapter
$ The temperature range (°C) listed on
the ID label should not be exceeded
during operation
For Servicing:
External Adjustment for Verified
Scales of Accuracy Class K and l
External adjustment is blocked in legal
metrology (switch cover is sealed)
External adjustment only possible by
removing the seal
If the seal is broken, the validity of
verification will become void and you
must have your scale re-verified
Using a Verified Scale in Legal
Metrology with Internal Adjustment
Equipment:
§ Before use in legal metrology, the
“internal adjustment" function should
be carried out at the installation
location
Characteristics
Which of the following features are
available depends on the connected
weighing platform These features are
configured in the Setup menu:
External calibration/adjustment blocked
in verified weighing instruments
External calibration/adjustment with
the default weight value or standard
weight (not available on verified
instruments):
setup
wp-1
1.9.: Calibration and Adjustment
Specify the weight for external
calibration/adjustment:
setup
wp-1
1.18.: (enter adjustment weight)
Block the J key to prevent use of the
functions described above:
setup
wp-1
1.9.: (Block key 1910)
Calibrate first; then adjust automatically
or manually (not for verified weighing
instruments):
setup
wp-1
1.10.: Calibration/adjustment sequence
Block external or enable calibration/
adjustment:
setup
wp-1
1.16.: (External adjustment)
Calibration Switch
Verification access switch
Calibration
switch
SIWX*DCP
SIWX*DCS
SIWX*BBP
SIWX*BBS
§ Switch on the left for use in legal
metrology
Operation
SIW*DCP/AC:
SIWSDCS:
SIW*BBP:
SIWSBBS:
SIW*DCP/AC:
SIWSDCS:
SIW*BBP:
SIWSBBS:
SIW*DCP/AC:
SIWSDCS:
SIW*BBP:
SIWSBBS:
SIW*DCP/AC:
SIWSDCS:
SIW*BBP:
SIWSBBS: