Series PM174 Powermeter and Power Quality Analyzer Installation and Operation Manual BG0411 Rev.
LIMITED WARRANTY The manufacturer offers the customer a 24-month functional warranty on the instrument for faulty workmanship or parts from date of dispatch from the distributor. In all cases, this warranty is valid for 36 months from the date of production. This warranty is on a return to factory basis. The manufacturer does not accept liability for any damage caused by instrument malfunction.
 Read this manual thoroughly before connecting the device to the current carrying circuits. During operation of the device, hazardous voltages are present on input terminals. Failure to observe precautions can result in serious or even fatal injury or damage to equipment. All trademarks are property of their respective owners.
Table of Contents Chapter 1 General Information......................................................... 7 Chapter 2 Installation...................................................................... 11 Mechanical Installation ................................................................................... 11 Panel Mounting ........................................................................................................11 DIN Rail Mounting ......................................................
Device Options.........................................................................................................46 Communication Ports ...............................................................................................46 Network Address......................................................................................................48 Counters Setup ........................................................................................................48 Control Setpoints Setup ......
Administration ............................................................................................... 109 Changing a Password ............................................................................................. 109 Upgrading Device Firmware ......................................................................... 109 Monitoring Devices........................................................................................ 111 Viewing Viewing Viewing Viewing Real-time Data ...............
Chapter 1 General Information Chapter 1 General Information The PM174 is a compact, multi-function, three-phase AC powermeter and power quality analyzer specially designed to meet the requirements of users ranging from electrical panel builders to substation operators. Bright 3-row LED display provides easy local meter readings. The display module is freely detachable and can be located at a distance of up to 1000 meters from the device.
Chapter 1 General Information • Event recorder for logging internal diagnostics events, control events and I/O operations • 16 data recorders; programmable data logs on a periodic basis and on any internal and external trigger • Two waveform recorders; simultaneous 6-channel AC recording in a single plot; sampling rate of 32, 64 and 128 samples per cycle; 20 pre-fault cycles; up to 30 seconds of continuous recording at a rate of 32 samples per cycle • IEEE 1159 Power Quality recorder (onboard PQ anal
Chapter 1 General Information RS-422/RS-485 • Communication protocols: Modbus RTU, Modbus ASCII, Modbus/TCP, DNP3, DNP3/TCP, GE EGD producer, and Profibus DP (with firmware V24.2.
Chapter 1 General Information Parameter Import & Export kVA Predicted Demand kW Maximum Demand Import kW Maximum Demand Export kvar Maximum Demand Import kvar Maximum Demand Export kVA Maximum Demand Total Energy Total kWh Import & Export Total kvarh Import & Export Total kvarh Net Total kVAh Energy per Phase kWh Import per phase kvarh Import per phase kVAh per phase TOU Registers 8 TOU energy registers (kWh and kvarh import & export, kVAh, 2 pulse sources) 8 TOU maximum demand{ XE "demand" } registers 8 ta
Chapter 2 Installation Mechanical Installation Chapter 2 Installation Mechanical Installation Panel Mounting Figure 2-1 Dimensions Figure 2-2 STEP 1 (ANSI 4" round cutout): Mount the display module in cutout Series PM174 Powermeters 11
Chapter 2 Installation Mechanical Installation Figure 2-3 STEP 1 (DIN 92x92mm square cutout): Mount the display module in cutout Figure 2-4 STEP 2: Assemble the four locating studs 12 Series PM174 Powermeters
Chapter 2 Installation Mechanical Installation Figure 2-5 STEP 3: Slide and position the meter on locating studs Figure 2-6 STEP 4: Affix the meter using the thumb nuts Series PM174 Powermeters 13
Chapter 2 Installation Mechanical Installation DIN Rail Mounting The PM174 can be mounted on a 35-mm DIN rail. The display module is mounted separately on the switchboard panel and is connected to the meter by a communication cable (see “Remote Display Installation”).
Chapter 2 Installation Remote Display Installation Remote Display Installation Mechanical Installation Standard Cutouts (ANSI 4" round or DIN 92x92 mm square) Figure 2-9 Display cutout dimensions STEP 1: Insert the display module into cutout. STEP 2: Fasten washers and nut on screws.
Chapter 2 Installation Remote Display Installation Special Cutout Figure 2-11 Panel cutout dimensions Figure 2-12 Display mounting 16 Series PM174 Powermeters
Chapter 2 Installation Remote Display Installation Electrical Connection The remote display is connected to the meter via a 3-wire or 5-wire communication cable provided with two 15-pin D-type connectors. At distances of up to 100 m, the display receives power through the communication cable directly from the meter. Connect pins 1 and 8 on both sides as shown in Figure 2-13. At distances above 100 m, supply power from a separate 12V DC power source (a 12V AC/DC adapter can be used).
Chapter 2 Installation Electrical Installation Electrical Installation Before installation ensure that all incoming power sources are shut OFF. Failure to observe this practice can result in serious or even fatal injury and damage to equipment. Typical Installation DIGITAL INPUTS + - + 2 1 ANALOG INPUTS ANALOG OUTPUTS COM.1 RELAYS COM.
Chapter 2 Installation Electrical Installation Terminals Analog Inputs/Outputs Relay Outputs AC Current Inputs Digital Inputs 18 19 20 1 2 DIGITAL INPUTS 26 27 28 29 1 2 21 22 23 24 25 + - + 2 1 ANALOG INPUTS ANALOG OUTPUTS RELAYS + 1 1 - 3 2 V1 5 V2 8 V3 AC Voltage Inputs 11 VN COM1 Port O 1A CT. 690V P 5A CT. OPT.
Chapter 2 Installation Electrical Installation Chassis Ground Connection Connect the chassis ground of the device to the switchgear earth ground using a dedicated wire greater than 2 mm2/14 AWG. Wiring Diagrams For AC input ratings, see “Technical Specifications” in Appendix A.
Chapter 2 Installation Electrical Installation Figure 2-18 4-Wire Wye 3-Element Direct Connection Using 3 CTs. Wiring Mode = 4LL3 or 4Ln3 Figure 2-19 4-Wire Wye 3-Element Connection Using 3 PTs, 3 CTs.
Chapter 2 Installation Electrical Installation Figure 2-20 3-Wire 2-element Open Delta Connection Using 2 PTs, 2 CTs. Wiring Mode = 3OP2 Figure 2-21 4-Wire Wye 2½-Element Connection Using 2 PTs, 3 CTs.
Chapter 2 Installation Electrical Installation This configuration provides accurate power measurements only if the voltages are balanced. Figure 2-22 3-Wire Open 2½-Element Delta Connection Using 2 PTs, 3 CTs. Wiring Mode = 3OP3 Figure 2-23 4-Wire 3-Element Delta Direct Connection Using 3 CTs.
Chapter 2 Installation Electrical Installation Wiring Mode = 4LL3 or 4Ln3 Figure 2-24 24 3-Wire 2½-Element Broken Delta Connection Using 2 PTs, 3 CTs.
Chapter 2 Installation I/O Connections I/O Connections For I/O ratings, see “Technical Specifications” in Appendix A. Relay Outputs LOAD LOAD N 10A FUSE LINE 21 22 23 24 25 + - + 2 1 ANALOG INPUTS ANALOG OUTPUTS 18 19 20 1 2 DIGITAL INPUTS 26 27 28 29 1 2 RELAYS + 1 1 - 3 2 V1 S/N O 1A CT. 690V + P 5A CT. OPT.
Chapter 2 Installation I/O Connections Analog Outputs LOAD SHIELD + _ + PROTECTIVE GROUND _ 21 22 23 24 25 + - + 2 1 ANALOG INPUTS ANALOG OUTPUTS 18 19 20 1 2 DIGITAL INPUTS 26 27 28 29 1 2 RELAYS + 1 1 - 3 2 V1 S/N POWER SUPPLY 90-264VAC 50/60Hz 5 V2 STANDARD 10W 85-290VDC O 1A CT. 690V + P 5A CT. OPT.
Chapter 2 Installation Communications Connections Communications Connections Several communication options are available for the PM174: COM1 (check the label on the back of your meter): RS-232/RS-422/RS-485 56K Dial-up modem Ethernet 10/100BaseT Profibus DP (with firmware V24.2.1 and later) COM2: RS-422/RS-485 The RS-232/RS-422/RS-485 port is a standard port for COM1. Other options are ordered separately.
Chapter 2 Installation PM174 RS232 MALE CON. Communications Connections IBM PC/COMPATIBLE 25-PIN DB25 FEMALE CON. IBM PC/COMPATIBLE 9-PIN DB9 FEMALE CON. PM174 RS232 MALE CON. RS-232 SIMPLE 3-WIRE CONNECTION 9-PIN RS-232 SIMPLE 3-WIRE CONNECTION 25-PIN Figure 2-29 COM1: RS-232 Cable Drawings COM1 RS-422/485 Connection V2 (12) 10-16VDC LOW DC (24) 18-36VDC (48) 36-72VDC 8 N S COM.
Chapter 2 Installation Communications Connections COM1 Dial Up Modem Connection AC0140 05-12001-3 Figure 2-31 COM1: Telephone Line Connection COM1 Ethernet Connection AC0139 RJ45 05-12001-4 Figure 2-32 COM1: Ethernet Connection Series PM174 Powermeters 29
Chapter 2 Installation Communications Connections COM1 Profibus Connection V2 (12) 10-16VDC LOW DC (24) 18-36VDC (48) 36-72VDC 8 N S COM.1 : RS-232/422/485 STANDARD ETHERNET 3 MODEM PROFIBUS ATTENTION V 5 9 VN Static-Sensitive Devices Handle Only at Static-Safe Workstations N/- 10 POWER SUPPLY COM.2 L/+ 12 RS-422/RS-485 COM.
Chapter 2 Installation Communications Connections V2 (12) 10-16VDC LOW DC (24) 18-36VDC (48) 36-72VDC 8 COM.1 : RS-232/422/485 STANDARD ETHERNET 3 MODEM PROFIBUS N S ANALOG IN/OUT : +1mA + 7 0-20mA 3 0-1mA - 9 4-20mA ATTENTION V 5 9 VN 10 POWER SUPPLY COM.2 L/+ 12 RS-422/RS-485 COM.
Chapter 2 Installation Communications Connections V2 (12) 10-16VDC LOW DC (24) 18-36VDC (48) 36-72VDC 8 COM.1 : RS-232/422/485 STANDARD ETHERNET 3 MODEM PROFIBUS V 11 VN COM.1 5 9 1 6 N S ANALOG IN/OUT : +1mA + 7 0-20mA 3 0-1mA 4-20mA - 9 ATTENTION Static-Sensitive Devices Handle Only at N/Static-Safe Workstations 10 POWER SUPPLY COM.
Chapter 3 Display Operations Indicators and Controls Chapter 3 Display Operations Displayed Parameters Load Bar Graph Wh/varh Pulse LED Measurement Units Port Activity LEDs Navigation Buttons Indicators and Controls Display Diagnostics diSP Con.Err The display may indicate a connection error as shown on the left picture if it fails to establish a connection with the meter. Check the connection between the display module and the meter body.
Chapter 3 Display Data Display Energy Pulse LED The PM174 has a red “Energy Pulse” LED. It flashes at a constant rate when a load is applied to the meter. There are two modes of LED operation: normal and test. In normal mode, the LED pulses indicate imported Wh at a rate of 1,000 pulses per kWh. In test mode, the LED pulses indicate either imported Wh, or imported (inductive) varh at a rate of 10,000 pulses per kWh/kvarh. The energy test mode can be enabled through the Display Setup menu.
Chapter 3 Display Operations Data Display Primary and Secondary Volts Volts can be displayed in primary (default) or secondary units. The volts display mode can be changed through the Display Setup menu. Phase Power Readings In configurations with the neutral wire, in addition to total three-phase powers, the meter can show per-phase power readings. By default, they are disabled. See Display Setup on how to enable per-phase power readings in your meter.
Chapter 3 Display Data Display Pressing both the UP and DOWN arrow buttons together returns to the first page within the current display. The SELECT button operates once it’s released. The button has two functions: • When pressed briefly, it switches to programming mode. • When pressed together with the ENTER button for more than 5 seconds, it resets Min/Max records, maximum demands, or energies depending on the currently displayed page.
Chapter 3 Display Operations Data Display 5 S T 6 7 8 9 10 11 12 13 14 15 16 Common Measurements (Main Display) Neutral current In Frequency Hz Total var kvar/Mvar Phase L1 powers Ph.L1 (if enabled) PF kW/MW Phase L1 powers kVA/MVA (if enabled) Ph.L1 kvar/Mvar Phase L2 powers Ph.L2 (if enabled) PF kW/MW Phase L2 powers kVA/MVA (if enabled) Ph.L2 kvar/Mvar Phase L3 powers Ph.L3 (if enabled) PF kW/MW Phase L3 powers kVA/MVA (if enabled) Ph.
Chapter 3 Display Data Display 5 S 6 T 7 8 9 10 11 Min/Max and Maximum Demands Maximum volts V1/V12 V2/V23 V3/V31 Hi Maximum amps I1 I2 I3 Hi Maximum total VA kVA/MVA Maximum total PF PF (absolute) kW/MW Hi Maximum total W Maximum neutral In current Hz Maximum frequency kvar/Mvar Hi Maximum total var Maximum volt demands V1/V12 V2/V23 V3/V31 Hd Maximum ampere I1 demands I2 I3 Hd Maximum VA demand kVA/MVA PF at maximum VA PF demand kW/MW Hd Maximum W import demand 12 Hd 38 kvar/Mvar Maximum var
Chapter 3 Display Operations Data Display Power Quality/Harmonics Display Press the PQ/ESC button. The THD/TDD LED is illuminated. Press the button again to move to the individual harmonics, or to the flicker and unbalance displays. Use the UP and DOWN arrow buttons to scroll through harmonics and power quality measurements. Note that voltage harmonics readings are line-to-neutral in the 4LN3, 3LN3 and 3BLN3 wiring modes, and line-to-line in all other modes. PQ ESC 1 thd. S 2 T 3 thd. tdd.
Chapter 3 Display Data Display 4 Flicker/Unbalance C.Unb I% unb Current negative sequence unbalance, percent Energy Display Press the ENERGY button. The MVAh, Mvarh, or MWh LED is illuminated. If TOU registers are configured in the meter, press the button again to scroll through all active TOU registers. Use the UP and DOWN arrow buttons to scroll through { XE "demand" }energy pages.
Chapter 3 Display Operations Status Display 1 S T 2 8 TOU Energy Register 1 Tariff 1 reading rEG.1 trF.1 MWh Tariff 1 reading rEG.1 trF.2 MWh rEG.1 trF.8 MWh Tariff 8 reading … ENERGY ENTER S T 1 2 8 TOU Energy Register 8 Tariff 1 reading rEG.8 trF.1 MWh Tariff 1 reading rEG.8 trF.2 MWh rEG.8 trF.8 MWh Tariff 8 reading Status Display The meter has separate status information pages accessible through the primary device menu.
Chapter 3 Display Using the Menus 8 Status Display Cnt.3 9 Counter #3 Cnt.4 10 Counter #4 batt Backup battery status (Normal/Low) nor/Lo Using the Menus Navigation Buttons THD/TDD ESC S T SELECT ENERGY ENTER The PM174 has a menu-driven setup. To enter the menus, press and release the SELECT button. The SELECT button selects (highlights) an active window in which you can select or change a desired menu item. The button operates once it’s briefly pressed and released.
Chapter 3 Display Operations Using the Menus 1. 2. Press the SELECT button to activate the OPS window. Press the ENTER button to enter the main menu. To change the meter setup, or to clear the accumulated values: 1. 2. Press the SELECT button to activate the CHG window. Press the ENTER button to enter the main menu. Entering the Password The setup change menu can be secured by a four-digit user password. The meter is primarily shipped with the password preset to 0 and password protection disabled.
Chapter 3 Display Menu Operations Viewing and Changing Setup Items A second level menu normally consists of three items: the upper static window indicates the menu name, while the middle window represents a list of setup parameters you can scroll through, and the lower item shows the present parameter value. To select a parameter you want to view or change: 1. Highlight the middle window by pressing the SELECT button. bASc bASc S ConF Pt 1.0 4Ln3 2.
Chapter 3 Display Operations Menu Operations 2. Use the UP and DOWN arrow buttons to scroll to the desired option. To change the option: Press the SELECT button to activate the lower window. Use the UP and DOWN arrow buttons to select the desired option. Press ENTER to confirm your changes and to store your new setting, or press ESC to discard changes. 1. 2. 3. To exit the menu, press ESC. The following table lists available options.
Chapter 3 Display Menu Operations Device Options OPtS P.cAL rEAc This menu allows you to change the user-configurable device options or put the meter into energy test mode. To enter the menu, select the “OPtS” entry from the main menu, and then press the ENTER button. To select a setup option: 1. 2. Press the SELECT button to activate the middle window. Use the UP and DOWN arrow buttons to scroll to the desired option. To change the option: 1. 2. 3.
Chapter 3 Display Operations Menu Operations 2. Use the UP and DOWN arrow buttons to scroll to the desired option. To change the option: 1. 2. 3. Press the SELECT button to activate the lower window. Use the UP and DOWN arrow buttons to select the desired option. Press ENTER to confirm your changes and to store your new setting, or press ESC to discard changes. To exit the menu, press ESC. The following tables list available port options.
Chapter 3 Display Menu Operations Label Parameter Options Default bAud Baud rate 300-115200 bps 19200 bps The port baud rate dAtA Data format and 7E, 8N, 8E parity 8N Description 7E data format should not be used with the Modbus RTU and DNP3 protocols Network Address A. 192. 168. 000.203 This menu allows you to configure the device IP address and the default gateway address for the Ethernet port. To enter the menu, select “nEt” from the main menu, and then press the ENTER button.
Chapter 3 Display Operations Menu Operations To exit the menu, press ESC. The following table lists available counter options. Parameter Options Default Description Pulse source None = disabled Inp.1 = DI1 Inp.
Chapter 3 Display Menu Operations Control Setpoints Setup SEtP ESC SP.1 trG.1 rtHi.C1 SP.1 On.1 200 The PM174 provides 16 control setpoints with programmable operate and release delays. Each setpoint evaluates a logical expression with up to four arguments using OR/AND logic. Whenever an expression is evaluated as “true”, the setpoint performs up to four concurrent actions that can send a command to the output relays, increment or decrement a counter, or trigger a recorder.
Chapter 3 Display Operations A.In.1 ESC A.In.1 Lo 0 A.In.1 Hi 230 A.In.1 dEc.P Menu Operations Label Parameter On.1-On.4 Operate limit Options The threshold (in primary units) at which the conditional expression would be evaluated to true. Not applicable for digital triggers. Description OFF.1OFF.4 Release limit The threshold (in primary units) at which the conditional expression would be evaluated to false. Defines the hysteresis for analog triggers. Not applicable for digital triggers. Act.
Chapter 3 Display Menu Operations Analog Outputs Setup A.Ou.1 ESC A.Ou.1 OutP rt.U1 A.Ou.1 Lo 0 This entry appears only if the meter is ordered with optional analog outputs. For more information on configuring analog inputs in your meter, see “Programming Analog Outputs” in Chapter 4. To enter the menu, select the “A.Ou.1” or “A.Ou.2” entry from the main menu for the AO1 and AO2 output respectively, and then press the ENTER button. To change the analog output options: 1. 2. 3. 4. 5. A.Ou.
Chapter 3 Display Operations Menu Operations associated with the analog expander with address 0, while outputs A2-1 through A2-8 are associated with the analog expander with address 1. The menu uses three entries: 1. The upper window indicates an analog expander’s output channel. 2. The middle window selects a setup parameter to view or change. 3. The lower window displays the parameter value. Use the UP and DOWN arrow buttons to scroll to the desired analog expander channel.
Chapter 3 Display Menu Operations Timers Setup t-r t-r.1 0 The PM174 is provided with four interval timers. When enabled, a timer generates periodic events in predefined intervals that can trigger setpoints to produce periodic actions like periodic data trending. To enter the menu, select the “t-r” entry from the main menu and press the ENTER button. Use the UP and DOWN arrow buttons to scroll to the desired timer. To change the time period for the timer: 1. 2. 3. 4.
Chapter 3 Display Operations Menu Operations Label Fund. dAtE Parameter Fundamental component display mode Date order Options diS, En dnY, ndY, Ynd (d=day, n=month, y=year) Default Disabled Description Disables or enables fundamental values in the main display mm.dd.yy Defines the date order in the RTC display Meter Security AccS ESC The Access Control menu allows you to change the user password and enable or disable password protection.
Chapter 3 Display Menu Operations To select a setup option, use the UP and DOWN arrow buttons from the upper window. To change the time, date, or daylight savings setting: dAtE 1. 01.23. 05. 2. 3. 4. Highlight an item you want to change by pressing briefly the SELECT button. When you enter the time setup display, the hours and minutes are frozen to allow you to adjust them. Adjust the selected item with the UP and DOWN arrow buttons.
Chapter 3 Display Operations Menu Operations The Reset menu allows you to separately reset minimum/maximum log records, maximum demands and counters. To reset the desired registers: 1. Highlight the middle window by pressing briefly the SELECT button. Select the desired entry by scrolling through the list with the UP and DOWN arrow buttons until the desired entry appears. Press the SELECT button briefly to highlight the lower item. Press and hold the ENTER button for 5 seconds. Release the button.
Chapter 4 PAS Application Configuration Database Chapter 4 PAS Application Software Supplemental PAS software can be used for configuring the PM174 through communication ports, for retrieving real-time and recorded data, and for remote upgrading device firmware. For information on how to install PAS on your PC, see the “PAS Getting Started” guide supplied on the installation CD. Configuration Database To communicate with your meters, create a separate site database for each device.
Chapter 4 PAS Application Software Setting up Communications your meter’s COM1 port can be equipped with an RS-232/RS422/RS-485 serial interface, with a dial-up modem for communicating through public telephone lines, with an Ethernet module for communicating through the Internet, or with a Profibus port. PAS does not support communications through Profibus. To configure your communications with the PM174: 1. 2. 3. Select Configuration from the Tools menu.
Chapter 4 PAS Application 2. Setting up Communications In the “Protocol” box, select the same communications protocol as you have in your meter. The default protocol setting in your meter for all ports is Modbus RTU. For more information on configuring the protocol parameters, refer to the “PAS Getting Started” guide. Communicating through a Dial-up Modem Configuring a Modem 1. 2. 3. 4. On the Connection tab, select a local modem installed on your PC.
Chapter 4 PAS Application Software Setting Up the Meter 3. 4. 5. Click on the “IP address” and type in the IP address of your meter. The default IP address preset at the factory is 192.168.0.203. In the “Protocol” box, select the communications protocol for the TCP port. The meter can provide Modbus/TCP connections on TCP port 502 and DNP3/TCP connections on port 20000. The host port is set automatically as you select the protocol. Select “Modbus RTU” for Modbus/TCP or “DNP3” for DNP3/TCP.
Chapter 4 PAS Application Authorization Downloading Setup to the Meter You can update each setup in your meter one at a time or download all setups together from the site database. To update a particular setup in your device, check the On-line button on the PAS toolbar, select a meter site from the list box on the toolbar, and then select the desired setup group from the Meter Setup menu. Click on the tab of the setup you want to download to the meter, and then click Send.
Chapter 4 PAS Application Software Configuring Communications in your Meter To change the port settings in your meter, select desired port parameters, and then click Send. For the available communication options, see “Communication Ports” in Chapter 3. NOTES 1. In meters with the Ethernet option, the device provides the permanent Modbus TCP server on port 502.
Chapter 4 PAS Application Configuring Communications in your The following table lists available network options. Parameter Options Default Device IP Address 192.168.0.203 Network Subnet Mask 255.255.255.0 Network Default Gateway 192.168.0.1 TCP Service Port 502 = Modbus/TCP 20000 = DNP3/TCP 502 Ì The TCP service port can also be changed trough the COM1 serial port setup: changing the protocol for the port automatically changes the TCP port for the Ethernet.
Chapter 4 PAS Application Software Configuring Communications in your Meter exchange is individually configurable and can produce and send data to the same or to different consuming nodes. For EGD implementation and configuration details, see the PM174 GE EGD Communications Protocol Reference Guide. To enter the Setup dialog, select the device site from the list box on the PAS toolbar, select Communications Setup from the Meter Setup menu, and then click on the GE EGD Producer Setup tab.
Chapter 4 PAS Application 6. Configuring Communications in your data. PAS automatically truncates your range if you specify more data than the exchange can accommodate. To indicate the end of the address range list, put a zero start point ID in the range following the last range you specified. Click on Send to download your new setup to the meter. The device will reject your setup if you specified an inexistent start point ID for one of the address ranges.
Chapter 4 PAS Application Software General Meter Setup General Meter Setup This section describes how to configure the PM174 for your particular environment and application using PAS. Basic Meter Setup Before operating your meter, provide the device with basic information about your electrical network. To enter the Setup dialog, select the device site from the list box on the PAS toolbar, and then select General Setup from the Meter Setup menu.
Chapter 4 PAS Application General Meter Setup Parameter Options Default Description Nominal frequency 50,60 Hz 60 Hz The nominal line frequency 1, 2, 5, 10, 15, 20, 30, 60 min, E=external sync 30 min The length of the demand period for power demand calculations. If the external synchronization is selected, a pulse front on the digital input DI1 denotes the start of the demand interval.
Chapter 4 PAS Application Software General Meter Setup Parameter Options Default Description Energy Roll Value 1000.0 kWh 10000.0 kWh 100000.0 kWh 1000000.0 kWh 10000000.0 kWh 100000000.0 kWh 100000000.0 The value at which energy counters roll over to zero Phase Energy Calculation Disabled, Enabled Disabled Enables phase energy calculations Analog Expander 1 Option Disabled 0-20 mA 4-20 mA 0-1 mA ±1 mA Disabled Enables outputs for the AX-8 analog expanders through port COM2.
Chapter 4 PAS Application General Meter Setup 1. When the reactive power calculation mode is selected, active and reactive powers are measured directly and apparent power is calculated as: S = P 2 + Q2 This mode is recommended for electrical networks with low harmonic distortion, commonly with THD < 5% for volts, and THD < 10% for currents. In networks with high harmonics, the following method is preferable. 2.
Chapter 4 PAS Application Software General Meter Setup Parameter Options DST start month DST start week DST start weekday The date when Daylight Savings Time First Month-weekSunday in begins. The DST switch point is weekday specified by the month, week of the Week = 1st, 2nd, 3rd, April month and weekday. By default, DST 4thor Last (last week starts at 2:00 AM on the first Sunday of the month) in April of each year.
Chapter 4 PAS Application General Meter Setup The available options are shown in the following table. Parameter Options Pulse input mode PULSE MODE, KYZ MODE Default Description PULSE MODE In pulse mode, either leading, or trailing edge of the input pulse is recognized as an event. In KYZ mode, both leading and trailing edges of the input pulse are recognized as separate events. Pulse polarity NORMAL (N.O.), INVERTING (N.C.
Chapter 4 PAS Application Software General Meter Setup The available relay options are shown in the following table: Parameter Options Default Description Operation mode UNLATCHED Unlatched mode: the relay goes into its active state when the control setpoint is in active (operated) state, and returns into its non-active state when the setpoint is released.
Chapter 4 PAS Application General Meter Setup Parameter Options Default Description Retentive mode NO YES NO Applicable for latched relays. In non-retentive mode, the relay always returns to its non-active state upon power up. In retentive mode, the relay status is restored to what it was prior to loss of power. Pulse width 20-1000 ms 100 ms The actual pulse width is a multiple of the 1/2-cycle time rounded to the nearest bigger value. The pause time between pulses is equal to the pulse width.
Chapter 4 PAS Application Software General Meter Setup To configure the Analog Inputs in your device, select General Setup from the Meter Setup menu, and click on the Analog Inputs tab. If you are programming your device online, analog inputs are designated as not available if they are not present in the device. The available options are described in the following table.
Chapter 4 PAS Application General Meter Setup zero. The device does not allow you to access this setting. Whenever the direction of the input current is changed to negative, the device automatically uses your full engineering scale settings for +1 mA with a negative sign. Scaling Analog Inputs for 0-2 mA and ±2 mA The input scales for 0-1 mA and ±1 mA analog inputs are always programmed for 0 mA and +1 mA regardless of the desired input range.
Chapter 4 PAS Application Software General Meter Setup When you select an output parameter for the analog output channel, the default engineering scales are set automatically. They represent the maximum available scales. If the parameter actually covers a lower range, you can change the scales to provide a better resolution on the analog outputs.
Chapter 4 PAS Application General Meter Setup Power Factor” above). The device does not allow you access to this setting if the parameter is directional. Whenever the sign of the output parameter is changed to negative, the device automatically uses your full engineering scale settings for +1 mA with a negative sign. Scaling Analog Outputs for 0-2 mA and ±2 mA The output scales for 0-1 mA and ±1 mA analog outputs are programmed for 0 mA and +1 mA regardless of the desired output current range.
Chapter 4 PAS Application Software General Meter Setup expander with address 0, while channels 9 through 16 are associated with the analog expander with address 1. The available analog expander outputs options are described in the following table. Option Output parameter Range See Appendix B Description Selects the measured parameter to be transmitted through the analog output channel.
Chapter 4 PAS Application General Meter Setup counter can also be incremented in response to any internal or external event, and checked and cleared through the Control Setpoints. The following table lists available options.
Chapter 4 PAS Application Software General Meter Setup Using Control Setpoints The PM174 has an embedded logical controller that runs different actions in response to user-defined internal and external events. Unlike a PLC, the meter uses a simplified programming technique based on setpoints that allows the user to define a logical expression based on measured analog and digital values that produce a required action. The meter provides 16 control setpoints with programmable operate and release delays.
Chapter 4 PAS Application Option General Meter Setup Range Release limit Description The threshold (in primary units) at which the conditional expression would be evaluated to false. Defines the hysteresis for analog triggers. Not applicable for digital triggers. Actions Action See Appendix C The action performed when the setpoint expression is evaluated to true (the setpoint is in operated state) Delays Operate delay 0.1-999.9 sec Release delay 0.1-999.
Chapter 4 PAS Application Software General Meter Setup examples of pulsed events are pulse inputs (transition pulses on the digital inputs), internal pulsed events (energy pulses and time interval pulses), and events generated by the interval timers. The logical controller automatically clears pulsed events at the end of each scan, so that triggers that used pulsed events are prevented from being triggered by the same event once again.
Chapter 4 PAS Application General Meter Setup Using Setpoint Events and Actions When a setpoint status changes, i.e., a setpoint event is either asserted or de-asserted, the following happens in your device: 1. 2. 3. The new setpoint status is logged to the setpoint status register that can be monitored through communications from the SCADA system or from a programmable controller in order to give an indication on the expected event.
Chapter 4 PAS Application Software Configuring Summary Energy and TOU Registers Configuring Summary Energy and TOU Registers The PM174 provides 8 total (summary) energy and 8 concurrent tariff energy and maximum demand registers that can be linked to any internal energy source or to any external pulse source that delivers pulses through the device digital inputs. The meter tariff structure supports 8 different tariffs using an arbitrary tariff schedule.
Chapter 4 PAS Application Configuring Summary Energy and The available options are shown in the following table: Parameter Options Default Description Summary/TOU Registers TOU Unchecked Checked Unchecked Links tariff registers to the selected energy source Use Profl Unchecked Checked Unchecked Enables automatic daily profiling for energy usage registers (both total and tariff registers if TOU is enabled) Dmd Profl Unchecked Checked Unchecked Enables automatic daily profiling for maximum dem
Chapter 4 PAS Application Software Configuring Summary Energy and TOU Registers The daily profile setup allows you to specify the daily tariff change points with a 15-minute resolution for 4 seasons using 4 different daily schedules for each season. To configure your daily profiles: 1. Select the desired season and day type. 2. Select the start time for each tariff change point and the corresponding active tariff number. 3. Repeat the setup for all active profiles.
Chapter 4 PAS Application Configuring Summary Energy and The meter’s TOU calendar allows you to configure any tariff schedule based on any possible utility regulation. The calendar itself has 32 entries that allow you to specify profiles for working days and holidays through all seasons in any order that is convenient for you, based on simple intuitive rules. There are no limitations on how to define your schedule.
Chapter 4 PAS Application Software Configuring Recorders Configuring Recorders The PM174 is provided with a 1-Megabyte onboard nonvolatile memory for data, event and waveform recording. Before using recorders, the device memory should be partitioned between log files. The device memory is fully configurable. You can define how much memory to allocate for each log file. If you want to change the factory settings, follow the guidelines in the section below.
Chapter 4 PAS Application Option Configuring Recorders Range Description Record size Parameters The size of the file record for a single channel or a section. Set automatically depending on the file and on the number of parameters in the data record 0-16 The number of parameters in a single data record Memory is allocated for each file statically when you set up your files and will not change unless you re-organize the files.
Chapter 4 PAS Application Software Configuring Recorders Configuring the Event Recorder To configure the Event log file: 1. Double click on the Event Log file partition with the left mouse button. 2. 3. Select a file type for your file. Select the maximum number of records you want to be recorded in the file. Click OK, then send your new setup to the meter or save to the device database. 4.
Chapter 4 PAS Application 2. 3. 4. 5. 6. 7. 92 Configuring Recorders Select a file type for your file. Select the number of parameters you want to be recorded in the file records. Select the maximum number of records you want to be recorded in the file. Click OK, and then send your new setup to the meter, or save to the device database.
Chapter 4 PAS Application Software Configuring Recorders automatically update the “Parameter” box with the following parameter in the group. Add the name for your data log file in the “Name” box. It will appear in the data log reports. Save your new setup to the device database, and send it to the meter. 8. 9. Factory Preset Periodic Data Logs Data logs #1 and #2 are factory preset for periodic recording of the standard power quantities as shown in the following table. No.
Chapter 4 PAS Application Configuring Recorders 3. 4. Select the TOU Daily profile file type. Select the number of season tariffs in your TOU schedule. Add one additional parameter if you selected to record the Summary (TOU total) registers as well. Select the maximum number of records you want to be recorded in the file assuming that a new record will be added once a day. Click OK and send your setup to the meter or save to the database. 5. 6.
Chapter 4 PAS Application Software Configuring Recorders 3. Select the maximum number of records you want to be recorded in the file.
Chapter 4 PAS Application 7. 8. 9. Configuring Recorders Select the number of cycles to be recorded prior to the event, and a total number of cycles in the waveform. Add the name for your waveform log file in the “Name” box. It will appear in the waveform reports. Save your waveform setup to the device database, and send it to the meter.
Chapter 4 PAS Application Software Configuring Recorders Event ID PQE212 PQE221 PQE222 PQE223 PQE231 PQE232 PQE233 PQE31 PQE32 PQE33 PQE4 IEEE 1159 category Trigger parameter Instantaneous swell Momentary interruption Momentary sag Momentary swell Temporary interruption Temporary sag Temporary swell Sustained interruption Undervoltage Overvoltage Voltage unbalance PQE52 PQE53 Harmonics THD Interharmonics THD PQE6 Voltage fluctuations (flicker) Frequency variations ½ cycle RMS voltage ½ cycle RMS v
Chapter 4 PAS Application 4. Configuring Communication Download your setup to the device. The available PQ recorder options are shown in the following table. The picture above shows the default PQ recorder settings. Option Range Default PQ Log Thresholds Threshold, % 0-200.0% Hysteresis, % 0-50.0% Waveform Log Checked On Start Unchecked Checked On End Unchecked 1-2 Log No. Description Defines the operating threshold for the PQ trigger in percent of the nominal (reference) value 5.
Chapter 4 PAS Application Software Configuring Communication Protocols Initially these registers are reserved and none of them points to an actual data register. To build your own Modbus register map: 1. 2. 3. 4. Select Protocol Setup from the Meter Setup menu and click on the Modbus Registers tab. Click on the Default button to cause the assignable registers to reference the actual default meter register 11776 (0 through 119 are not allowable register addresses for re-mapping).
Chapter 4 PAS Application Configuring Communication The following table describes available DNP options. Refer to the DNP3 Data Object Library document available from the DNP User’s Group on the DNP3 object types.
Chapter 4 PAS Application Software Configuring Communication Protocols Parameter Options Default Description Binary Counter Change Event Object 32-bit 32-bit 16-bit 16-bit 32-bit 32-bit 32-bit 16-bit 16-bit 16-bit 32-bit +Time The default BC change event object variation for requests with qualifier code 06 when no specific variation is requested 32-bit -Flag The default frozen BC object variation for requests with qualifier code 06 when no specific variation is requested Frozen Binary Counter Ob
Chapter 4 PAS Application Configuring Communication Configuring DNP Class 0 The most common method of getting static object information from the meter via DNP is to issue a read Class 0 request. The PM174 allows you to configure the Class 0 response by assigning ranges of points to be polled via Class 0 requests. To view or change the factory-set DNP Class 0 assignments, select Protocol Setup from the Meter Setup menu and click on the DNP Class 0 Points tab.
Chapter 4 PAS Application Software Configuring Communication Protocols A change event point index is normally the same as for the corresponding static object point. If you wish to use independent numeration for event points, enable re-mapping event point indices via DNP Options setup (see above) so they would start with index 0. You should define a separate event setpoint for each static object point you wish to be monitored for change events.
Chapter 4 PAS Application 4. 5. 6. Configuring Communication Less than (under) - a new event is generated when the point value drops below the specified threshold, and then when it returns above the threshold plus a predefined return hysteresis – applicable for AI objects. A hysteresis for the return threshold is 0.05 Hz for frequency and 2% of the operating threshold for all other points. Check the “Ev On” box for the points you wish to be included into event poll reports.
Chapter 4 PAS Application Software Remote Device Control Remote Device Control This section describes online operations on the meter you can perform through PAS. To access device control options you should have your device online. Remote Relay Control PAS allows you to send a command to any relay in your device or release a latched relay, except of the relays that are linked to an internal pulse source. These relays cannot be operated outside of the device.
Chapter 4 PAS Application 1. 2. Remote Device Control From the “Status” box, select the desired flag status. Click on Send. Device Diagnostics Device diagnostic messages may appear as a result of the PM174 built-in diagnostic tests performed during a start-up and device operation. To enter the Device Diagnostics dialog, check the On-line button on the PAS toolbar, select Device Control from the Monitor menu, and then click on the Device Diagnostics tab.
Chapter 4 PAS Application Software Remote Device Control which is not affected by loss of power and may be read and cleared via communications or via PAS. Refer to PM174 communication guides for the diagnostic register address and layout. See Device Diagnostic Codes in Appendix E for the list of diagnostic codes and their meanings. Frequent hardware failures may be the result of excessive electrical noise in the region of the device. If the meter continuously resets itself, contact your local distributor.
Chapter 4 PAS Application Remote Device Control To reset the desired accumulation registers or to clear a file: 1. 2. 3. 108 Click on the corresponding button, and then confirm your command. If a target has more than one component, you are allowed to select components to reset. Check the corresponding boxes, and then click OK.
Chapter 4 PAS Application Software Administration Administration Changing a Password PAS allows you to remotely change the password, and enable or disable the password security check in your meter. To change the password, check the On-line button, select Administration from the Monitor menu, and then select Change Password. To change the password: 1. Type in a new 4-digit password 2. Repeat the password in the Confirm box 3. Check the “Enable network protection” to enable password checking 4. Click Send.
Chapter 4 PAS Application 4. Point to the firmware upgrade file for your meter, click Open, and then confirm upgrading the meter. You are asked for the password regardless of the password protection setting in your meter. 5. Type the meter password, and click OK. If you did not change the password in the meter, enter the default password 0. 6. Wait until PAS completes upgrading your device. It takes about 3-4 minutes at 115,200 bps to download the file to the meter. 7.
Chapter 4 PAS Application Software Monitoring Devices Monitoring Devices Viewing Real-time Data For information on the real-time data monitoring and recording options, see the “PAS Getting Started Guide”. To get real-time data from your device, select the device site from the list box on the PAS toolbar, point to RT Data Monitor on the Monitor menu, and then select a data set you want to view.
Chapter 4 PAS Application Retrieving Log Files Viewing Real-time Waveforms To retrieve the real-time waveforms, select the device site from the list box on the PAS toolbar, and then select RT Waveform Monitor from the Monitor menu or click on the button on the PAS toolbar. Use the blue arrowhead buttons and for one-time or continuous waveform polling. Use the Stop button continuous polling. to stop See Viewing Waveforms for information on using waveform views.
Chapter 4 PAS Application Software Viewing Recorded Files PAS offers you different options for easy analysis of retrieved trend and fault data. Data can be displayed in primary or secondary units. You can use filtering to find out and work with a subset of events that meet the criteria you specify, or use sorting to rearrange records in the desired order. PAS can also link events and the corresponding data records and waveforms together to allow you more effective analysis of recorded events.
Chapter 4 PAS Application Viewing Recorded Files establishes links to retrieved waveforms and data records where it finds a relationship with the event. The event ID for which PAS finds related data is blue colored. To check a list of the event links, click on the colored event ID. Click on a list item to move to the waveform or data log record. Selecting Primary and Secondary Units Voltages and currents can be displayed in primary or secondary units.
Chapter 4 PAS Application Software Viewing Recorded Files To change the default sorting order based on the date and on the local toolbar or click time, click on the Sort button on the report window with the right mouse button and select “Sort...”, check the desired sort order, and then click OK. Selecting Voltage Units Voltages can be displayed in primary or secondary units.
Chapter 4 PAS Application Viewing Recorded Files formerly CBEMA) curve chart. To view an ITI curve chart, click on the “ITI” button on the window toolbar. To view the event details, click on the event point with the left mouse button. To directly move to the related power quality report entry or to a waveform record, click on the corresponding list item with the left mouse button.
Chapter 4 PAS Application Software Viewing Recorded Files The IEEE 1159 statistics is reported within the selected time range on either a weekly, or yearly basis. Statistics of power quality incidents for voltage imbalance, harmonics, flicker and frequency variations are given on a per week basis. The report shows the number of incidents, and a total time and a percentage of the observation time within which the characteristic exceeds the standard limit.
Chapter 4 PAS Application Viewing Recorded Files Customizing Reports If you wish to add a logo image, header and footer to your reports: 1. Select “Report Setup…” from the Reports menu, or click on the report window with the right mouse button, and then select “Report Setup…”. 2. Click on the Change button and select a logo image file. Check the “Show” box to include your logo into a report. Type the header text in the Page Header box. Check the “Enabled” box to include the header into a report.
Chapter 4 PAS Application Software Viewing Recorded Files Printing Reports To get a hardcopy of the report on the printer, select Print… from the File menu. To check the report, as it will look when printed, select Print Preview from the File menu. Viewing the Data Log Data log files can be displayed in a tabular view or in a graphical view as a data trend graph. Viewing Data Trend To view data in a graphical form, click on the Data Trend button on the local toolbar.
Chapter 4 PAS Application Viewing Recorded Files Click on the button on the local toolbar to view overlapped waveforms, and click on the button to view non-overlapped waveforms. Waveform data is recorded in series that may contain many cycles of the sampled waveform. A waveform window displays up to 128 waveform cycles.
Chapter 4 PAS Application Software Viewing Recorded Files cycles, the scroll bar appears under the waveform pane allowing you to scroll through the entire waveform. Selecting Waveform Channels To select the channels you want to view on the screen, click on the waveform window with the right mouse button, select “Channels...”, check the channels you want displayed, and then click OK. Using the Marker Lines Waveform and RMS panes have two blue dashed marker lines.
Chapter 4 PAS Application Viewing Recorded Files Viewing Symmetrical Components PAS can calculate the symmetrical components for voltages and currents at the point indicated by the left marker line. To enable or disable the symmetrical components, click on the waveform window with the right mouse button, select “Options...”, check or uncheck the “Symmetrical components” box on the “Channels” tab, and then click OK.
Chapter 4 PAS Application Software Viewing Recorded Files Viewing a Spectrum Chart Click on the button to view a spectrum chart. To change a channel, click on the window with the right mouse button, select “Channels...”, check the channel you want displayed, and then click OK. A spectrum is calculated over four cycles of the waveform beginning from the point where the left marker line is located. If there are more than one waveform views open, PAS gives the priority to the overlapped waveform view.
Chapter 4 PAS Application Viewing Recorded Files PAS can give you indication on whether harmonic levels in the sampled waveforms exceed compliance limits defined by the power quality standards or local regulations. To review or change harmonic limits: 1. Click on the spectrum window with the right mouse button and select “Limits…”. 2. Select a harmonics standard, or select “Custom” and specify your own harmonic limits.
Chapter 4 PAS Application Software Viewing Recorded Files Harmonics that exceed selected compliance levels are colored in red on the graph and in the tables. Viewing a Spectrum Table Click on the button on the local toolbar. The spectrum table displays voltage, current, active power and reactive power harmonic components both in percent of the fundamental and in natural units, and phase angles between the harmonic voltage and current.
Chapter 4 PAS Application Viewing Recorded Files coordinated waveforms that have the same time span as the selected waveform. 3. Check the sites your want to see displayed. 4. Click on the “Channels” button and select channels for each site. Click OK. 5. To change the channels, click on the waveform window with the right mouse button and select “Channels...”.
Chapter 4 PAS Application Software COMTRADE and PQDIF Converters COMTRADE and PQDIF Converters The COMTRADE and PQDIF file converters allow you to convert retrieved waveforms into COMTRADE or PQDIF file format, and data log tables – into PQDIF format. Manual Converting To manually convert your waveforms or a data log into COMTRADE or PQDIF format: 1. Click on the Export PAS toolbar. button on the 2. Select the database and a waveform or data log table you want to export, and then click Open. 3.
Chapter 4 PAS Application COMTRADE and PQDIF Converters PQDIF file names contain a site name followed by a timestamp of the first event recorded to the file, and may look like 12KVSUB_20040928T133038.pqd. Automatic Converting PAS allows you to automatically convert waveform and data logs into COMTRADE or PQDIF format at the time you upload data from your devices via the Upload Scheduler. To automatically convert your waveform or data log tables into COMTRADE or PQDIF format: 1.
Appendix A Technical Specifications Appendix A Technical Specifications Environmental Conditions Operating temperature: -20°C to 60°C (-4°F to 140°F) Storage temperature: -25°C to 80°C (-13°F to 176°F) Humidity: 0 to 95% non-condensing Construction Dimensions see Figure 2-1 Weight: 1.23kg (2.7 lb.) Materials Case enclosure: plastic PC/ABS blend Display body: plastic PC/ABS blend Front panel: plastic PC PCB.: FR4 (UL94-V0) Terminals: PBT (UL94-V0) Connectors-Plug-in type: Polyamide PA6.
Appendix A Technical Specifications 5A secondary Operating range: continuous 10A RMS Burden: < 0.1 VA Overload withstand: 15A RMS continuous, 300A RMS for 1 second 1A secondary Operating range: continuous 2A RMS Burden: < 0.02 VA Overload withstand: 6A RMS continuous, 80A RMS for 1 second Relay Outputs 2 relays rated at 3A/250 VAC; 3A/30 VDC, 2 contacts (SPST Form A) 2 Wire size: 14 AWG (up to 1.
Appendix A Technical Specifications Update time: 1 cycle Communication Ports COM1 (Optional modules) Serial EIA RS-232 optically isolated port Isolation: 2,000 V RMS Connector type: DB9 female. Baud rate: up to 115.2 kbps. Supported protocols: Modbus RTU, Modbus ASCII and DNP3. RS-422/RS-485 optically isolated port Isolation: 2,000 V RMS Connector type: DB9 female. Baud rate: up to 115.2 kbps. Supported protocols: Modbus RTU, Modbus ASCII and DNP3.
Appendix A Technical Specifications Keypad: 6 push buttons Communication: EIA RS-485 port with 12V supply voltage Connector type: DB15, 15 pins 2 Wires size: up to 14 AWG (up to 1.5 mm ) Distance: up to 1000 m (3200 feet) Standards Compliance Accuracy per ANSI C12.20 –1998 UL File no.
Appendix A Technical Specifications Measurement Specifications Parameter Voltage Full Scale @ Input Range 120VxPT @ 120V Accuracy % % FS Reading 0.2 0.01 Range Conditions 10% to 120% FS 400VxPT @ 690V 0 to 1,150,000 V Starting voltage 1.5% FS @ 120V 1.5% FS @ 690V Line current CT 0.2 0.02 1% - 200% FS 0 to 40,000 A Active power 0.36×PT×CT @ 120V 0.2 0.02 |PF| ≥ 0.5 1 -10,000,000 kW to +10,000,000 kW 0.3 0.04 |PF| ≤ 0.9 1 -10,000,000 kvar to +10,000,000 kvar 0.2 0.02 |PF| ≥ 0.
Appendix B Parameters for Analog Output Appendix B Parameters for Analog Output The following table lists parameters that can be provided on the meter’s analog outputs and on the external analog expander outputs. Display Code nonE Designation NONE rt.U1 rt.U2 rt.U3 rt.U12 rt.U23 rt.U31 rt.C1 rt.C2 rt.C3 V1/12 RT 1 V2/23 RT 1 V3/31 RT 1 V12 RT V23 RT V31 RT I1 RT I2 RT I3 RT rt.Ac.P rt.rE.P rt.AP.P rt.PF rt.PF.LG rt.PF.Ld rt.U.AG rt.UL.AG rt.C.
Appendix C Setpoint Triggers and Actions Appendix C Setpoint Triggers and Actions Setpoint Triggers Display Code Designation Description nonE NONE St1.On St2.On St1.OFF St2.OFF STAT STAT STAT STAT rL1.On rL2.On rL1.OFF rL2.OFF RELAY RELAY RELAY RELAY POS.Ph.r nEG.Ph.r Pq.E POS PHASE REVERSAL NEG PHASE REVERSAL PQ EVENT A.In1.Hi A.In2.Hi A.In1.Lo A.In1.Lo HI AI1 HI AI2 LO AI1 LO AI2 rtHi.C1 rtHi.C2 rtHi.C3 rtLo.C1 rtLo.C2 rtLo.C3 HI I1 RT HI I2 RT HI I3 RT LO I1 RT LO I2 RT LO I3 RT rtHi.
Appendix C Setpoint Triggers and Actions 136 Display Code Designation Description ArHi. S ArPF.LG ArPF.Ld HI kVA AVR HI PF LAG AVR HI PF LEAD AVR ArnEU.C ArHi.Fr ArLo.Fr ArHU.Ub ArHC.Ub HI In AVR HI FREQ RT LO FREQ RT HI V UNB% RT HI I UNB% RT Hi d.U1 Hi d.U2 Hi d.U3 Hi d.C1 Hi d.C2 Hi d.C3 Hi d.P.i Hi d.P.E Hi d.q.i Hi d.q.i Hi d. S HiSd.P.i HiSd.P.E HiSd.q.i HiSd.q.i HiSd. S HiAd.P.i HiAd.P.E HiAd.q.i HiAd.q.i HiAd. S HiPd.P.i HiPd.P.E HiPd.q.i HiPd.q.i HiPd.
Appendix C Setpoint Triggers and Actions Display Code Designation Description YEAr Mon M.dAY hour Min SEc YEAR MONTH DAY OF MONTH HOURS MINUTES SECONDS FG1.On … FG8.OFF EVENT FLAG 1 ON … EVENT FLAG 8 OFF SP1.On … SP16.
Appendix C Setpoint Triggers and Actions 138 Display Code Designation Description FLG3.On FLG4.On FLG1.OFF FLG2.OFF FLG3.OFF FLG4.
Appendix D Parameters for Monitoring and Appendix D Parameters for Monitoring and Data Logging The following table lists parameters measured by the meter that are available for data logging and monitoring through communications. The left column shows data abbreviations used in PAS. Parameter groups are highlighted in bold.
Appendix D Parameters for Monitoring and Designation V1 THD V2 THD V3 THD I1 THD I2 THD I3 THD I1 KF I2 KF I3 KF I1 TDD I2 TDD I3 TDD V12 V23 V31 RT TOTAL kW kvar kVA PF PF LAG PF LEAD kW IMP kW EXP kvar IMP kvar EXP V AVG V LL AVG I AVG RT AUX In FREQ V UNB% I UNB% AVR PHASE V1 V2 V3 I1 I2 I3 kW L1 kW L2 kW L3 kvar L1 kvar L2 kvar L3 kVA L1 kVA L2 kVA L3 PF L1 PF L2 PF L3 V1 THD V2 THD V3 THD I1 THD I2 THD I3 THD I1 KF I2 KF I3 KF I1 TDD I2 TDD 140 Description V1/V12 Voltage THD 2 V2/V23 Voltage THD 2 V3
Appendix D Parameters for Monitoring and Designation I3 TDD V12 V23 V31 AVR TOTAL kW kvar kVA PF PF LAG PF LEAD kW IMP kW EXP kvar IMP kvar EXP V AVG V LL AVG I AVG AVR AUX In FREQ V UNB% I UNB% RMS (10-min) V1 V2 V3 V ZERO-SEQ I ZERO-SEQ V UNB% I UNB% HRM TOT (10-min) V1 THD V2 THD V3 THD I1 THD I2 THD I3 THD V1 THD/I V2 THD/I V3 THD/I I1 TDD I2 TDD I3 TDD PHASORS V1 Mag V2 Mag V3 Mag I1 Mag I2 Mag I3 Mag V1 Ang V2 Ang V3 Ang I1 Ang I2 Ang I3 Ang DEMANDS V1 DMD V2 DMD V3 DMD I1 DMD I2 DMD Series PM174 Pow
Appendix D Parameters for Monitoring and Designation I3 DMD kW IMP BD kvar IMP BD kVA BD kW IMP SD kvar IMP SD kVA SD kW IMP ACC DMD kvar IMP ACC DMD kVA ACC DMD kW IMP PRD DMD kvar IMP PRD DMD kVA PRD DMD PF IMP@kVA MXDMD kW EXP BD kvar EXP BD kW EXP SD kvar EXP SD kW EXP ACC DMD kvar EXP ACC DMD kW EXP PRD DMD kvar EXP PRD DMD HRM DMD V1 THD DMD V2 THD DMD V3 THD DMD I1 THD DMD I2 THD DMD I3 THD DMD I1 TDD DMD I2 TDD DMD I3 TDD DMD SUMM ACC DMD SUM REG1 ACC DMD SUM REG2 ACC DMD … SUM REG8 ACC DMD SUMM BLK
Appendix D Parameters for Monitoring and Designation kVAh L2 kVAh L3 %HD V1 V1 %HD01 V1 %HD02 … V1 %HD50 %HD V2 V2 %HD01 V2 %HD02 … V2 %HD50 %HD V3 V3 %HD01 V3 %HD02 … V3 %HD50 %HD I1 I1 %HD01 I1 %HD02 … I1 %HD50 %HD I2 I2 %HD01 I2 %HD02 … I2 %HD50 %HD I3 I3 %HD01 I3 %HD02 … I3 %HD50 ANG V1 V1 H01 ANG V1 H02 ANG … V1 H50 ANG ANG V2 V2 H01 ANG V2 H02 ANG … V2 H50 ANG ANG V3 V3 H01 ANG V3 H02 ANG … V3 H50 ANG ANG I1 I1 H01 ANG I1 H02 ANG … I1 H50 ANG ANG I2 I2 H01 ANG I2 H02 ANG … I2 H50 ANG ANG I3 I3 H01 ANG
Appendix D Parameters for Monitoring and Designation V3 H01 I1 H01 I2 H01 I3 H01 kW L1 H01 kW L2 H01 kW L3 H01 kvar L1 H01 kvar L2 H01 kvar L3 H01 kVA L1 H01 kVA L2 H01 kVA L3 H01 PF L1 H01 PF L2 H01 PF L3 H01 HRM TOT POW kW H01 kvar H01 kVA H01 PF H01 FLICKER V1 Pst V2 Pst V3 Pst V1 Plt V2 Plt V3 Plt MIN PHASE V1 MIN V2 MIN V3 MIN I1 MIN I2 MIN I3 MIN V1 THD MIN V2 THD MIN V3 THD MIN I1 THD MIN I2 THD MIN I3 THD MIN I1 KF MIN I2 KF MIN I3 KF MIN I1 TDD MIN I2 TDD MIN I3 TDD MIN MIN TOTAL kW MIN kvar MIN kV
Appendix D Parameters for Monitoring and Designation V3 THD MAX I1 THD MAX I2 THD MAX I3 THD MAX I1 KF MAX I2 KF MAX I3 KF MAX I1 TDD MAX I2 TDD MAX I3 TDD MAX MAX TOTAL kW MAX kvar MAX kVA MAX PF MAX MAX AUX In MAX FREQ MAX MAX DMD V1 DMD MAX V2 DMD MAX V3 DMD MAX I1 DMD MAX I2 DMD MAX I3 DMD MAX kW IMP SD MAX kW EXP SD MAX kvar IMP SD MAX kvar EXP SD MAX kVA SD MAX MAX HRM DMD V1 THD DMD MAX V2 THD DMD MAX V3 THD DMD MAX I1 THD DMD MAX I2 THD DMD MAX I3 THD DMD MAX I1 TDD DMD MAX I2 TDD DMD MAX I3 TDD DMD
Appendix D Parameters for Monitoring and Designation … TOU REG2 TRF8 TOU REG3 TOU REG3 TRF1 TOU REG3 TRF2 … TOU REG3 TRF8 TOU REG4 TOU REG4 TRF1 TOU REG4 TRF2 … TOU REG4 TRF8 TOU REG5 TOU REG5 TRF1 TOU REG5 TRF2 … TOU REG5 TRF8 TOU REG6 TOU REG6 TRF1 TOU REG6 TRF2 … TOU REG6 TRF8 TOU REG7 TOU REG7 TRF1 TOU REG7 TRF2 … TOU REG7 TRF8 TOU REG8 TOU REG8 TRF1 TOU REG8 TRF2 … TOU REG8 TRF8 TOU MAX DMD REG1 DMD1 TRF1 MAX DMD1 TRF2 MAX … DMD1 TRF8 MAX TOU MAX DMD REG2 DMD2 TRF1 MAX DMD2 TRF2 MAX … DMD2 TRF8 MAX TOU
Appendix D Parameters for Monitoring and Designation DMD7 TRF8 MAX TOU MAX DMD REG8 DMD8 TRF1 MAX DMD8 TRF2 MAX … DMD8 TRF8 MAX Description Tariff #8 register TOU Maximum Demand Register #8 Tariff #1 register Tariff #2 register … Tariff #8 register 1 In 4LN3, 4LL3, 3LN3, 3LL3, 3BLN3 and 3BLL3 wiring modes, the voltages will be line-to-neutral; for any other wiring mode, they will be line-to-line voltages.
Appendix E Data Scales Appendix E Data Scales The maximum values for volts, amps and power in the PM174 setup and in communications are limited by the voltage and current scale settings. See Device Options in Chapter 4 on how to change the voltage scale in your meter. The following table defines the meter data scales.
Appendix F Device Diagnostic Codes Appendix F Device Diagnostic Codes Diagnostic Code 2 3 5 6 7 Description Reason Hardware Hardware Hardware Hardware Hardware 8 9 Memory/Data fault Hardware watchdog reset CPU exception Run-time software error Software watchdog timeout Power Down/Up Warm restart 10 Configuration reset 11 13 RTC fault Low battery 15 EEPROM fault failure failure failure failure failure Normal power-up sequence External restart via communications or by firmware upgrade Corrupted