Instruction manual

3700R/3740 Refrigerated Sampler
Section 3 Programming Guidelines
3-64
Figure 3-13 Simplified Start Time Diagram
The start time for a routine is either the programmed start time
or the time at which the delay-to-start time expires. The pro-
grammed start time is determined by entries made in Displays
#90, #91, #92, and #93. The delay-to-start time is determined by
the entry in Display #310 of the Start Time Delay configure
option. These input displays are shown in Table 3-6 and in the
charts in Figures 3-8, 3-10, 3-11, and 3-12.
Note
Unless the sampling routine contains a programmed start time
(Displays #91, #92, or #93), the sampler will delay the start
time according to the amount of time specified in the Start
Time Delay configure option. A programmed start time entry
always overrides any settings made in the Start Time Delay
configure option.
Teledyne Isco 4200 Series Flow Meters and 4100 Series Flow
Loggers provide two essential functions for certain sampling
applications. They send flow-pacing signals — flow pulses — to
the sampler for flow-paced sampling. The second function pro-
vides disable (inhibit) or enable signals to the sampler. A sampler
receiving a disable signal will suspend the sampling routine until
it receives an enable signal.
The sampler will disregard a disable signal it receives from a
flow meter during the delay-to-start time. If the sampler is dis-
abled when the start time occurs, it will suspend the routine
until it is enabled. Once enabled, the sampler will begin the time
or flow intervals. The sampler will initiate a sample event when
enabled for all time-paced routines. For flow-paced routines, it
will initiate a sample event when the sampler is configured to
take a sample-upon-enable in the Enable Pin configure option.
START
SAMPLIN
G
Delay to
Start Time
Determined
by start time
entry or,start
time delay
Sampling
Routine
START
TIME