Instruction manual

3700R/3740 Refrigerated Sampler
Section 3 Programming Guidelines
3-82
3.15 Programming
Examples
The examples throughout this section demonstrate the steps
used to program the sampler for several different sampling rou-
tines in the extended programming mode. Each programming
example concludes with the run state displays that appear for
that routine.
Example Time-Paced Sampling: Uniform Intervals shows how to
program the sampler to take time-paced samples at uniform time
intervals of 30 minutes starting at 8:00 a.m. tomorrow, placing
one 500 ml sample in each bottle.
Example: Time-Paced Sampling: Uniform Intervals
1. Press ENTER/PROGRAM to access the interactive state.
. . . STANDBY . . .
10:34:50 19-JUN-04
2. Access the program sequence by selecting “PROGRAM”.
[
PROGRAM
, NONUNIFORM
TIME INTERVALS]
3. Select “TIME”.
[
TIME
, FLOW STORM]
PACED SAMPLING
4. Select “UNIFORM”.
[
UNIFORM
, NONUNIFORM]
TIME INTERVALS
5. Enter “0” to set the hours at zero. Press ENTER/PROGRAM to store the number “0” and move to the minutes entry.
SAMPLE EVERY
0
HOURS, 10 MINUTES
6. Enter “30” to set the minutes entry to 30.
SAMPLE EVERY
0 HOURS,
30
MINUTES
7. Enter “1” to set the number of bottles per sample event to 1.
1
BOTTLES PER
SAMPLE EVENT (1-24)
8. Select “SAMPLES”.
CHANGE BOTTLES BASED
ON [TIME,
SAMPLES
}
9. Enter “1” to set the number of samples per bottle to 1.
CHANGE BOTTLES EVERY
1
SAMPLES (1-8)
10. Enter “500” to set the sample volume at 500 ml.
SAMPLE VOLUMES OF
500
ml (10-1000)
11. Select “YES”.
ENTER START TIME
[
YES
, NO]
12. Enter the start time and date: 8:00 tomorrow morning.