Product data

70
Signum 3 Ex
Net-total Formulation Application R
Net-total Formulation
Application R
With this application, you can weigh in
different components up to a defined
total Each component is saved in the
net-total memory
Characteristics
Weigh in up to 999 components in
series
Net total formulation cannot be
combined with a level 1 or level 2
application
Current component number displayed
in the text lines (indicating the
component to be added)
Toggle the display from “component
mode” to “additive mode” by pressing
the w key
Component mode: Display the weight
of the component currently on the
platform (for 1 second after it is
saved; then the platform is tared)
Additive mode: Display the weight of
all components on the platform (after
it is saved, the net weight of the last
component added is displayed briefly
Activate Info mode with the I key
Automatic component printout when it
is saved Configured in Setup under:
appl 3:
Net Tot: 3.17.
If the 3.17.2 menu item is selected,
the entire component record is printed
If the 3.17.3 menu item is selected, the
following items are generated only once
for the first component:
Blank line, date, time, ID1 through ID4,
header lines 1 and 2 For subsequent
components, each “component” item
(“Comp xx”) is followed by a blank line
Automatic taring of container weight
Configured in Setup under:
Appl: a.tare: 3.7.
Restore factory default settings
Configured in Setup under:
Appl: def.app: 9.1.
Net-total Formulation Application R