Signum 3 Ex
Operation
Operation
(
1) Zero the scale
J
2) Start calibration (eg, when
adjustment prompt flashes
WP symbol)
c.ext.def appears
for two
seconds
You are prompted to place the
required weight on the plat-
form (eg, 10,000 g)
3) Position the calibration/
adjustment weight on the
weighing platform
The difference between the
weight value and the true
weight of the sample will be
displayed with plus/minus
signs
Ext. calibration A printout will be generated if
Targ. + 10000 g the process is cancelled using
Diff. + 1 g the ( key
J 4) Activate calibration/adjust-
ment (press the ( key to
cancel)
The adjustment weight is dis-
played once adjustment is fi -
ished
-------------------- A GMP-compliant printout is
24.02.2012 10:15 generated
TypSIWX
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-02
--------------------
Ext. Calibrate
Targ. + 10000 g
Diff. + 1 g
Ext. adjustment
Diff. + 0 g
--------------------
24.02.2012 10:15
Name:
-------------------
Example:
External calibration and manual adjustment with default weights (weighing parameters: factory settings)