User guide

4Resting ECG
4.4 Modes of Operation and Procedural Overview
AT-101
Page 26
Art.-no.: 2.510526 Rev.: c
4.4.1 Automatic Mode
Two, user defined automatic mode formats are available. The following can be pro-
grammed freely for each of the 2 formats before recording:
Lead Format
Chart Speed
With the optional interpretation program it is also possible to select the rhythm
lead(s), measurement table, average cycles with optional markings and interpreta-
tion statements for the printout.
For further information and to define the auto formats see paragraph 5.2 ECG Set-
tings.
« To start an automatic ECG recording in Auto Mode Format 1, press the START
key.
« To start the automatic ECG recording in Auto Mode Format 2, press the SHIFT
key followed by the START key.
After approximately 10 seconds the recording is analysed and the printout gives the
following:
ECG recording of all leads in either Standard or Cabrera format according to sel-
ection
Sensitivity
Heart Rate
Speed
Filter Settings
Time and Date
Interpretation statements (option)
Average Cycles
intervals
Axis
Sokolow Index (ECG index for hypertrophy)
Detailed Measurement Table
The softkey options change at the end of the recording to enable you to save (1) the
recording or to obtain an extra copy (2). When a recording has been saved, it remains
stored by the AT-101 until deleted, even when the unit is switched off. Accessing re-
cordings in the memory is detailed on paragraph 4.5.
When an external printer is connected, the printout is automatically directed to the ex-
ternal printer. When the external printer is unconnected or switched off, the AT-101
automatically switches to the internal thermal printer.
A progrees bar (3) is displayed when printing to an external printer.