OPERATING MANUAL
1) OP ER ATING THE UNIT FOR THE FIRST TIME
Check there is no con den sa tion on the unit. If there is, al low some
time to dry.
Check there is a link be tween + and S+ and be tween – and S– on
the SENSE BLOCK (on rear panel).
Check there is a link be tween the in puts of the In ter lock (CON A,
rear panel).
Set the CV and CC po ten ti om eters to min i mum (fully anti clock -
wise).
For rec om mended ca ble di am e ters and torque (see table 4 - 1)
With high out put cur rent make sure to use low re sis tive con -
nec tions be tween the power sup ply and the load:
- Mount the ca ble lugs di rectly on the tinned out put strips
fol lowed by a washer, a split washer and a nut (see fig. 4 - 1).
Al ways in this or der!
- Never place wash ers be tween the lugs and the strips
be cause this can re sult in ex ces sive heat!
- Only use nuts and wash ers sup plied with the unit.
Switch on unit.
Dis able the Key lock func tion, see next para graph 2).
Check the unit is not in Re mote CV or Re mote CC (LED’s for this
func tion should be off). Press the RE MOTE/LO CAL but ton un til
both LED’s are off.
Turn on the out put by press ing the OUT PUT ON/OFF but ton.
Turn both the CV and CC po ten ti om e ter a few turns clock wise.
A volt age should now be pres ent on the out put.
By press ing the DIS PLAY CV/CC SETTING but ton, the me ters
will show the set ting of the CV and CC po ten ti om e ter.
By press ing the DIS PLAY LIMITS but ton, the volt me ter will show
the set ting of the CV-limit and the CC-limit potentiometer.
Check the cool ing of the unit is not ob structed.
2) KEY LOCK
If the func tion KEY LOCK is ac ti vated, it is no lon ger pos si ble to
op er ate the RE MOTE/LO CAL but ton and the OUT PUT ON/OFF
but ton. This func tion can be use ful to pro tect the out put from ac ci -
den tal shut down. The func tion KEY LOCK does not in flu ence the
operating of the potentiometers.
Ac ti vate Key lock:
Pressing the but tons DIS PLAY SET TINGS and DIS PLAY
LIMITS at the same time for more than 3 sec onds, ac ti vates the
func tion KEY LOCK. The mo ment this func tion is ac ti vated, the
LED’s for RE MOTE CV / CC and for OUT PUT ON will blink a few
times.
Dis able Key lock:
Pressing the same but tons again for 3 sec onds, dis ables the
Key lock function. The LED’s for RE MOTE CV / CC and for OUT -
PUT ON will blink again to in di cate the new set ting.
3) AN A LOG PRO GRAM MING
Set dipswitch 1 of SW1 in po si tion ON to se lect CON E for pro -
gram ming.
Dis able Key lock.
Set the unit in RE MOTE CV for volt age pro gram ming and/or in
RE MOTE CC for cur rent pro gram ming.
Use the RE MOTE/LO CAL but ton and push this but ton sev eral
times un til the right set ting is ac ti vated.
Note that pushing the RE MOTE/LO CAL but ton will shut down the
out put to avoid ac ci den tal dam age to the load.
Con nect the pro gram ming volt age source(s) (0 - 5 V) to the an a -
log pro gram ming con nec tor CON E on the rear panel (see fig.
4 - 2 and fig. 4 - 3). Al ways use a shielded ca ble (max. 30 me ter)
for pro gram ming.
Turn the out put on again with the OUT PUT ON/OFF but ton.
If only the volt age is pro grammed, the max i mum cur rent can still
be set with the CC po ten ti om e ter and vice versa. If this is not de -
sir able the unit can be or dered with Op tion P001 in or der to have
a fixed set ting for the CV or the CC po ten ti om e ter on the front
panel.
fig. 4 - 1
Low re sis tive ca ble con nec tion by mount ing the
ca bles di rectly on the tinned out put strips
fig. 4 - 2
Pro gram ming by volt age:
left volt age -, right cur rent pro gram ming
fig. 4 - 3
Pro gram ming by cur rent
left volt age -, right cur rent pro gram ming
Unit Out put ca bles
[mm2]
Bolts Torque
[Nm]
SM7.5-80 16 M5 5
SM18-50 10 M5 5
SM70-AR-24 4 M5 5
SM400-AR-4 1 M5 5
ta ble 4 - 1
Rec om mended ca ble di am e ters and
mount ing torque