Unit installation
Platform Specifications
Three different types of weighing technology are used in the Signum Series, offering different performance levels.
Signum Regular:
Model code SIWRDCP-1,-2,-3
Type, accuracy class DG SI 300, l
Weighing capacity (kg) -3 -6
Resolution code -I -N -R -L -I
Weighing capacity (kg) 3 3 3/6 6666
Readability d 0.1 1 1/2 1 2 0.5 0.2
Verifiable/verified version -BCE -NCE -RCE -BCE
Readability e (g) 1 1/2 1 2
Preload (kg) 1.2 1 .2
Repeatability [s] (g) 0.2 Meets EN 45501 0.2
Linearity (g) 0.3 0.4
Ambient temperature –10°C to +40°C (14°F to 104°F)
(only for use in
legal metrology)
Calibration weight (g) 2,000 5,000 5,000 5,000 5,000 5,000
- Accuracy class M1 M1 M1 M1 M1 M1
Signum Regular:
Model code SIWRDCP-1,-2,-3
Type, accuracy class DG SI 300, l
Weighing capacity (kg) -15
Resolution code -N -R -L -I
Weighing capacity (kg) 6/15 15 15 15 15
Readability d 2/5 2 5 1 0.5
Verifiable/verified -NCE -RCE -BCE
Readability e (g) 2 5
Preload (kg) 3 3 3
Repeatability [s] (g) Meets EN 45501 0.2
Linearity (g) 0.8
Ambient temperature –10°C to +40°C (14°F to 104°F)
(only for use in
legal metrology)
Calibration weight (g) 5,000
- Accuracy class M1
Signum Regular:
Model code SIWRDCP-1,-2,-3
Type, accuracy class DG SI 300, l
Weighing capacity (kg) -35
Resolution code -N -R -L -I -M
Weighing capacity (kg) 15/35 35 35 35 15/35
Readability d 5/10 5 10 2 1 5/10
Verifiable/verified version -NCE -RCE -BCE
Readability e (g) 5/10 5 10
Preload (kg) 6
Repeatability [s] (g) 1
Linearity (g) 1.5
Ambient temperature –10°C to +40°C (14°F to 104°F)
(only for use in
legal metrology)
Calibration weight (g) 10,000
- Accuracy class M1
96