User guide

5 Setup
5.3 System Settings
AT-101
Page 46
Art.-no.: 2.510526 Rev.: c
5.3.4.2 Communications Test
5.3.4.3 Installing New Software Options (Upgrade)
Use the upgrade option to install any available software options (e.g. Measurement).
To install new options in the AT-101, a code must be entered. This code must be ob-
tained from SCHILLER. To install software option proceed as follows:
5.3.4.4 Update the Software
When this is selected, test options are given for the RS-232 communication port. Use
this test if the RS-232 port is suspected of malfunction. A special test plug is used to
carry out the UART test.
PRINT
SETUP
EXIT
COMM
TEST
BASE
INIT.
SOFTWARE
1. Enter the TEST and INFO screen
SETUP > SYSTEM SETTINGS >TEST AND INFO
2. Select SOFTWARE.
3. Select UPGRADE.
4. Enter the upgrade code.
When the correct code is entered, acceptance of the code is indicated by a series
of beeps. The option can be used immediately.
V More than 10 attempts to enter the incorrect code blocks the unit
« Use the Update option to update the current software.
Details to update the software is given in the AT-101 Service Handbook.
PRINT
SETUP
EXIT
COMM
TEST
BASE
INIT.
SOFTWARE
EXIT
UPGRADE
UPDATE
EXIT
UPGRADE
UPDATE