Product data
65
Signum 3 Ex
Classification Application W Classification Application W
Minimum Load
The minimum load for the first class
is configured in Setup under:
class
: 3.6.
Once the limit is exceeded by the load,
initialization can begin�
Once the application is initialized, a
weight value below the minimum load is
designated Class 0; no class is displayed�
The minimum load required for
automatic taring of the container
weight on the platform (first weight),
or for automatic printout of results,
is configured in Setup under:
Appl: m.weigh: 3.5.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform� If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (=1000 intervals)
on the weighing platform for the first
class to activate autotaring or autoprint�
Display
The result of a given measurement is
shown as either a weight value or a class
number�
– Weight display:
The current weight is shown in the
measured value line and the current
class in the text line�
– Display of classes:
The current class is shown in the
measured value line, and the current
weight in the text line�
Digital I|O Interface
The Classification application supports
the digital input/output-interface�
There are 4 control lines, or outputs,
which are activated as follows (see also
the diagram below):
– With 3 classes:
– Class 1
– Class 2
– Class 3
– Set
– With five classes:
– Classes 1/2
– Classes 2/3/4
– Classes 4/5
– Set
In Setup menu:
Appl 2:
class: 4.7.
you can define whether these control
ports are
– off
– always on
– activated at stability
The “SET” output normally changes
its voltage level when the current
weight exceeds the minimum load�
Alternatively, you can assign the
“Ready for use” function to this port�
Configured in Setup under:
Appl 2:
class: 4.3.
Digital I|O Interface
Control lines when working with 3 classes
Digital I|O Interface
Control lines when working with 5 classes
Klasse
123
Mindestlast
Maximallast
Set
Klasse 3
Betriebsbereit
Klasse 1
Klasse 2
Klasse
12345
Mindestlast
Maximallast
Set
Klasse 1/2
Klasse 4/5
Betriebsbereit
Klasse 2/3/4