Product data

52
Signum 3 Ex
Neutral Measurement Application Z
nM
Parameter for Saving Weight Values
The reference weight is saved when the
scale has stabilized
“Stability” is defined as the point at
which the fluctuation of a measured
value lies within a defined tolerance
range The narrower the tolerance
range, the more stable the platform is
at “stability”
In Setup menu:
appl1
:
Neutr.M:3.11.
You can define whether the value is
saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range) If you select
“increased stability”, the reference
weight saved will be more accurate and
the results more reproducible, but the
response time of the weighing platform
might be longer
Accuracy Level for Calculation of
Reference Value
The resolution applied for calculating
the reference weight is defined in Setup
under:
appl1
:
Neutr.M: 3.9.
The resolution for calculating the
reference weight is increased if “+1
decimal place,” “+2 decimal places” or
“Internal resolution” is selected With
the “+1 decimal place” setting, the net
value is determined to one additional
decimal place (ie, display accuracy
x 10); “+2 decimal places” increases
display accuracy x 100, and so on up to
the maximum resolution available
Decimal Places in Displayed Result
In neutral measurement, not only whole
numbers but also decimal numbers
(for example, 125 o electrical cabling)
can be displayed The number of
decimal places displayed in neutral
measurement is configured in Setup
under:
appl1
:
Neutr.M: 3.10.
Minimum Load
The minimum load required for
initialization of the weighing platform
is configured in Setup under:
appl1
:
Neutr.M: 3.6.
Once the limit is exceeded by the load,
initialization can begin
If the load on platform is too light, the
following will occur when you try to
save a value:
Error code Inf 29 appears,
A warning signal is emitted (double-
beep)
The weighing platform is not initialized
The preset calculation factor is saved
The minimum load required for
automatic taring of the container
weight on the platform (first weight)
is configured in Setup under:
appl
: m.weigh: 3.5.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you
must place at least 1000 g (= 1000
intervals) on the weighing platform for
initialization
Neutral Measurement Application Z
nM