Product data

68
Signum 3 Ex
Totalizing Application s
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the
) key
§ Select Setup: Press the k key
repeatedly until appl is displayed
§ Confir appl: Press the ) key
§ Select the Totalizing application:
Press the k key repeatedly and
confirm with the ) key
Totalizing Application Parameters
3 6 Minimum load for autosave
361* 1 digit
362 2 digits
363 5 digits
364 10 digits
365 20 digits
366 50 digits
367 100 digits
368 200 digits
369 500 digits
3610 1000 digits
316 Autosave mode
3161* Off
3162 On
317 Individual/Component printout
when saved
3171 Automatic printing off
3172* Print the entire standard
print configuration
every time with the
O key
322 Source of values for data saved
automatically
3221* Application 1
3222 Application 2
323 Saved value
3231* Net
3232 Calculated
3233 Net and Calculated
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times
Minimum Load
The minimum load required for
automatic taring of the container
weight on the platform (first weight)
is configured in Setup under:
Appl: m.weigh: 3.5.
The minimum amount that a
component must weigh before it
can be saved in totalizing memory
is configured in Setup under:
appl 3
:
totaliz: 3.6.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals)
on the weighing platform for autotaring
(only with the “Autotare first weight”
option selected)
Printout
In Setup menu:
appl 3:
totaliz: 3.17.
You can configure whether a printout
is generated manually by pressing
the p key or automatically when a
weight value is stored in the totalizing
memory
If you select the 3.17.1 menu item,
printouts can only be generated
manually by pressing the p key
(single printout)
If the 3.17. 2. menu item is selected
(Print one component), the component
record is printed
The total data record is printed when
you clear the totalizing memory
(by pressing the c key)
Totalizing Application s