Instructions

In the anus (rectal)
This is the safest method. It is particularly suitable for infants and small
children. Guide the thermometer point carefully 1 - 2 cm into the anus.
Measuring period: approx. 10 seconds
Under the arm (axillary)
From a medical viewpoint this type of measurement leads to inaccurate values.
It is therefore unsuitable for carrying out exact measurements.
Trouble shoting
Auto detection is error. Please return to the distributor.
Temperature at the probe above 43.9°C
Temperature at the probe below 32.0°C
The device detects low battery. Please change the battery.
Battery replacement
1. When the symbol "
" appears, the battery needs to be replaced.
2. Using a coin, unlock the battery cover by turning it counter-clockwise and
carefully remove the battery with a non-metal instrument from the battery
holder. Avoid of using any sharp metal object in this operation.
3. Discard old battery carefully.
4. Replace the battery with a 3V, CR2032 lithium type in the chamber with the
negative(-) side down and the positive(+) side up.
5. Using a coin, replace and secure the battery cover by turning it clockwise.
Warning :
1. Discard old battery carefully , out of reach of young children.
Swallowing the battery may be fatal. If the battery is swallowed,
contact a hospital immediately to have it
removed. Do not dispose of
the battery in a fire. It may explode .
2. For battery disposition, please follow chapter "Disposal note".
Clean & disinfection instruction
Do not use corrosive detergent to clean the unit.
Clean and disinfect the thermometer before and after use.
Cleaning process:
1. Pre-cleaning with cold water wash several times.
2. Clean the probe with a solution of 5-10% mild detergent about 10~15
minutes.
3. Continuously clean with cold water to avoid any chemical residues
remaining on the device.
4. Rinse the device with drinking water or distilled water to ensure the
removal of all remnants of the cleansing agent.
5. If necessary, repeat step 2 through step 4 in order to make sure the device
is clean.
6. A final wipe down by a clean wipe is necessary to avoid water stains.
Disinfection process:
1. To disinfect the device when completed the cleaning process.
2. Wiping the probe with a cloth dipped in a disinfectant solution such as 70-
75% isopropyl (rubbing) alcohol : contact time 10 min for disinfection.
3. The disinfectant free rinse without residues.
Drying process:
1. After cleaning and disinfection process above.
2. Put the device to air-dry in the room temperature (20-30°C), exposure time
around 30~60 min.
3. And store the device in a dry, clean, and dust-free environment at room
temperature.
Care and storage
1. Do not drop or crush the device. It is not shock proof.
2. Do not dismantle or modify the device other than changing the batteries.
3. Do not store the device in an extreme environment with direct sunlight,
high/low temperatures or in a location with high humidity or severe dust.
Specifications
Measuring range: 32.0°C to 43.9°C
Temperature below 32.0°C: display "L“ for low.
Temperature above 43.9°C:display "H“ for high.
Aaccuracy: +/- 0.1°C between 35.5°C and 42.0°C; other: +/- 0,2°C
Memory: last memory
Battery (include): 1 x CR2032
Battery life: more that 5.000 measurements
Dimension: 117 x 30 x 16 mm
Weight: approx. 20g (inluded battery)
Operation condition: +16°C to +35°C;
at a relative humidity range of 15% to 90%,
Storage condition: -10°C to 50°C; at a relative humidity up to 90%
Warranty
We give a guarantee period of 2 years from the date of purchase. We
guarantee perfect design of this article. We will rectify all material or
manufacturing errors during the guarantee period. If, contrary to expectations,
defects are found, please send the device back to our customer service
department. Please enclose the proof of purchase and a description of the
defect so that the problem can be rectified as soon as possible. The guarantee
excludes damages caused by improper handling, related to the battery or to the
packaging.
Referenced norms
The thermometer complies with the requirements of European norm ISO
80601-2-56 and ISO 13485.
Disposal note
Defect devices or devices which are taken out of operation at the end
of their service life may still contain residues which could present a
risk to the environment. Please dispose of the device in a suitable and
correct way at a collection point for old electrical items.
Batteries must be disposed of correctly. There are containers for
battery disposal at battery dealers and in communal collection points.
Explanation of symbols
K-jump Health Co., Ltd.,
No.56, Wu Kung 5th Rd.
New Taipei Industrial Park,
New Taipei City, 24890
Taiwan
Polygreen Germany GmbH
Ruhlsdorfer Str. 95
14532 Stahnsdorf, Germany
Importer & after-sales service
SCALA Electronic GmbH, Ruhlsdorfer Str. 95, 14532 Stahnsdorf, Germany
Rev. 25/01/2019/GB