Instruction manual

3700R/3740 Refrigerated Sampler
Section 3 Programming Guidelines
3-74
Figure 3-17 Storm Sampling: Time Pacing for the Second Bottle Group
The “TIME MODE” message marks the beginning of the first
section. This message is followed by Display #15, below.
Use this display to set amount of time between the time the
sampler is enabled and the first sample delivered to the first
bottle group. (The sampler may be enabled at the start time.)
The acceptable range of values for the delay to first group sample
is 0 to 999 minutes. If the delay is 0, the sampler will take the
first sample when enabled by the flow meter. Subsequent
time-paced samples will occur at the time interval entered in
Display #23.
-- MINUTE DELAY TO
FIRST GROUP SAMPLE
Display #15