Product data
63
Signum 3 Ex
Application 2: Checkweighing O
Checkweighing Toward Zero Application
Checkweighing Toward Zero Application
Example: 2
Checkweighing samples with a target weight of 1250 g and a tolerance range from -10 g to +30 g
Settings (different from the factory settings):
Setup: Application 2: Checkweighing toward zero (parameter 4� 7� 2)
Setup: Printout, PRTPROT 7�6, then select the menu line items of your choice
O
1�) Begin input of target and
tolerance values
2�) Place a sample with the target
weight (in this example,
1250 g) on the platform
O
3�) Save target value
1240
4�) Enter value for lower limit
(in this example, 1240 g)
O
5�) Save value for the lower limit
1280
6�) Enter value for the upper limit
(in this example, 1280 g)
O
Save the value for the upper
limit
7�) Weigh samples
p
8�) Printing results
Note: If automatic printout of
results is enabled, you do not
need to press the p key�
The results are printed
automatically�
Setp + 1.250 kg Target
Min + 1.240 kg Minimum
Max + 1.280 kg Maximum
G# + 1.256 kg Gross weight
T + 0.000 kg Tare weight
N + 1.256 kg Net weight
Lim + 0.48 % Percentage of deviation from
target*
W.Diff+ 0.006 kg Absolute deviation from target
-------------------
* When displayed in relation to
target value:
If the weight is lighter than
the lower limit, the display
shows: LL
If the weight is heavier than
the upper limit, the display
shows: HH