Product data
66
Signum 3 Ex
Classification Application W
Example:
Defining three classes�
Settings (different from the factory settings):
Setup: Application 2: Classification
Setup: Printout, PRTPROT 7�6, then select the menu line items of your choice
O
1�) Begin input of class delimiters
110
2�) Enter the upper limit for
Class 1 using the keypad
(in this example, 110 g)
O
3�) Save the upper limit for
Class 1
130
4�) Enter the upper limit for
Class 2 using the keypad
(in this example, 130 g)
O
5�) Save the upper limit for
Class 2
6�) Place the sample on the
weighing platform
Read the result
p
7�) Printing results
Note: If automatic printout of
results is enabled, you do not
need to press the p key�
The results are printed
automatically�
Lim1 + 0.110 kg see chapter:
wRef + 0.130 kg Configuring Printouts
G# + 0.118 kg
T + 0.000 kg
N + 0.118 g
Class 2
-------------------
+
Totalizing Application s