Product data

54
Signum 3 Ex
Averaging (Animal Weighing) Application V
Averaging (Animal
Weighing) Application V
With the Averaging application, you
can use your weighing platform for
calculating weights as the average
of a number of individual weighing
operations These individual operations
are also known as “subweighing
operations
This function is used to determine
weights under unstable ambient
conditions or for weighing unstable
samples (such as live animals)
Characteristics
Averaging started manually or auto-
matically Configured in Setup under:
appl1: anim.wg: 3.18.
With manual start selected, the
averaging routine begins when
you press a key (provided the start
conditions are met) With automatic
start selected, averaging begins when
you place the first load on the platform
(provided the start conditions are met)
Enter the number of subweighing
operations using the keypad
Info mode
Toggle the display from “result of last
measurement” to “current weight” by
pressing the
w key
Autmatic printout configured in Setup
under:
Appl1:
anim.wg: 3.20.
Automatic taring of container weight
Configured in Setup under:
appl1:
anim.wg
: 3.7.
Automatic start of averaging when
the scale is turned on and a sample
placed on the platform (provided start
conditions are met) Configured in
Setup under:
Appl: a.start: 3.8.
Exit application, delete parameters:
The number of measurements remain in
memory until it is deleted, overwritten
or the application is changed using the
c key
The number of measurements also
remains saved after the scale is turned
off
You can assign different functions to
the c key for deleting applications
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application Configured in
Setup under:
Appl: sel.cf: 3.24.
(Sel CF function key c in
applications)
Tare function:
If you store a tare (weight value) by
pressing the ) key, you can later
enter a value manually The tare value
you enter is added to the stored tare
value
Setting: menu code 3251 (factory
setting)
A tare value entered manually
overwrites a stored tare value (weight
value) If you enter a tare value
manually, a tare value (weight value)
stored later overwrites the manually
entered value Setting: menu code
3252
Configured in Setup under: Signum
®
2
Appl: tare. f: 3.25.
Restore factory default settings
Configured in Setup under:
Appl: DEF.App: 9.1.
A number of subweighing operations
are required to form the basis for
calculation of an average weight
You can enter the desired number
of subweighing operations using the
keypad
The number you enter is active until
it is overwritten by another number
It also remains in memory when you
switch to a different application
program, or turn off the scale
There are three ways to start the
averaging routine:
Manual start with preset number of
sub-weighing operations:
Place the sample on the platform and
press the O key
Manual start with user-defined number
of subweighing operations:
Place the sample on the platform
and enter the number of weighing
operations using the keypad Press the
r key to save the number entered
and begin weighing
Automatic start with preset number of
subweighing operations:
Measurement begins when you place
the first sample on the platform,
provided the start conditions are met
Averaging (Animal Weighing) Application V