Product data

51
Signum 3 Ex
Counting Application Z
Neutral Measurement Application Z
nM
Neutral Measurement
Application Z nM
With this application you can use your
weighing platform to measure the
length, surface and volume of parts
that have roughly the same specific
weight The o symbol is displayed
as the weight unit
Characteristics
Save the reference weight “wRef” from
the weighing platform
Enter the reference weight “wRef” using
the keypad
Enter the calculation factor “nRef”
using the keypad
Activate Info mode with the I key
Toggle the display between
measurement and weight using the
w key
The level of accuracy (display
resolution) can be set when the
calculated reference weight is applied
Automatic taring of container weight
Configured in Setup under:
Appl: a.tare: 3.7.
Automatic initialization when the
scale is switched on The indicator is
initialized with the most recently used
calculation factor “nRef” and reference
weight “wRefConfigured in Setup
under:
Appl: a.start: 3.8.
Exit application, delete parameters:
The value of the reference sample
weight in the reference memory remains
active until it is deleted, overwritten or
the application is changed using the
c key The reference sample weight
also remains saved after the scale is
turned off
You can assign different functions to
the c key for deleting applications
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application Configured in
Setup under:
Appl: sel.cf: 3.24.
(Sel CF function key c in
applications)
Tare function:
1) If you store a tare (weight value)
by pressing the ) key, you can later
enter a tare value manually The tare
value you enter is added to the stored
tare value Setting: menu code 3251
(factory setting)
2) A tare value entered manually
overwrites a stored tare value (weight
value) If you enter a tare value
manually, a tare value (weight value)
stored later overwrites the manually
entered value Setting: Menu code
3252
Configured in Setup under:
Appl
: tare. f: 3.25.
Restore factory default settings
Configured in Setup under:
Appl: def.app: 9.1.
In order to calculate the length, surface
or volume of a given sample, the
average weight of a reference quantity
of the sample must be known (in the
example below, the reference is 1 meter
of electrical cable) There are three ways
to enter the reference weight in the
program:
Calculation:
Place the reference quantity (defined
by the calculation factor) on the
connected weighing platform and
calculate the reference sample weight
by pressing the O key
Place any amount of the sample
material on the connected weighing
platform, enter the calculation factor
through the keypad, and press the
O key to calculated reference sample
weight
How the reference weight is calculated
depends on the application setting for
resolution The value is either rounded
off in accordance with the display
resolution, or saved with 10-fold or 100-
fold resolution, or with the maximum
internal resolution of the weighing
platform
Enter the reference weight (ie, the
weight of one meter of electrical cable)
using the keypad and press
O to save
it
The initial application values remain
active until deleted by pressing the
c key or until overwritten by a new
value They remain saved after the scale
is switched off
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the
) key
§ Select Setup: Press the k key
repeatedly until appl is displayed
§ Confirm appl: Press the ) key
§ Select the Neutral Measurement
application:
Press the k key repeatedly and
confirm with the ) key
Neutral Measurement Application Parameters
36 Minimum load for initialization
361* 1 digit
362 2 digits
363 5 digits
364 10 digits
365 20 digits
366 50 digits
367 100 digits
368 200 digits
369 500 digits
3 610 1000 digits
39 Reference value calculation
resolution
391* Display accuracy
392 Display accuracy
+ 1 decimal place
393 Display accuracy
+ 2 decimal places
394 Internal resolution
310 Decimal places in displayed result
3101* None
3102 1 decimal place
3103 2 decimal places
3104 3 decimal places
311 Parameter for saving weight values
3111* With stability
3112 With increased stability
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times