Product data

40
Signum 3 Ex
Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select Setup: Press the k key repeat-
edly until Setup is displayed
§ Open the Setup menu: Press the
) key
$ Select weighing platform 1 WP1":
Press the ) key or
Setup
wp-1
1
19 Calibration and Adjustment
191 * Ext calibration/adjustment; default weight
193 External calibration/adjustment; weight can be selected under menu item 1181
194 Internal calibration/adjustment (models with built-in motorized calibration weight only)
198 Set preload
199 Clear preload
1910 J key locked
110 Calibration/Adjustment sequence
1101 Calibration with automatic adjustment
1102 * Calibration with manual adjustment
111 Zeroing range
1111 1 percent/max cap
1112 * 2 percent/max cap
112 Initial zero point range
1121 Factory setting (depending on the model type)
1122 2 percent/max cap
1123 * 5 percent/max cap
113 Tare/zero at power on:
1131 * On
1132 Off, load previous tare values
1133 Only zero at power on
116 External adjustment
1
)
1161 * Activated
1162
2
) Blocked
117 Adjustment weight unit
1171 Gram
1172 * Kilogram
1173 Pound
1
)
118 Entering adjustment weight
1181 External user adjustment weight (enter value, eg: 10,000g)
§ Save the settings with the ) key and exit Setup:
Press the ( key several times
1
) Not available on scales verified for use in legal metrology
2
) = Factory setting for use in legal metrology
* Factory Settings
Operation
Operation
40