Product data

65
Signum 3 Ex
Classification Application W Classification Application W
Minimum Load
The minimum load for the first class
is configured in Setup under:
class
: 3.6.
Once the limit is exceeded by the load,
initialization can begin
Once the application is initialized, a
weight value below the minimum load is
designated Class 0; no class is displayed
The minimum load required for
automatic taring of the container
weight on the platform (first weight),
or for automatic printout of results,
is configured in Setup under:
Appl: m.weigh: 3.5.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (=1000 intervals)
on the weighing platform for the first
class to activate autotaring or autoprint
Display
The result of a given measurement is
shown as either a weight value or a class
number
Weight display:
The current weight is shown in the
measured value line and the current
class in the text line
Display of classes:
The current class is shown in the
measured value line, and the current
weight in the text line
Digital I|O Interface
The Classification application supports
the digital input/output-interface
There are 4 control lines, or outputs,
which are activated as follows (see also
the diagram below):
With 3 classes:
– Class 1
– Class 2
– Class 3
– Set
With five classes:
– Classes 1/2
– Classes 2/3/4
– Classes 4/5
– Set
In Setup menu:
Appl 2:
class: 4.7.
you can define whether these control
ports are
off
always on
activated at stability
The “SET” output normally changes
its voltage level when the current
weight exceeds the minimum load
Alternatively, you can assign the
“Ready for use” function to this port
Configured in Setup under:
Appl 2:
class: 4.3.
Digital I|O Interface
Control lines when working with 3 classes
Digital I|O Interface
Control lines when working with 5 classes
Klasse
123
Mindestlast
Maximallast
Set
Klasse 3
Betriebsbereit
Klasse 1
Klasse 2
Klasse
12345
Mindestlast
Maximallast
Set
Klasse 1/2
Klasse 4/5
Betriebsbereit