3700R/3740 Refrigerated Sampler Installation and Operation Guide Part #60-3703-176 of Assembly #60-3704-105 Copyright © 1996. All rights reserved, Teledyne Isco, Inc.
Foreword This instruction manual is designed to help you gain a thorough understanding of the operation of the equipment. Teledyne Isco recommends that you read this manual completely before placing the equipment in service. Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction. If the problem persists, call or e-mail the Teledyne Isco Technical Service Department for assistance.
3700R/3740 Refrigerated Sampler Safety 3700R/3740 Refrigerated Sampler Safety General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed in the following general warnings: WARNING This instrument has not been certified for use in “hazardous locations” as defined by the National Electrical Code.
3700R/3740 Refrigerated Sampler Safety Hazard Symbols The equipment and this manual use symbols used to warn of hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual. The lightning flash and arrowhead within the triangle is a warning sign alerting you of “dangerous voltage” inside the product. Pinch point.
3700R/3740 Refrigerated Sampler Table of Contents Section 1 Introduction 1.1 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1.1 3700 Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1.2 3740 Refrigerator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 Programmable Features . . . . . . . . . . . . . . . . . . . . . . . . . .
3700R/3740 Refrigerated Sampler Table of Contents 3.1.1 Sample Events and the Sampling Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 3.2 Types of Samples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 3.2.1 Types of Sample Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 3.2.2 Types of Sample Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 3.
3700R/3740 Refrigerated Sampler Table of Contents 3.14 Foreign Languages and Metric Units of Measure . . . . . . . . . . . . . . . . . . . . . . . . 3-81 3.15 Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-82 3.16 Standby State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-98 3.16.1 Operable Keys in Standby State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-99 3.
3700R/3740 Refrigerated Sampler Table of Contents 5.21.1 Peristaltic Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23 5.21.2 Volumetric Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23 5.21.3 Sample Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23 Appendix A Replacement Parts Lists Appendix B Accessories B.1 B.2 B.3 B.4 B.5 B.6 General Accessories . . . . . . .
3700R/3740 Refrigerated Sampler Table of Contents 3-12 Extended Programming Mode: STORM Programming . . . . . . . . . . . . . . . . . . . 3-62 3-13 Simplified Start Time Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-64 3-14 Start Time Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-66 3-15 Start Time Diagram for Nonuniform Time Routines . . . . . . . . . . . . . . . . . . . .
3700R/3740 Refrigerated Sampler Table of Contents Flow-Paced Sampling / Time-Based Bottle Sets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-88 Time-Paced Samples at Nonuniform Intervals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-91 Time-Paced Samples at Nonuniform Clock Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-93 Sampling with Stops and Resumes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-94 Storm Sampling . . .
3700R/3740 Refrigerated Sampler Section 1 Introduction This section contains an overall description of the sampler and a list of technical specifications. 1.1 Description The 3700R/3740 Refrigerated Sampler is a programmable liquid sampler designed for sequential sampling. It is one of Teledyne Isco’s 3700 Series of portable and refrigerated samplers.
3700R/3740 Refrigerated Sampler Section 1 Introduction arm to the next bottle. A desiccator is installed in the control box to prevent moisture damage to the electronics, pump, and distributor systems. 1.1.2 3740 Refrigerator If you want to convert a presently owned 3700 Portable Sampler into a refrigerated unit, or to have both portable and refrigerated options with the same controller, the 3740 Sampler Refrigerator is available.
3700R/3740 Refrigerated Sampler Section 1 Introduction 1.2.1 Flexible Sampling Intervals The 3700R is designed to collect up to 24 separate sequential (discrete) samples and is fully programmable for true composite sampling. Both sequential and composite samples may be collected at user-definable time intervals (time pacing) or at equal flow volume intervals using flow pulse inputs from an external flow meter (flow pacing). The flow interval is set from 1 to 9999 flow pulses.
3700R/3740 Refrigerated Sampler Section 1 Introduction where you need a series of identical sets of samples. Multiple bottle compositing can be used when you need to acquire a series of larger volume composite samples by taking a number of smaller samples and distributing them over several bottles. In the extended programming mode, you can switch multiplexed bottles or bottle sets after a programmed number of samples have been deposited or after a programmed period of time has elapsed.
3700R/3740 Refrigerated Sampler Section 1 Introduction current date. The start time delay is the period between the time you press the start sampling key and the time the routine actually starts. It is adjustable from 0 (zero) to 9999 minutes. Other features are available. Program storage allows you to store up to three separate programs, eliminating the need to reprogram the sampler for recurrent sampling routines. A program lock prevents unauthorized program alterations.
3700R/3740 Refrigerated Sampler Section 1 Introduction matic rinsing of the suction line when concerns of cross contamination arise. A programmable setting for sampling retries is available. If the suction line becomes clogged and no liquid is detected in the line, the sampler can be programmed to repeat a purge cycle – up to three times – to clear the clogged line. 1.4.2 Pump Tubing and Suction Lines The pump tubing is Silastic™ medical grade silicon rubber.
3700R/3740 Refrigerated Sampler Section 1 Introduction • 7500 ml (2 gallon) polyethylene (rectangular) bottles with unlined caps. Single Bottle Composite Configurations • 9400 ml (2.5 gallon) glass (round) bottle with Teflon lined cap. • 9400 ml (2.5 gallon) polyethylene (round) bottle with unlined cap. • With minor changes to the refrigerator, you can install a larger composite sample bottle (4 or 5.5 gallon round polyethylene). Call the factory. Note The 12-bottle configuration (2.
3700R/3740 Refrigerated Sampler Section 1 Introduction Set points or pairs of set points — pairs can be level and rainfall rates, level and elapsed time, rainfall and elapsed time, and so on — form the sampler enable control condition. A control condition is simply the set of parameters defining the conditions in which a flow meter will enable the sampler. For example, a flow meter can be programmed with a control condition which is satisfied when the flow meter detects 1/4 inch of rainfall in 15 minutes.
3700R/3740 Refrigerated Sampler Section 1 Introduction 1.7.4 Liquid Level Actuator 1.8 Technical Specifications, Controls, and Connectors The Liquid Level Actuator is used to provide level-sensitive control of the sampler. The actuator can be used as a simplified alternative to a flow meter. The technical specifications, controls, and connectors of the 3700R/3740 are listed in Table1-1. Refer to Figure 2-6 for a picture of the controls and connectors.
3700R/3740 Refrigerated Sampler Section 1 Introduction Table 1-1 3700R Technical Specifications, Controls, and Connectors (Continued) Clock Specifications Real Time Clock Accuracy 1 minute/month, typical. Sample Specifications Sample Volume Accuracy With the liquid detector enabled and automatic compensation for head: typically, the greater of ± 10% or ± 20 ml, over a head range of 1 to 12 feet and sampler supply voltage of 10 to 13 volts. Sample Volume Repeatability ± 10 ml, typical.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures 2.1 Summary of Setup Procedure The following sections detail the preparations made before using the refrigerator: 1. Install the sampler. The suction line can extend to a liquid source up to 99 feet from the sampler, note that the maximum lift for the peristaltic pump with either the Teflon or vinyl line is 26 feet.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures Cut the line to the shortest length feasible: this aids the downhill routing. Avoid loops of coiled suction line which may hold residual amounts of liquid which would cross contaminate sample volumes. A shorter line will also extend battery life and pump tube life because a shorter pumping cycle is used to deliver the sample volume. 2.2.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures 2.2.2 Attaching the Teflon Suction The Teflon line is attached to the pump tubing by inserting the line into the pump tubing and securing it with a suitable clamp. 2.2.3 Strainers The 3/8 inch ID vinyl suction lines are shipped from the factory with our standard weighted polypropylene strainer (Figure 2-3) installed on one end of the suction line and a tubing coupling on the other end.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures 2.2.4 Intake Placement The proper placement of the sampler intake assures the collection of representative samples. The intake should be placed in the main flow, not in an eddy or at the edge of flow. The vertical position of the intake in the flow is important. An intake at the bottom may result in excess heavy solids and no floating materials, while placement at the top may result in the opposite. 2.2.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures b a t t e r y c o n n e c to r t o t h e s am p l e r ’s p o w e r c o n n e c t o r. A rechargeable nickel-cadmium battery is most commonly used; however, lead-acid batteries are available. The Teledyne Isco Nickel-Cadmium battery has an operating capacity of seven standard sampling programs after an 18 hour charge.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures Flow Meter Cable Power Cable Flow Meter Cable Port Thermostat Knob Flow Meter Connector Suction Line Port Control Base Figure 2-6 Flow Meter Connection and Suction Line Port Note Install the bottle rack so that bottle number 1 is oriented towards the front of the refrigerator. 2.5.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures bottles and fits down inside the circle of bottles. It is held in place by three large draw cords attached to the bottom of the rack. Each of these draw cords is stretched up and hooked onto a corresponding projection located on the ring. The sets of glass and plastic bottles are interchangeable in the sample bottle rack.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures in the sample bottle rack. Combinations of glass and plastic bottles may be used if desired. Note: All 8 bottles should be installed in the rack to ensure that they are properly held in place. The order in which the bottles are filled is indicated on the plate located on the bottom of the rack. Since the bottle rack is used for both 24 and 8 bottle configurations, the plate is numbered 1 through 24 and 1 through 8.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures Liquid Deflector Figure 2-8 Liquid Deflector Figure 2-9 Sample Bottle Rack with 24 Plastic Sample Bottles 2-9
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures Bottle #1 Figure 2-10 Sample Bottle Rack with 24 Glass Sample Bottles Figure 2-11 Sample Bottle Rack with 8 Glass Sample Bottles 2-10 Front ↓
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures Figure 2-12 Sample Bottle Rack with 8 Plastic Sample Bottles 2.5.4 Bottle Locating Base with 2 Sample Bottles The locating base can be permanently installed over the wire rack mounting plate. It is held in place by five screws, washers, and nuts and a bottle locating base retainer. Four of the screws, washers, and nuts replace the four locating posts on the wire rack mounting plate.
3700R/3740 Refrigerated Sampler Section 2 Setup and Operating Procedures 2.5.6 Single Bottle Composite Sampling Two and one-half (2.5) gallon glass and plastic containers are available from Teledyne Isco for composite sampling. Each container lid has a hole for routing the pump tube into the container to keep the pump tube from vibrating. When using the single bottle, remove the distributor arm and bottle rack.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines This section discusses in detail the sampling features of the 3700R Sampler and the procedures for programming your sampler. Teledyne Isco has found that the most effective way to learn to program the sampler is to read this section with a sampler in front of you. The time that you spend in the office or laboratory actually programming the sampler and observing its operation is usually time well spent.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines • The Run State section discusses what is occurring when your sampler is operating in the run state. 3.1 Description of Sampling Operation 3.1.1 Sample Events and the Sampling Cycle This information introduces the types of sampling performed by the 3700R and the terminology that describes those capabilities. A sample event is the complete cycle involved in taking a sample.
00R/3740 Refrigerated Sampler Section 3 Programming Guidelines You can characterize sample types by sample distribution and sample pacing. Sample distribution is the order in which samples are placed in the bottles. Sample pacing is the interval between samples. 3.2.1 Types of Sample Pacing There are two types of sample pacing: time-pacing and flow-pacing. In time-paced sampling, the interval between samples is a time interval.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines You can combine the types of sample pacing and sample distribution. Thus, you can take time- or flow-paced sequential samples and time- or flow-paced composite samples. Each combination of pacing is noted below. Time-Paced Sequential – In a time-paced sequential sample, samples are collected at equal increments of time. Each bottle receives one sample.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Bottle Number (8 of 24 Bottles) 3 4 5 6 1 2 2 4 6 8 1 3 5 7 7 8 Sample Event Figure 3-2 Samples-Per-Bottle Multiplexing Bottles-Per-Sample Multiplexing – I n b o t t l e s - p e r- s a m p l e multiplexing, you fill a programmed number of bottles at each sample event. (The programmed number of bottles is a bottle set; as many as 24 bottles can be used.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.3 Types of Sampling Available in Extended Programming Mode The extended programming mode expands the variations of sample pacing and distribution. The extended features listed below are used in more complex sampling routines. Note, however, that the sampling capabilities described in the Section 3.2 are available in both the basic and the extended programming modes.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines number of samples. This allows you to control the number of samples in each bottle. When you switch bottles or sets on a time basis, the total volume of each bottle or bottle set may vary in samples-per-bottle multiplexing. 3.3.3 Multiple-Bottle Compositing The extended programming mode offers a third multiplexing possibility: multiple-bottle compositing.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.4.1 Storm-Event Sample Distribution Schemes A bottle group can consist of one or more bottles. Thus, the first-flush sample can be a large sample delivered to a single bottle or distributed to several bottles with a multiplexing scheme.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Standby State – In the standby state, the sampler is waiting for your instructions. From standby, you can start a sampling routine, placing the sampler in the run state. You can also access the interactive state, which is discussed in an upcoming section. Run State – In the run state, the sampler is executing the sampling routine and operating under program control.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Extended Programming – The extended programming mode increases the number of available features, making more complex sampling routines possible. For example, you can use nonuniform time intervals, and create an intermittent sampling schedule with the Stops and Resumes feature.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines . . . STANDBY . . .
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-1 Configure Option Functions (Continued) Configure Option Programming Mode Function Load Stored Extended Loads one of as many as three previously saved sampling programs. Extended Saves current sampling program. Extended Directs sampler to take a sample at the beginning of a flow-paced program. Directs sampler to sample at time-switches.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Most configure options are available in both programming modes. Table 3-1 summarizes the configure sequence options; each option is marked as available in both modes or available only in the extended mode. A detailed discussion of each configuration option is found in the sections beginning with Set Clock through Exit Configuration.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines dure, see the section Basic Programming Procedure. The examples given demonstrate several types of sampling programs in the basic mode. Examples of extended mode sampling programs are also shown. 4. If desired, connect the sampler to a flow meter or other interfacing equipment. Start the sampler. 5. As the routine is executed, the displays will report the sampler’s progress.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Figure 3-7 3700 Control Panel Control Keys – Control keys let you turn the sampler on or off, start or resume the currently entered program, and control the sampler manually. Control key functions are: • ON/OFF - Pressing the ON/OFF key when the sampler is off will turn the sampler on, reactivate the display, and place the sampler in the standby state.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines • When you press the STOP key, the sampler will display a message similar to the one shown below. This message indicates the total number of pump counts that occurred while the pump was running. This information can be used to set the desired number of purge counts required to purge the suction line. See the Purge Counts configure option. Press any key, except STOP and ON/OFF, to return to the standby display..
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines • EXIT PROGRAM - Pressing the EXIT PROGRAM key while in the program sequence will return the sampler to standby. If you press the EXIT PROGRAM key while in the run state, it will halt the program; the message “PROGRAM HALTED” will be displayed. • CLEAR ENTRY - When entering a number, use the CLEAR ENTRY key to return to the original entry.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines “PROGRAM” and the sample volume entry “250” will be blinking. Italic characters are used to represent blinking words or numbers in illustrations in this manual. [PROGRAM, CONFIGURE] SAMPLER SAMPLES VOLUMES OF ml (10 - 1000) 250 Display Numbers – Nearly all input displays have a number assigned to them. The number is used to cross-reference the input displays with an explanatory listing found in Appendix C.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines key. The CLEAR ENTRY key will clear any typed number and the original number will reappear. The left arrow will erase the most recently typed number. Some numeric input displays prompt you to enter more than one value. In the illustration below, the prompt asks for five numeric entries: hours, minutes, day, month, and year. You can use the left and right arrow keys to move back and forth between each of the five entries.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines programming mode procedure and examples before using the extended programming mode. Most of the procedures used in the extended programming mode duplicate those of the basic mode and are not repeated in the extended programming procedures section. Flow charts accompany examples for both programming modes. These charts diagram the program sequence structure for each mode. Figure 3-8 charts the Basic Programming Mode structure.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines . . . STANDBY . . .
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Note If you will be using very short sample intervals, be sure the interval is longer than the duration of the sampling cycle. For truly representative sampling, the interval in either time- or flow-paced sampling should be longer than the duration of the sampling cycle. If the intervals are too short, you will not miss any sample events, although some events will occur at improper times.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines If you are using bottles-per-sample multiplexing to add preservatives to consecutive samples, be sure to consider the volume of the preservative into account. Suction Head- You can configure the sampler, through the Liquid Detector configure option, to add the suction-head setting to the program sequence. The suction-head display will appear after you enter the sample volume.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Before programming the sampler – E s p e c i a l l y i f y o u a r e unfamiliar with the settings used in the previous routine, or if the settings have been changed, verify the configure-option settings. You must change the settings if they do not match your bottle size or the suction line used with your unit.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Checking the Configure Option Settings (Continued) BOTTLE VOLUME IS 1000 ml 9. Select “YES” if you want to use the nonstandard bottle volume. Select “NO” if you want to revise the entry. The display shown in step 8 will reappear; use it to enter the revised bottle volume. Press the ENTER/PROGRAM key to accept the entry and advance to step 10 1001 ml! SURE? . . . ARE YOU {YES, NO} 10.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Checking the Configure Option Settings (Continued) RETRY UP TO 0 TIMES WHEN SAMPLING (0-3) 19. To verify the programming mode setting, press the ENTER/PROGRAM key. SELECT OPTION: (← →) PROGRAMMING MODE 20. Select “BASIC.” Press the ENTER/PROGRAM key to accept the selection [BASIC, EXTENDED] PROGRAMMING MODE 21. Scroll through the remaining options with the arrow keys. Use the techniques above to change or verify the settings.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines The steps in Example Time-Paced Sequential Sampling program the sampler to take samples every 30 minutes. Each bottle is to receive one 250 ml sample volume. The sampling routine is to begin according to the start time delay: one minute after you press the START SAMPLING key. Example: Time-Paced Sequential Sampling 1. If the sampler is not already on, press the ON/OFF key to turn it on. The standby display shown here will appear.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Sequential Sampling (Continued) This display appears as the sampler counts down the time remaining to the start time. The first line reports the bottle which will receive the next sample volume. The second line reports the start time at the left and the current time on the right. BOTTLE 1 AT 5:44 5:42:33 When the start time arrives, the sampler will take the first sample.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Flow-Paced Sequential Sampling (Continued) 5 SAMPLE EVERY PULSES (1-9999) 5. Select “NO.” MULTIPLEX SAMPLES? [YES, NO] 6. Enter “500” to set the sample volume at 500 ml. SAMPLE VOLUMES OF (10 - 1000) 500 ml 7. Select “NO.” ENTER START TIME? [YES, NO ] After this message displays briefly, the sampler automatically returns to the standby state. PROGRAMMING SEQUENCE COMPLETE 8.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Flow-Paced Sequential Sampling (Continued) ↓ ↓ This display informs you that the sampling routine is completed. The number of sample events displays in the upper right corner. The second line indicates the current time and date. DONE . . . 24 SAMPLES 22:48:32 20-JUN-04 Example Checking the Configure Option Settings shows samples-per-bottle multiplexed sampling. Samples are taken every 15 minutes.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Multiplexed (samples-per-bottle) Sampling ENTER START TIME? [YES, NO] 11. When this display first appears, it will show you the sampler’s “guess” at the start time. Type in the desired start time. For this example, the start time is 6:00 on 6/20/91. TAKE FIRST SAMPLE AT 6:00 20-JUN-04 After this message is displayed briefly, the sampler automatically returns to the standby state. PROGRAMMING SEQUENCE COMPLETE . . . 12.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Multiplexed (samples-per-bottle) Sampling 3 OF 4, BOTTLE 1 PUMPING 200 ml ↓ ↓ When the final sample has been deposited in the first bottle, the display changes to indicate the new sample bottle.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Flow-Paced Multiplexed (bottles-per-sample) Sampling MULTIPLEX SAMPLES? [YES, NO] 6. Select “BOTTLES PER SAMPLE.” [BOTTLES PER SAMPLE, SAMPLES PER BOTTLE] 7. Enter the number of bottles per sample event: “2.” 2 BOTTLES PER SAMPLE EVENT (1-24) 8. Enter “500” to set the sample volume at 500 ml. SAMPLE VOLUMES OF (10 - 1000) 500 ml 9. Select “YES.” ENTER START TIME? [YES, NO] Enter the start time for the sampling routine.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Flow-Paced Multiplexed (bottles-per-sample) Sampling When the sampler receives the tenth pulse, the first line reports the current bottle number. BOTTLE 1 When the pump begins to run forward to deliver the sample volume, the second line appears. After the pump completes the sample delivery, it reverses to purge the liquid from the pump tube past the liquid detector.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines capacity of the 2.5 gallon bottle. The total sample volume collected at the end of the routine would be 12,000 ml (48 samples x 250 ml/sample), well within the capacity of the 4-gallon bottle. Example: Time-Paced Composite Sampling 1. If the sampler is not already on, press the ON/OFF key to turn it on. Press ENTER/PROGRAM to access the interactive state. . . . STANDBY . . . 10:34:50 19-JUN-04 2.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Composite Sampling (Continued) ↓ ↓ 13. Press the ENTER/PROGRAM key to reenter the interactive state. . . . STANDBY . . . 10:38:50 19-JUN-04 14. Access the program sequence by selecting “PROGRAM.” [PROGRAM, CONFIGURE] SAMPLER 15. Select “TIME.” [TIME, FLOW] PACED SAMPLING 16. Enter “0” to set the hours at zero. Press ENTER/PROGRAM to store the number “0” and move to the minutes entry. SAMPLE EVERY 0 HOURS,1 MINUTES 17.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Composite Sampling (Continued) When the pump reverses for the pre-sample purge, the second line disappears. At the end of the pre-sample purge, the pump runs forward to deliver the sample, and the second line appears on the display. This display remains through the end of the post-sample purge.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Calibration 1. Press ENTER/PROGRAM to access the interactive state. . . . STANDBY . . . 9:34:50 19-JUN-04 2. Access the program sequence by selecting “PROGRAM.” [PROGRAM, CONFIGURE] SAMPLER 3. Step through the program until the “CALIBRATE SAMPLER?” input display appears (Step 4). [TIME, FLOW] PACED SAMPLING Other program sequence displays. ↓ ↓ 4. Select “YES.” CALIBRATE SAMPLER? [YES, NO] 5.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines ENTER/PROGRAM key while the desired option name is displayed. Once you’ve accessed the input display, use the keypad to enter numbers and make choices. Pressing the EXIT PROGRAM key in the configure option list will return the sampler to standby. Pressing the EXIT PROGRAM key in a configure option input display will return you to the list of configure options without changing the display’s setting.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Note The sampler uses the bottle configuration setting to position the distributor. If you enter a bottle configuration that does not correspond to the actual configuration used in the sampler, the sampler may position the distributor over locations without a bottle and fill the refrigerator instead of a sample bottle. 3.9.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.9.7 Bottle Volume The sampler will accept bottle volumes between 100 ml and 30,000 ml. It uses the bottle volume to calculate the range of acceptable sample volumes in the sample-volume input display and to check for probable overfill when running a sampling routine. Table 3-2 lists Teledyne Isco’s standard bottle sizes for composite samplers.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines with the diameter, type, and length of the suction line; the sampler must revise the internal pump tables. The electronic pump count uses the tables as reference for sample delivery. There are three types of suction lines available for use with the 3700R sampler: 1/4-inch vinyl, 3/8-inch vinyl, and 3/8-inch Teflon. If you specify 1/4-inch line, the sampler will prompt you for the line length immediately.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines repeats according to the number of rinse cycles entered (Display #241). Rinse cycles contribute to the wear on pump tubing; therefore, if you use rinse cycles, you may need to replace the tubing more frequently. The Tubing Life configure option allows you to monitor pump-tubing wear.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-3 Sampling Capabilities Through the Program Sequence Pacing Mode Feature Time-Pacing: Basic & Extended Samples taken at regular time intervals from 1 minute to 99 hours, 59 minutes. Extended Samples taken at irregular time intervals by specifying the time and date of each sample. You can enter dates as far as one month in advance of the current date.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-3 Sampling Capabilities Through the Program Sequence (Continued) Storm Distribution Sequential Extended One sample volume placed in each bottle. Samples-per-Bottle Multiplexing Extended More than one sample volume placed in each bottle. You can switch bottles after a specified number of samples have been deposited. Bottles-per-Sample Multiplexing Extended Sample volumes placed in more than one bottle at each sample event.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines factory default settings by loading a program. When you reinitialize the sampler, all four programs return to the default program settings. Example: Load Stored Program 1. Press ENTER/PROGRAM to access the interactive state. . . . STANDBY . . . 10:34:50 19-JUN-04 2. Access the configure sequence by selecting “CONFIGURE.” [PROGRAM, CONFIGURE] SAMPLER 3. Press the LEFT ARROW or RIGHT ARROW key to scroll through the configure options.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines is used at two sites, each requiring a specific sampling routine and different suction-line lengths, the sampling routines can be stored for each site, but you must reenter the suction-line length settings each time the line changes. (Display #260) The sampler is shipped from the factory with four programs: the current program and three stored programs, numbered from 1 to 3. All four programs contain the factory default program settings.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Save Current Program (Continued) 8. Select the number you want to use as the program “name.” Select “NONE” when you do not want to save a program When the sampler has saved the program, it will display the next configure option. When you are done configuring the sampler, press EXIT/PROGRAM to return to Standby. SAVE PROGRAM AS [#1, #2, #3, NONE] 9.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.9.17 Sampling Stop/Resume The Sampling Stop/Resume configure option lets you add stop and resume settings (Display #s 100, 101, and 102) to the program sequence. The Stop/Resume settings appear in the key time section of the program sequence. (You can enter as many as 24 times: 12 stop times and 12 resume times.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Note If you enter a start time, the sampler will disregard the start-time delay. The sampler will use the start-time delay setting when you do not enter the start time in the program sequence. When using a start time delay of greater than 0, take care when the sampler is being inhibited by another sampler, a flow meter, or a Liquid Level Actuator. 3.9.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines A sampler that has been inhibited while running a routine will move to the next bottle set when enabled, if at least one sample volume has been deposited in the current bottle set. This is true for both time-switched and count-switched routines, and for bottles sets composed of one or more bottles.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Presample Purge Rinse Cycle Pump Volume 1 A. Pulse at Beginning of Pre-sample Purge Pump Volume 2 Pump Volume 3 Postsample Purge Event Mark Bottle Number B. Pulse at Beginning of Forward Pumping Only Event Mark Bottle Number Event Mark C. Continuous During Entire Pump Cycle Bottle Number D.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines diameter and length settings. If your observations indicate that a greater or lesser number of pump counts in either purge is needed, you can change the count settings. Entries are to be between 0 and 9,999 for both purges. You can obtain a purge count using the pump forward or pump reverse keys. (Display #s 340 and 341) 3.9.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines TEST” and beeps every three seconds until you turn it off. If either the RAM or ROM fail the test, service the sampler. Contact Teledyne Isco Technical Service for assistance. LCD Test – After successful RAM and ROM tests, the sampler will then test the LCD by first filling the display with solid rectangles and then printing the alphabet and other characters. Each character position in the display should contain a character.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-4 Factory Configure Option Settings (Continued) Configure Option Display Number Factory Setting 243 0 PROGRAMMING MODE 250 BASIC LOAD STORED PROGRAM 255 SAVE CURRENT PROGRAM 260 # of retries FLOW MODE SAMPLING sample at start time 270 NO sample at switch times 271 NO NONUNIFORM TIME 280 MINUTES CALIBRATE SAMPLER 290 DISABLE SAMPLING STOP/RESUME 300 DISABLE sample at stop times 301 NO sample at resume tim
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-5 Factory Program Settings Program Setting Factory Setting [TIME, FLOW] PACED SAMPLING 10 TIME UNIFORM/ NONUNIFORM TIME INTERVALS 11 UNIFORM [TIME, FLOW, STORM] PACED SAMPLING 12 TIME [TIME, FLOW] BOTTLE GROUP 14 FLOW -- MINUTE DELAY TO FIRST TIMED SAMPLE 15 0 MINUTES [DURING, AFTER] TIME MODE 16 DURING _MINUTE DELAY TO SECOND GROUP SAMPLES 17 30 21 1 HOUR 0 MINUTES SAMPLE EVERY -- PULSES (1 - 9999) 22 10 T
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-5 Factory Program Settings (Continued) Program Setting Display Number Factory Setting CHANGE SETS AFTER SAMPLES 45 1 SAMPLE CONTINUOUSLY? [YES, NO] 48 NO SAMPLE VOLUMES OF --- ml EACH (10 - MAX) 50 200 --- COMPOSITE SAMPLES (0 - MAX) 60 10 70 10 ENTER START TIME? [YES, NO] 90 NO STORM ENABLED AFTER HH:MM MM/DD 93 ENTER FIRST SWITCH TIME? [YES, NO] 95 NO 100 0 110 2 HOURS 0 MINUTES SUCTION HEAD OF -- FEET (
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Each of these extended features is discussed briefly below. You may find it helpful to refer to Figures 3-10, 3-11, and 3-12, which chart the program sequence in the extended mode. 3.10.1 Extended Mode Sample Pacing In the extended programming mode, the Sample Pacing section allows you to select flow pacing, or one of two types of time-pacing: uniform or nonuniform.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines grammed number of samples. You can use select either “TIME,” “SAMPLES,” or “SAMPLE/BTL” settings with both time- and flow-paced sampling. 3.10.3 Extended Mode Sample Volumes The extended mode Sample Volumes section is identical to the basic Sample Volume section. 3.10.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines [PROGRAM, CONFIGURE] Configure To Configure SAMPLER Program Display #12 Sample Pacing [TIME, FLOW, STORM ] Storm PACED SAMPLING Flow Time Display #11 Display #22 [UNIFORM, NONUNIFORM] TIME INTERVALS Uniform Time SAMPLE EVERY ---- PULSES (1 - 9999) Nonuniform Time Display #21 Display #20 SAMPLE EVERY -- HOURS -- MINUTES MODIFY SEQUENCE? [YES, NO] Yes No Display #25 TAKE --- SAMPLES (1 - MAX) Clock-Time Minutes Display
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Continued from Figure 3-10 Display #50 Sample Volume SAMPLE VOLUMES OF --- ml EACH (1-MAX) Display #70 SUCTION HEAD OF -- FEET (1-MAX) See Chapter 3 Liquid Detector configure option Display #80 CALIBRATE SAMPLE VOLUME? [YES, NO] See Example 7 Calibration Procedure Display #90 Key Times ENTER START TIME? [YES, NO] Yes No Flow Time Display #92 Display #91 START FLOW COUNT AT HH:MM DD-MM TAKE FIRST SAMPLE AT HH:MM DD-MM Change
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines First Bottle Group [PROGRAM, CONFIGURE] SAMPLER TIME MODE Configure FIRST BOTTLE GROUP Display #15 Program --- MINUTE DELAY TO FIRST GROUP SAMPLE Display #12 [TIME, FLOW, STORM ] PACED SAMPLING Display #23 Storm TAKE -- TIMED SAMPLE EVENTS (1 - MAX) Display #24 SAMPLE INTERVALS OF -- MINUTES (1 - 999) Display #30 __ BOTTLES PER SAMPLE EVENT (1 - MAX) Display #31 __ SAMPLES PER BOTTLE (1 - MAX) Display #50 SAMPLE VOLUMES OF --- ml
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Stops and Resumes – The Stops and Resumes settings are available only when you have enabled the Sampling Stop/Resume configure option. You must first enter the number of stops and resumes, from 0 to 24. (Enter 0 if you want to omit the stop and resume settings without returning to the configure sequence and disabling the Stops and Resumes option.) Then enter the stop and resume clock times. The first entry will be a stop time.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines START SAMPLING START TIME Delay to Start Time Determined by start time entry or,start time delay Sampling Routine Figure 3-13 Simplified Start Time Diagram The start time for a routine is either the programmed start time or the time at which the delay-to-start time expires. The programmed start time is determined by entries made in Displays #90, #91, #92, and #93.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines If the sampler is enabled when the start time occurs, it will begin the routine. The sampler will initiate a sample event at the start time for all time-paced routines. It will initiate a sample event at the start time for flow-paced routines when configured for sample at start time in the Flow Mode configure option. Refer to Figure 3-14.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines I I I I Figure 3-14 Start Time Diagram 3-66 I I
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Figure 3-15 Start Time Diagram for Nonuniform Time Routines 3-67
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.12 Stormwater Sampling Many stormwater runoff monitoring permits require a first-flush grab sample within the first 30 minutes of the storm event and a time- or flow-paced composite sample during the remainder of the event. The 3700 Sampler’s STORM pacing is ideal for monitoring storm events. Storm pacing allows you to: • Divide the sampler’s bottles into two groups. The first bottle group receives the first flush grab sample.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines • How to start and restart the monitoring installation Equipment Checklist – The equipment listed below is suggested for typical storm-event monitoring: • 3700 Sampler and power source.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines meter before you press the start sampling key on the sampler. If you are using the flow meter to detect a storm event and enable the sampler when the event occurs, the sampler must be disabled by the flow meter when you press the start sampling key. If the sampler is not disabled, it will begin the routine as soon as you press the key. This means that the flow meter must not detect the enable control conditions.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Table 3-7 Recommended Configurations for Storm Sampling (Continued) Configure Option Display Number Recommended Setting Enter head manually 242 NO # of retries 243 0 Programming Mode 250 Select “EXTENDED.” Load Stored Program 255 Load only a stored STORM program. Save Current Program 260 Not applicable. Sample at start time 270 Disregarded by STORM routines. Sample at switch times 271 Disregarded by STORM routines.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.13 Programming the Sampler for a Storm Routine The extended programming mode contains two program sequences. The first sequence contains the displays for time- and flow-paced routines. The second sequence contains the displays for STORM paced routines. Figures 3-10 and 3-11 show first sequence displays; Figure 3-12 shows the STORM displays. The storm input displays are divided into three sections.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Figure 3-16 Storm Sampling: Flow Pacing for the Second Bottle Group 3-73
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Figure 3-17 Storm Sampling: Time Pacing for the Second Bottle Group The “TIME MODE” message marks the beginning of the first section. This message is followed by Display #15, below. -- MINUTE DELAY TO FIRST GROUP SAMPLE Display #15 Use this display to set amount of time between the time the sampler is enabled and the first sample delivered to the first bottle group. (The sampler may be enabled at the start time.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines How to allocate bottles to the bottle groups – Use the next three displays to determine the both the sample distribution and the number of bottles allocated to the bottle groups. TAKE 2 TIMED SAMPLE EVENTS (1 - MAX) Display #23 3 BOTTLES PER SAMPLE EVENT (1 - MAX) Display #30 1 SAMPLES PER BOTTLE(1 - MAX) Display #31 A bottle group can contain one or more bottles. You can allocate all bottles but one to the first bottle group.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines After you’ve completed the entries for the first bottle group, the sampler will present the following message: -- BOTTLES, ----- ml AVAILABLE This message marks the beginning of the second section. It reports the number of bottles allocated for the second bottle group and the total volume of their combined capacity.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines For example, if the delay to the first group were 0 and the routine required two timed sample events at 30-minute intervals, the sampler would take 30 minutes to fill the first bottle group. The first sample event would occur after the programmed delay to group sample; the second sample event would occur 30 minutes later. Use Display #24 to enter the time interval for the second bottle group.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines has not been adjusted to reflect the higher rate. If this occurs, the sampler will continue the first bottle group until all samples have been taken. You can adjust the pulse interval by • Reprogramming the flow meter to change the number of flow pulses sent to the sampler by the flow meter. • Reprogramming the sampler to change the flow pulse interval in the flow-pacing segment of the STORM routine.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Depending on whether the second bottle group is time paced or flow paced, the third delay period will be either the delay to second group sample or the “During/After” delay. • Delay-to-Start Time. The start time is simply the time the sampler begins the routine. The sampler begins timing the delay-to-start time when you press the start sampling key.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Figure 3-18 Start Time Diagram for STORM Routines Note Before programming a sampler in the extended programming mode, the sampler must be configured for extended programming. Refer to the section Programming Mode for a discussion of the Programming Mode configure option. Example Checking the Configure Option Settings shows the steps required to access the Programming Mode option.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.14 Foreign Languages and Metric Units of Measure The 3700R Sampler provides displays in French, Spanish, and German. Additionally, the software supports entries in metric units of measure. Metric units include volumes in milliliters, suction-head and suction-line length in decimeters, and suction-line ID in millimeters.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.15 Programming Examples The examples throughout this section demonstrate the steps used to program the sampler for several different sampling routines in the extended programming mode. Each programming example concludes with the run state displays that appear for that routine.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Sampling: Uniform Intervals (Continued) TAKE FIRST SAMPLE AT 8:00 20-06 JUN After this message displays briefly, the sampler will automatically return to the standby state. PROGRAMMING SEQUENCE COMPLETE 13. When the sampler is properly installed, press the START SAMPLING key to run the program. . . . STANDBY . . . 10:37:23 19-JUN-04 14.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Multiplexed (samples-per-bottle) Sampling 1. Press ENTER/PROGRAM to access the interactive state. . . . STANDBY . . . 5:37:23 19-JUN-04 2. Access the program sequence by selecting “PROGRAM.”. [PROGRAM, CONFIGURE] SAMPLER 3. Select “TIME.” [TIME, FLOW, STORM] PACED SAMPLING 4. Select “UNIFORM.” [UNIFORM, NONUNIFORM] TIME INTERVALS 5. Enter “0” to set the hours at zero.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Multiplexed (samples-per-bottle) Sampling This display is used with samples-per-bottle multiplexing. The first line indicate the number of the upcoming sample and the total number of samples each bottle is to receive. It also indicates the current bottle number. Thus, this display indicates that the upcoming sample is the first of four samples and bottle one is to receive the next sample.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example Extended Flow-Paced Multiplexed (bottles-per-sample) Sampling shows how to program the sampler to take samples at intervals of 10 pulses. Samples will be multiplexed: 800 ml sample volumes will be placed into two bottles at each sample event. Each bottle is to receive one sample volume. The routine will use the start time delay. Example: Extended Flow-Paced Multiplexed (bottles-per-sample) Sampling 1.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Extended Flow-Paced Multiplexed (bottles-per-sample) Sampling (Continued) This display appears after you press the start sampling key to start the routine. It reports the scheduled start time on the first line. The second line reports the current time and date. START FLOW COUNT AT 10:37:23 20-06 JUN When the scheduled start time arrives, the sampler will begin to count flow pulses.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines In Example Flow-Paced Sampling / Time-Based Bottle Sets, samples are to be taken at intervals of 1,000 flow pulses. Samples will be multiplexed by placing 100 ml sample volumes into two bottles at each sample event. To avoid missing any samples, the sampling routine starts immediately after you press the start sampling key.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Flow-Paced Sampling / Time-Based Bottle Sets (Continued) ENTER FIRST SWITCH TIME? [YES, NO] 12. Enter the time at which you want the sampler to move to the second bottle set. At the first bottle set change, 12:00, the bottle set changes will be placed on schedule. FIRST SET CHANGE AT 19-06 JUN 12:00 After this message is displayed briefly, the sampler will automatically return to the standby state. PROGRAMMING SEQUENCE COMPLETE 13.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Flow-Paced Sampling / Time-Based Bottle Sets (Continued) When the next sample event occurs, this display reports the current sample event and the current bottle. When the sample delivery is completed, the distributor will move to the next bottle of the set and the following display will appear. SAMPLE 2, BOTTLE 1 SAMPLE 2, BOTTLE 1 PUMPING 100 ml This cycle will be repeated, as samples are deposited in the bottles of the set.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Samples at Nonuniform Intervals 1. Press ENTER/PROGRAM to access the interactive state. . . . STANDBY . . . 10:34:50 19-JUN-04 2. Access the program sequence by selecting “PROGRAM.” [PROGRAM, CONFIGURE] SAMPLER 3. Select “TIME.” TIME, FLOW, STORM] PACED SAMPLING 4. Select “NONUNIFORM.” [UNIFORM, NONUNIFORM] TIME INTERVALS 5. Select “YES” to modify the sequence of nonuniform time entries. MODIFY SEQUENCE? [YES, NO] 6.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Time-Paced Samples at Nonuniform Intervals (Continued) ENTER START TIME? [YES, NO] After this message is displayed briefly, the sampler will automatically return to the standby state. PROGRAMMING SEQUENCE COMPLETE 15. Press the START SAMPLING key to run the program. . . . STANDBY . . . 10:37:23 19-JUN-04 16. To start the sampling routine with the first bottle, accept the blinking “1” by pressing the ENTER/PROGRAM key.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Nonuniform times can be entered in either the method shown in Example Time-Paced Samples at Nonuniform Intervals, or by specifying a clock time and date for each sample. On the following page, Example Time-Paced Samples at Nonuniform Clock Times shows the displays used to enter the clock times.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Sampling with Stops and Resumes 1. Press ENTER/PROGRAM to access the interactive state. . . . STANDBY . . . 10:34:50 19-JUN-04 2. Access the program sequence by selecting “PROGRAM.” [PROGRAM, CONFIGURE] SAMPLER 3. Select “TIME.” [TIME, FLOW, STORM] PACED SAMPLING 4. Select “UNIFORM.” [UNIFORM, NONUNIFORM] TIME INTERVALS 5. Enter “0” to set the hours at zero. Press ENTER/PROGRAM to store the number “0” and move to the minutes entry.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Sampling with Stops and Resumes (Continued) RESUME SAMPLING AT 1. 11:00 19-JUN 16. Enter the time and date of the second stop time: 1:30 p.m. or 13:30. You must enter the times in military time. STOP SAMPLING AT 2. 13:30 19-JUN 17. Enter the time and date the program should resume: 4:00 p.m. or 16:00 in military time. STOP SAMPLING AT 2. 13:30 19-JUN 18. Enter the time and date the program should resume: 4:00 p.m.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Sampling with Stops and Resumes (Continued) DONE . . . 18 SAMPLES 18:10:35 21-JUN-04 Example Storm Sampling illustrates how to program a 24-bottle sampler to allocate six bottles to the first bottle group. The remaining 18 bottles, allocated to the second bottle group, will be used for flow-weighted samples. The sampler delays the first group 25 minutes after it is enabled.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Storm Sampling (Continued) 8. Enter “1” to place one sample in each bottle. 1 SAMPLES PER BOTTLE (1 - 2) Display 31 9. Enter the sample volume of “900” ml. SAMPLE VOLUMES OF (10 - 1000) Display 50 900 ml The sampler will display this message before automatically advancing to the next display. 18 BOTTLES, 18000 ml AVAILABLE 10. Select “FLOW.” [TIME, FLOW] SECOND BOTTLE GROUP Display 14 11.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Storm Sampling (Continued) The displays shown in step 22 alternate as the sampler’s clock approaches the start time. The first line reports the bottles which will receive the first sample volumes for each bottle group. The second line reports the start time at the left and the current time on the right.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines At this point, the following keys are operable: ON / OFF, PUMP REVERSE , PUMP FORWARD , START SAMPLING, ENTER / PROGRAM , MANUAL SAMPLE, NEXT BOTTLE, and DISPLAY STATUS. Program Halted – A sampling program can be halted with either the STOP or EXIT PROGRAM key; if this occurs, the sampler will display the halt message shown below.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Start Sampling – Once you have programmed and configured the sampler, use the START SAMPLING key to run the sampling program. The first display you’ll see after pressing the START SAMPLING key is illustrated below. If you want to have the sampling program begin with a bottle other than Bottle 1, enter the first bottle number here. If “1” is acceptable, press the ENTER/PROGRAM key to place the sampler into the run state.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Program Started Later Than Programmed Start Time 2. This display appears when you select “YES” in display #142 and a sample event is to occur at the start time. Enter the new start time. TAKE FIRST SAMPLE AT 12:30 19-JUN Display 93 3. This display appears when you select “YES” in display #142 and no sample is to occur at the start time for a flow-paced sampling routine. Enter the new start time.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.17 Displayed Selections and Messages Start Sampling – Once you have programmed and configured the sampler, use the START SAMPLING key to run the sampling program. There are two instances where the sampler will not be able to run the program immediately. When the program has been started after the programmed start time or after one or more programmed stop times.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.17.2 Display Status You can access a summary of the current program settings and the results of the most recent sampling routine with the DISPLAY STATUS key. This key is valid in both the run and standby states. The summary of the program settings consists of a number of informational displays describing the settings.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Review – Select “REVIEW” to review the program settings and sampling results. The sampler will present the display shown below. Use this display to leave the Display Status procedure or to review the program settings or sampling results. REVIEW PROGRAM [NO, SETTINGS, RESULTS] Display #150 No – Select “NO” to return to the previous operating state. If you entered Display Status from standby, the standby message will reappear.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines . . . PRINTING . . . PROGRAM INFORMATION PRINT ABORTED NO PRINTER DETECTED! Example: Reviewing the Sampling Results 1. Press display status to view the results of the sampling program. DONE . . . 12:34:50 24 SAMPLES 20-JUN-04 2. Select “REVIEW.” [REVIEW, PRINT] PROGRAM INFORMATION 3. Select “RESULTS.” REVIEW PROGRAM [NO, SETTINGS, RESULTS] 4. This display reports the start time and date of the program.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.17.3 Source of Sample Event The display illustrated in step 6 of Example Reviewing the Sampling Results shows the sample event number, the bottle(s) used, and the source of the sample event. Source refers to the programmed or configured setting that initiated the sample event. For example, the sampler will report “TIME” as a source if the sample was taken as one of the program’s time-paced samples.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines 3.17.4 Error Messages and Missed Samples The probable cause of any missed sample is reported after the sample number/source display. A missed sample display is illustrated below. SAMPLE MISSED: NO LIQUID DETECTED! Twelve causes are tracked and reported: Pump ’STOP’ Key Hit! – The sampler was halted with the stop key during the sample event. Pump Jammed! – The sampler was unable to take the sample because the pump jammed.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines you to monitor the sampler’s progress. The displays are listed with a brief explanation in Examples: Run State Displays: Time-Paced Sampling, Run State Displays: Flow-Paced Sampling, and Run State Displays: Composite Sampling. If you encounter a problem while running a sampling routine that results in missed samples, an asterisk will appear in the lower right corner of the display, as illustrated below.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Run State Displays: Time-Paced Sampling (Continued) Indicates the sample and bottle numbers of the next sample event. The current time is shown in the lower right corner. The sampling program requires four bottles per sample event and that the bottle sets be switched on a time basis. This display alternates with the display illustrated in item 8 at 3-second intervals.
3700R/3740 Refrigerated Sampler Section 3 Programming Guidelines Example: Run State Displays: Flow-Paced Sampling (Continued) NEXT BOTTLE CHANGE AT 10:00 19-JUN Indicates the time of the next bottle switch. SAMPLE 2, BTLS 1 - 4 AFTER 10 PULSES Indicates the time of the next bottle-set switch. NEXT SET CHANGE AT 10:00 19-JUN Indicates a sample in progress.
3700R/3740 Refrigerated Sampler Section 4 Options and Interfacing Equipment This section discusses major options available with the Teledyne Isco 3700R. 4.1 Connections to External Devices The 3700 Series Samplers can collect samples on a flow-paced basis using flow inputs from an external flow meter. The sampler requires a 5-to-15 VDC pulse of at least 25-millisecond duration to register a flow pulse.
3700R/3740 Refrigerated Sampler Section 4 Options and Interfacing Equipment Figure 4-1 6-pin Connector Diagram Table 4-1 Flow Meter Connector Wiring A +12 VDC B Common C Flow Pulses In D Bottle Number Out Master/Slave Out E Event Mark Out F Inhibit In Table 4-2 Printer Connector Wiring 4.1.2 Non-Isco Flow Meters 4-2 A +12 VDC B Common C Printer Source D Transmit E Receive F NC You can connect certain non-Isco Flow meters directly to a 3710 for flow-paced sampling.
3700R/3740 Refrigerated Sampler Section 4 Options and Interfacing Equipment Teledyne Isco also offers a connector prewired to a 22-foot (6.7-meter) cable that terminates in two wires. The black wire is connected to pin A; the white wire to pin C. Note that you will be unable to communicate anything other than the flow pulse contact closure with a non-Isco Flow meter. The non-Isco Flow meter will not be able to interpret event and bottle number information.
3700R/3740 Refrigerated Sampler Section 4 Options and Interfacing Equipment Figure 4-3 1640 Liquid Level Actuator The actuator can be set to control the sampler in two modes: Latch Mode – The sampler continues the sampling routine even though the liquid level has receded and liquid no longer contacts the sensor. Toggle Mode – The sampler will halt the routine when the liquid ceases to contact the sensor.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service The 3700R/3740 is ruggedly built to withstand difficult field operating conditions and will maintain maximum reliability when these simple maintenance procedures are followed. As with any piece of field operated equipment, preventive maintenance is necessary to keep it functioning properly. 5.1 Weather and Corrosion Resistance The 3700R/3740 is designed to be operated in relatively hostile environments.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5.3 Cleaning Protocols for Priority Pollutants The following sections are excerpted from U.S. Environmental Protection Agency Publications EPA-600/4-77-039 (“Sampling of Water and Wastewater” by Dr. Phillip E. Shelley) to provide an example of sampler cleaning procedures for priority pollutants.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5. Rinse thoroughly three times with tap water. 6. Rinse thoroughly three times with distilled water. 7. Rinse thoroughly with petroleum ether and dry by pulling room air through tubing. 8. Dry overnight in warm oven (less than 150° F), if possible. 9. Cap ends with aluminum foil. 5.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5.4.1 Cleaning the Bottle Locating Base and Sample Bottles 5.5 Pump Tubing 5.5.1 Inspection of Pump Tubing The bottle locating base may be cleaned with soapy water and a sponge. The sample bottles have a wide mouth to facilitate cleaning. The 2.5 gallon glass bottle may be washed with a brush and soapy water then washed in a dishwasher or autoclaved. The 2.5, 4, and 5.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service WARNING Pump may actuate without warning. To avoid injury, sampler must be off when pump housing cover is removed for inspection or tubing replacement. Note The black bands used to assist you in correctly positioning the tubing in the pump and detector are placed on one end only. If you turn the tubing around, you will not be able to use the bands for reference.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Machine Screw Pump Rotor Pump Shaft Locate bands here to place pump tubing correctly Figure 5-2 Interior of Liquid Detector and Pump Tube 5.5.2 Removing the Pump Tubing WARNING Pump may actuate without warning. To avoid injury, sampler must be off when pump housing cover is removed for inspection or tubing replacement. To remove the tubing: 1. Disconnect power from the unit. 2.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service pump. Correct placement is critical to prolong the life of the pump tube and to assure efficient operation and accurate sample volumes. Facing the liquid detector, place the inner edge of the end band against the upper left inlet of the liquid detector. Place the inner band at the lower outlet of the liquid detector. Figure 5-2 shows the correct placement of the tubing at the liquid detector. 2.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service line is removed from the pump tubing by detaching the tube coupling, as described in Section 2.2. If a complete new suction line-strainer assembly of the standard 10 or 25 ft. length is used, it should be installed as described in Section 2.2. 5.7.1 Bulk Suction Line If you are using bulk suction line to replace the old line, detach the old line from the tube coupling and remove the strainer.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service CAUTION Desiccant may produce irritating fumes when heated. Observe the following precautions: • Use a vented oven in a well ventilated room. • Do not remain in the room while recharging is taking place. • Use the recommended temperature. There have been reports of irritating fumes coming from the desiccant during reactivation. While our attempts to duplicate the problem have been unsuccessful, we still urge you to use caution.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Screws (10) Humidity Indicator Figure 5-3 Location of 10 Screws on the Control Box Frame 5-10
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Grounding Strap Desiccant Bags (3) Figure 5-4 Control Box Internal Desiccant 5.10 3700 Controller Mounting The 3740 Sampler Refrigerator is available to convert a 3700 Portable Sampler into a 3700R Refrigerated Sampler. To do this, remove the control box from the portable sampler following the instructions in the 3700 Portable Sampler Instruction Manual.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 3. Carefully insert the threaded rods into the four mating holes on the refrigerator and push the controller down until it is resting on the refrigerator, install the spacers and wing nuts on the four threaded rods. 4. Feed the pump tube down through the tubing feed through into the interior of the refrigerator and route the pump tube into the distributor arm. 5.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Figure 5-6 3740 Electrical Schematic 5.12 Refrigerator Electrical System An electrical circuit schematic for the refrigerator and the sampler’s 12 VDC power supply is shown above in Figure 5-6. WARNING Disconnect the refrigerator’s power before performing any service activities. Access to the electrical and refrigeration components can be gained by removing the back panel of the refrigerator.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Capillary Tube Condenser Coil Evaporator Plate Compressor Figure 5-7 Refrigeration Schematic Diagram 5.14 Servicing the 3700 Controller The electronic circuitry of the 3700R Sampler is solid-state and its reliability is high. If the unit should fail to operate properly, items such as a broken or intermittent connection in the power cable or wiring harness should be suspected. 5.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5. The main circuit board assembly of the 3700 Sampler is protected by an aluminum shield. To remove this shield, unscrew the stop nut that holds the short grounding strap and unscrew the four screws located in each corner of the shield. Lifting the shield off will then reveal the main circuit board shown in Figure 5-9. 6.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5.19 Precautions for Servicing CMOS Circuitry Most of the circuitry in the 3700 controller is made up of CMOS components. Because of the oxide gate structure of these devices, they are extremely susceptible to destruction caused by the discharge of static electricity through their inputs. WARNING Disconnect power from the refrigerator and controller when working on the unit.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Hex Standoff (4) Phillips Head Screw Connector Figure 5-9 Main Circuit Board Nut Nut Grounding Strap Screw Figure 5-10 Control Box Tray Removal 5-17
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Ground Strap Ground Strap Figure 5-11 Distributor Gear Case Removal 5.20 Preliminary Electronics Troubleshooting Steps Following are suggested areas to check before attempting to service the microprocessor CPU and associated circuitry. These checks should be made before looking at the CPU and memory. 1. Supply voltage is correct. 2. Wiring harnesses, connectors, and solder joints are in good condition. 3.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service must perform. The software is discussed only as is necessary to describe the operation of the circuitry (hardware). The following sections discuss the hardware in greater detail. A 12 VDC connector on the side of the control box is used to connect to an external 12 V source. A 3.75 Amp PTC device provides circuit protection.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service P8 to IC U15 and its associated circuitry where it is massaged into a YES or NO indication of liquid presence. IC U15, containing a switched capacitor low pass filter and two Op Amps, is the main circuit component of the detection system. A collection of transistors, resistors, and diodes are used to control the distributor motor.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service Figure 5-12 3700 Main Circuit Board Component Layout 5-21
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 201 202 203 Q201 204 Q202 R201 R202 205 R203 208 209 R204 206 210 207 C 201 212 211 D 201 P201 K201 IC201 Figure 5-13 Optical Counter PCB Component Layout 5.21 Sample Event Cycle and Delivery of Sample Volumes The sequence of steps in a typical sequential sample event is described below.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5.21.1 Peristaltic Pump The following is a brief description of the method by which the sample volume is determined. The 3700 Sampler uses a peristaltic pump to transport the sample from the source to the sample bottle. When compared with other suction lift sample gathering methods, a peristaltic pump has numerous advantages: simplicity, reliability, no metering chamber required, easily cleaned, etc. 5.21.
3700R/3740 Refrigerated Sampler Section 5 Routine Maintenance and Service 5-24
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists The following appendix contains illustrated replacement parts listings, including part descriptions and order numbers. Replacement parts can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco, Inc. Customer Service Department P.O. Box 82531 Lincoln, NE 68501 USA Phone: (800) 228-4373 (402) 464-0231 FAX: (402) 465-3022 E-mail: IscoInfo@teledyne.
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-2
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-3
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-4
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists DWG 603703266 REV B A-5
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists DWG 603703266 REV B A-6
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists DWG 603703266 REV B A-7
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-8
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists Screw, 10-32 x 3/4 Locknut, 10-32 A-9
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-10
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-11
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-12
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-13
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-14
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-15
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-16
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-17
3700R/3740 Refrigerated Sampler Appendix A Replacement Parts Lists A-18
3700R/3740 Refrigerated Sampler Appendix B Accessories The following appendix contains the most commonly ordered accessories for the 3700R Sampler, including part descriptions and order numbers. For additional accessories, contact Teledyne Isco Customer Service or your local sales representative. Accessories can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco, Inc. Customer Service Department P.O.
3700R/3740 Refrigerated Sampler Appendix B Accessories 1 /4" ID vinyl tubing, bulk 1000' . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-1680-057 3 /8" ID vinyl tubing, bulk 100' . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-1680-058 3 /8" ID vinyl tubing, bulk 500' . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3700R/3740 Refrigerated Sampler Appendix B Accessories 1000 ml wedge-shaped polypropylene bottles, qty of 100, with polyethylene foam lined caps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3700-046 350 ml round glass bottles, set of 24, includes Teflon lined caps . . . . . . . . . . . . . . . . . . . . . . . 68-2100-001 350 ml round glass bottles, qty of 100, includes Teflon lined caps . . . . . . . . . . . . . . . . . . . . . . 68-3700-048 B.6.
3700R/3740 Refrigerated Sampler Appendix B Accessories B-4
3700R/3740 Refrigerated Sampler Appendix C Display Index Note To access the display number, press the STOP key while the sampler displays the screen in question. The display numbers are available in the Standby and Program states. 1 2 20 4 3 20 20 5 20 6 20 7 20 6 intervals of 20 minutes 9 8 30 30 30 12 11 10 30 30 13 30 14 30 9 intervals of 30 minutes Start Time 15 30 16 30 17 18 19 Event No.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) This display appears only in the extended programming mode and follows Display #10 when you have selected “TIME.” Select “UNIFORM” for uniform time intervals, “NONUNIFORM” for nonuniform time intervals. Display11 [UNIFORM, NONUNIFORM] TIME INTERVALS This display appears after you select “PROGRAM” in Display #1.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) This display follows Display #20 when you select “YES” to modify the nonuniform time-interval sequence. Enter the number of sample events to occur at nonuniform time intervals. MAX varies according to the bottle size entered in Display #223 Display 25 TAKE --- SAMPLES (1 - MAX) This display follows Display #25. Enter the nonuniform clock times and dates for each sample event. Display 26 TAKE SAMPLES AT 1.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) This display appears if you select “TIME” in Display #40. Enter the amount of time each bottle is to receive samples. Display 42 CHANGE BOTTLES EVERY -- HOURS -- MINUTES This display appears if you select “TIME” in Display #41. Enter the amount of time each bottle set receives samples. Display 43 CHANGE SETS EVERY -- HOURS -- MINUTES This display appears after you select “SAMPLES” in Display #40.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) This display is part of the calibration sequence. Press the MANUAL SAMPLE key when a collection bottle is underneath the distributor and the suction line is in place. Display 81 PRESS MANUAL SAMPLE KEY WHEN READY... This display is part of the calibration sequence. When it first appears, it reports the programmed sample volume.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) This display follows Display #22. Enter the maximum amount of time you want between flow-paced sample events in a storm program. If the time between flow pulses exceeds this entry, the sampler will terminate the routine. Display 110 MAX FLOW INTERVAL OF -- HOURS, -- MINUTES This display appears when a routine is halted and you press the start sampling key.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) Bottles and Sizes configure option. This display appears when you select “PORTABLE” in Display #220. Select the number of bottles in the bottle base. Display 221 [1, 4, 12, 24] BOTTLES Bottles and Sizes configure option. This display appears when you select “REFRIG” in Display #220. Select the number of bottles in the base. Display 222 [1, 2, 4, 8, 12, 24] BOTTLES Bottles and Sizes configure option.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) Load Stored Program configure option. Select the sampling program you want to use. Select “NONE” to exit the display without loading a program. Display 255 LOAD PROGRAM [#1, #2, #3, NONE] Save Current Program configure option. Select the number with which you want to use to identify the current program when saved. Select “NONE” to exit the display without saving a program.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) Enable Pin configure option. Select “YES” to take a sample when the sampler becomes enabled; i.e., ceases to receive an inhibit signal from a flow meter or a Liquid Level Actuator. Note: No sample will be taken if the enable signal is used to initiate the start-time delay countdown. Select “NO” if you do not want a sample event when the sampler becomes enabled.
3700R/3740 Refrigerated Sampler Appendix C Display Index Table C-1 Display Index (Continued) Tubing Life configure option. Enter the number of pump counts required to trigger the pump-count warning if the factory setting is not suitable. This value is set to 500,000 when the controller is reinitialized. Display 352 ------ PUMP COUNTS TO WARNING Program Lock configure option. Select “ENABLE” to turn on the program lock. If you enable the program lock, input displays are protected by a pass-number: 3700.
3700R/3740 Refrigerated Sampler Appendix D Calculating Flow Increment Between Samples The 3700R Sampler will accept flow-proportional inputs from an external flow meter or flow logger. These electronic flow-input signals are transmitted to the sampler at fixed increments of total flow, for example, every 10,000 gallons. Each time 10,000 gallons of liquid flows past the flow meter, it sends a signal to the sampler, which registers it as a single flow pulse.
3700R/3740 Refrigerated Sampler Appendix D Calculating Flow Increment Between Samples Calculation of Number of Flow Pulses – O n c e t h e d e s i r e d flow increment between samples is known, the number of flow pulses to be programmed into the sampler may be calculated, assuming that the volume of the flow pulses from the flow meter is known. The number of flow pulses is calculated by dividing the flow increment between samples by the volume of each flow pulse.
3700R/3740 Refrigerated Sampler Appendix D Calculating Flow Increment Between Samples Sample Volume Considerations – Thus, a sample volume of 188 ml will result in the desired composite sample. Because of the basic uncertainty of the delivered sample volume exactly matching the programmed nominal sample volume and the 10 ml sample repeatability, it is good practice to select a nominal sample volume which is slightly smaller than the calculated sample volume.
3700R/3740 Refrigerated Sampler Appendix D Calculating Flow Increment Between Samples D-4
3700R/3740 Refrigerated Sampler Appendix E Glossary Bottles per sample – Bottles per sample is the number of bottles receiving sample volumes at each sample event. Composite sampling – In composite sampling, multiple sample volumes are placed in bottle sets. Typically, composite sampling uses a single container. A composite sample represents an average of the characteristics of the flow stream for the elapsed time of sampling.
3700R/3740 Refrigerated Sampler Appendix E Glossary Sample volume – The sample volume is the discrete, programmed amount of sample delivered to each bottle. A single sample event may deliver several sample volumes. Samples per bottle – Samples per bottle is the number of discrete sample volumes placed in each bottle. Sampling program – See sampling routine.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets This appendix provides Material Safety Data sheets for the internal desiccant used in the 3700R sampler. Specific questions regarding the use and handling of these products should be directed to the manufacturer listed in the MSDS.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant SECTION I -- PRODUCT IDENTIFICATION Trade Name and Synonyms: Chemical Family: Formula: Silica Gel, Synthetic Amorphous Silica, Silicon, Dioxide Synthetic Amorphous Silica SiO2.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant SECTION IV -- FIRE EXPLOSION DATA Fire and Explosion Hazard - Negligible fire and explosion hazard when exposed to heat or flame by reaction with incompatible substances. Flash Point - Nonflammable. Firefighting Media - Dry chemical, water spray, or foam.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant NOTE TO PHYSICIAN: This product is a desiccant and generates heat as it adsorbs water. The used product can contain material of hazardous nature. Identify that material and treat accordingly.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets 101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002 MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT® Packaged Desiccant SECTION IX -- SPECIAL PRECAUTIONS Avoid breathing dust and prolonged contact with skin. Silica gel dust causes eye irritation and breathing dust may be harmful.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets Indicating Silica Gel Material Safety Data Sheet Identity (Trade Name as Used on Label) Manufacturer : MSDS Number* : MULTISORB TECHNOLOGIES, INC. M75 (formerly Multiform Desiccants, Inc.) Address: CAS Number* : 325 Harlem Road Buffalo, NY 14224 Phone Number (For Information): 716/824-8900 Emergency Phone 716/824-8900 Date Prepared: July 6, 2000 G.E.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets Page 2 Section 5 - Health Hazard Data PRIMARY ROUTES OF ENTRY HEALTH HAZARDS Inhalation Ingestion NTP CARCINOGEN LISTED IN Skin Absorption Not Hazardous IARC Monograph Acute May cause eye, skin and mucous membrane irritation. Chronic Prolonged inhalation may cause lung damage. OSHA Not Listed Signs and Symptoms Drying and irritation. of Exposure Medical Conditions Asthma.
3700R/3740 Refrigerated Sampler Appendix F Material Safety Data Sheets F-8
3700R/3740 Refrigerated Sampler Index A D Accessories, B-1 Data Retrieval, 1-8 Delay, 3-49 Desiccant, 5-8 Diagnostics, 3-53 Display Index, C-1 Displays, 3-14, 3-17, 3-102, C-1 Run State, 3-107 Settings and Results, 3-103 B Basic Programming, 3-20 Bottle Locating Base, 2-5 Bottle Rack, 2-5 Bottles, 1-6, 3-39 C Circuit Boards, 5-18 Main Board, 5-19 Pump Control Board, 5-20 Cleaning, 5-1 Protocols for Priority Pollutants, 5-2 Refrigerator, 5-3 Compatible Equipment 4-20 mA Input, 4-3 Data Retrieval, 1-8 F
3700R/3740 Refrigerated Sampler Index P Parts, A-1 Password, 3-53 Power, 1-7, 2-4 Programming Basic, 3-20 Configure Sequence, 3-12 Examples, 3-19, 3-24, 3-82 Extended, 3-6, 3-57 Modes, 3-9, 3-43 Operating States, 3-8 Procedure, 3-13 Save Program, 3-46 Steps, 3-20 Stored Programs, 3-45 Pump Tubing, 1-6, 5-4 Installation, 5-6 Removal, 5-6 Replacement Alert, 3-53 R Refrigerator, 1-2 Defrosting, 2-12 Operation, 2-12 Replacement Parts, A-1 S Sampling Composite, 2-11, 3-7 Cycle, 3-2 Distribution, 3-3, 3-58 Eve
Warranty Teledyne Isco One Year Limited Factory Service Warranty * Teledyne Isco warrants covered products against failure due to faulty parts or workmanship for a period of one year (365 days) from their shipping date, or from the date of installation by an authorized Teledyne Isco Service Engineer, as may be appropriate. During the warranty period, repairs, replacements, and labor shall be provided at no charge.