User Manual
Table Of Contents
- Unpacking and Setup:
- Good Measurement Practices:
- Front Panel Controls:
- Balance Functions and Features:
- Autocalibration – To Calibrate the balance to an
- Autocalibration – To Calibrate the Balance to a M
- Autocalibration – To Calibrate the Balance and it
- RS-232 Command for CAL 1 Using the Internal Weight:
- (CAL 2) Linearity:
- (PCS) Front Panel Parts Counting:
- (HI-OK-LO) Checkweighing / Process Control Limits:
- (%) Percent Weighing:
- () Live Animal Weighing / Severe Environment Weighing:
- Floating Tare™ \(dual range models only\):
- Capacity Tracker:
- Selectable Filters, Stability Indicator, and Send-When-Stable:
- Density Determination:
- Below Balance Weighing:
- Specific Gravity Determination:
- Security Bracket:
- Options:
- RS-232 Interface:
- Troubleshooting Guide:
- Environmental Requirements:
- CE Mark:
- Specifications for SM Semi Micro Balances:
- Specifications for SA Analytical Balances:
- Specifications for SP Precision Balances:
- Specifications for SL Laboratory Balances:
- Specifications for SG General Balances and SE Gem Balances:
- Limited Warranty:
- Extended Warranty:
- Returned Goods Procedure:
- Disposal of Electrical and Electronic Equipment:
- Quick Setup Guide:
RS-232 Interface:
Note: Scientech balances are DTE (computer) devices. If communicating to another DTE device (ie computer), a Scientech
RS-232 cable, PN10697, or normal null modem cable is required. If you are communicating to a DCE device
(printer), a serial cable is required. Scientech offers a serial cable for your convenience.
Your balance is equipped with a bi-directional RS-232 compatible interface. The interface is uni-directional (balance send only) when the
local control is set to ON. To enable the bi-directional interface, set the local control to OFF (see the RS-232 Front Panel Configuration
section). The RS-232 connector is a 9 pin D subminiature connector located on the balance rear panel. The pin out and pin descriptions
are shown in Figure 5.
Figure 5 Connector Pin Out
The balance will not output any information under any circumstances unless pin 8 (CTS) is held in an ON state (+3 to +15 VDC).
Interfaces not utilizing the CTS handshake may tie pin 8 to pin 7 or pin 8 to pin 4 to defeat it.
This interface does not strictly adhere to the official RS-232 standard. However, it emulates what has become commonplace in the
microcomputer industry.
RS-232 Specifications:
Type EIA-RS-232C, DTE
Method Half-duplex, Asynchronous Transmission, Bi-directional
Format 300, 600, 1200, 2400, 4800, 9600 and 19,200 baud rate selectable (default is 9600)
Code: ASCII
For parity Even or Odd: Data bit 7, Stop bit 1.
For parity none: Data bit 7, Stop bit 2.
Note: Parity none, Data bit 8, Stop bit 1 does function, but is not technically correct.
RS-232 Front Panel Configuration:
Before using the RS-232 feature of your balance, it is necessary to configure the interface properly for your application. The RS-232
Configuration Mode allows you to set baud rate, front panel control, software protocol, parity, auto send, continuous send, and display
blanking. The balance is configured with defaults at the factory. The default modes are shown in parenthesis below.
It is not necessary to follow the entire configuration procedure if only one parameter is to be changed. Follow Steps 1 and 2, then use
the MODE switch to select the parameter to be changed, skipping the others. Select the desired setting using the SEND switch, save
the setting using the ZERO switch, then turn the balance off and then on again with the ON/OFF switch.
To enter the RS-232 configuration mode via the front panel, proceed as follows:
User Action Balance Response
1 Turn the balance off using the ON/OFF button. Display blanks.
2 Hold the SEND button down while turning the balance on using the
ON/OFF button.
OK and the current baud rate, br300, br600,
br1200, br2400, br4800, br9600, or br19,200
is displayed.
3 BAUD RATE SELECTION (9600 is default).
Press the SEND button until the desired baud rate is displayed. Then press
the ZERO button.
OK SAVEd is displayed followed by OK and
the chosen baud rate.
14