Instruction manual

3700R/3740 Refrigerated Sampler
Appendix C Display Index
C-7
Bottles and Sizes configure option. This display appears when you select “PORTABLE” in Display #220. Select the
number of bottles in the bottle base.
Display 221
[1, 4, 12, 24]
BOTTLES
Bottles and Sizes configure option. This display appears when you select “REFRIG” in Display #220. Select the num-
ber of bottles in the base.
Display 222
[1, 2, 4, 8, 12, 24]
BOTTLES
Bottles and Sizes configure option. Enter the bottle volume in milliliters.
Display 223
BOTTLE VOLUME IS
----- ml
Bottles and Sizes configure option. This display appears when you have entered a bottle volume in Display #223 that
exceeds a standard Teledyne Isco bottle size.
Display 224
---- ml ... ARE YOU
SURE?
[
YES
,
NO]
Suction Line configure option. Select “
1
/4 if you are using
1
/4-inch suction line, “
3
/8 if you are using
3
/8-inch suction
line.
Display 230
SUCTION LINE ID IS
[1/4, 3/8] INCH
Suction Line configure option. This display appears when you have selected “
3
/8” in Display #230. Select “VINYL” if
you are using vinyl suction line, “TEFLON” if you are using Teflon suction line.
Display 231
SUCTION LINE IS
[
VINYL
,
TEFLON
]
Suction Line configure option. Enter the length of the suction line. Do not include the tube coupling or the strainer in
the length measurement.
Display 232
SUCTION LINE LENGTH
IS -- FEET
(
3 - 99
)
Liquid Detector configure option. Select “ENABLE” to turn on the Liquid Detector, “DISABLE” to turn off the Liquid
Detector. If you turn off the detector, you must enter the suction head (Display #70) in the program sequence.
Display 240
[
ENABLE
,
DISABLE
]
LIQUID DETECTOR
Liquid Detector configure option. This display appears when you select “ENABLE” in Display #240. Enter the number
of rinse cycles. Rinse cycles condition the suction line to reduce cross-contamination.
Display 241
RINSE CYCLES
(
0 - 3
)
Liquid Detector configure option. This display appears when you select “ENABLE” in Display #240. Select “YES” to
add the suction-head setting (Display #70) to the program sequence. Select “NO” to omit the setting.
Display 242
ENTER HEAD MANUALLY?
[
YES
,
NO
]
Liquid Detector configure option. This display appears when you select “ENABLE” in Display #240. Enter the number
of retries: the number of times the sampler will try to detect the presence of liquid for each sample event.
Display 243
RETRY UP TO - TIMES
WHEN SAMPLING
(
0 - 3
)
Programming Mode configure option. Select “BASIC” to use the basic programming mode. To use the extended pro-
gramming mode, select “EXTENDED.
Display 250
[BASIC, EXTENDED]
PROGRAMMING MODE
Table C-1 Display Index (Continued)