Product data

60
Signum 3 Ex
Checkweighing Application O
Checkweighing
Application O
With the Checkweighing application,
you can check whether the sample on
the weighing platform matches a target
value or lies within a given tolerance
range
Checkweighing also makes it easy to fill
sample materials to a specified target
weight
Characteristics
Enter the nominal or target weight (set
point) and the tolerance range delimiters
either using the keypad or by saving the
weight value of a load on the platform
Enter the tolerance limits as absolute
values (Target, Min and Max) or via the
target value as percentage deviation
with user-defined percent limits or
with relative limits
Configured in the menu under:
Appl 2:
check.wg: 4.5.
The target value can be taken over
as a weighed value from a weighing
platform, and the upper and lower
tolerance limits are defined as a
percentage deviation from the target
value (setting code 452) The deviation
percentage can be changed using the
r key: 01%, 02%, 05%, 1%, 15%,
2%, 3%, 5% or 10%
The target value, lower tolerance limit
(minimum) and upper tolerance limit
(maximum) can be applied as weighed
values from the weighing platform
Target value and tolerance limits
checked during input; values must
conform to: upper limit > target >
lower limit > 1 digit
Checkweighing range: either 30% to
170% of the target, or from 10% to
infinity
Results are shown on main display on
the bar graph, as well as sent to control
output ports for further processing
Toggle the main display between weight
and tolerances limits by pressing the
w key If the weight in the readout is
outside the tolerance range, LL” (too
low) or HH” (too high) is displayed
Activate Info mode with the I key
Automatic printout configured in Setup
under:
Appl 2
:
check.wg: 4.6.
Automatic taring of container weight
Configured in Setup under: Appl
:
a.tare
: 3.7.
Automatic initialization when you
switch on the scale with most recently
saved application data Configured in
Setup under:
appl
: a.start: 3.8.
You can assign different functions to
the c key for deleting applications
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application Configured in
Setup under:
Appl
: sel.cf: 3.24.
(Sel CF function key c in
applications)
Tare function:
1) If you store a tare (weight value)
by pressing the ) key, you can later
enter a tare value manually The tare
value you enter is added to the stored
tare value Setting: menu code 3251
(factory setting)
2) A tare value entered manually
overwrites a stored tare value
(weight value)
If you enter a tare value manually,
a tare value (weight value) stored later
overwrites the manually entered value
Setting: menu code 3252
Configured in Setup under:
Appl: tare. f: 3.25.
Restore factory default settings
Configured in Setup under:
Appl: def.app: 9.1.
Checkweighing entails comparing the
current weight value to a defined target
value You can enter the value for this
target using the keypad, or by saving
the weight value indicated You can
also define upper and lower tolerance
limits based on this target You can do
this by:
Entering absolute values using the
keypad or placing the desired amount
of weight on the platform and saving
the value, or
by entering each value as a percentage of
the target weight
The initialization data remains valid until
deleted by pressing the c key or until
overwritten by a new value They remain
saved after the scale is switched off
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the ) key
§ Select Setup: Press the k key repeatedly
until Setup is displayed
§ Open the Setup menu: Press the ) key
§ Select the Checkweighing application:
Press the k key repeatedly and confirm
with the ) key
Application parameters: Checkweighing
42 Checkweighing range
421* 30 to 170%
422 10% to infinit
43 Control output for SET as:
431* Output “SET”
432 Ready to operate
44 Activation of outputs
441 Off
442 Always on
443 On at stability
444* On within checkweighing range
445 On at stability within
checkweighing range
45 Parameter input
451 Min, max, target value,
absolute values
452* Only target value with percent
limits
453 Target value with free percent
limits
454 Target value; min, max values
as relative weights to target
value
46 Automatic printing
461* Off
462 On
463 Only values within tolerance
464 Only values outside tolerance
47 Checkweighing toward zero
4 7 1* Off
47 2 On
* = Factory setting
§ Save the settings with the ) key and exit
Setup: Press the ( key several times
Application 2: Checkweighing O