Rotary Evaporator User Manual RE100-Pro LCD Digital Rotary Evaporator Please read the User Manual carefully before use, and follow all operating and safety instructions! www.scilogex.
Contents Preface ................................................................................. 1 Service ................................................................................. 1 Warranty .............................................................................. 1 1. Safety Instructions ....................................................... 2 2. Proper Use ................................................................... 5 3. Inspection ..................................................
Rotary Evaporator Preface Please provide the customer care representative with the Welcome to the “Rotary Evaporator User Manual”. Users should read this Manual carefully, follow the instructions and procedures, and be aware of all the cautions when using this instrument. following information: • Serial number(on the rear panel) • Certification • Description of problem (i.e.
Rotary Evaporator 1. Safety Instructions • Read the operating instructions in full before starting up and follow the safety instructions. • Keep the operating instructions in a place where they can be accessed by everyone. • • Ensure that only trained staff work with the appliance. Follow the safety instructions, guidelines, occupational health and safety and accident prevention regulations.
Rotary Evaporator • Do not operate the appliance in explosive atmospheres, with hazardous substances or under water. • Only process media that will not react dangerously to the extra energy produced through processing. This also applies to any extra energy produced in other ways, e.g. through light irradiation. of the distillation equipment. The cooler must work properly. Monitor the coolant flow rate at the cooler outlet.
Rotary Evaporator for stress relief. • • • • mechanism to lift the evaporator piston out of the heating bath. Evacuate the danger zone and warn those in the surrounding area! CAUTION! Avoid peroxide formation. Organic peroxides can accumulate in distillation and exhaust residues and explode while decomposing! Keep liquids that tend to form organic peroxides away from light, in particular from UV rays and check them prior to distillation and exhaust for the presence of peroxides.
Rotary Evaporator supply network if the device power switch is off or the plug is pulled out. • The socket for the mains cord must be easily accessible. • The voltage stated on the type plate must correspond to the mains voltage. • • Socket must be earthed (protective ground contact). Removable parts must be refitted to the appliance to prevent the infiltration of foreign objects, liquids etc. • Protect the appliance and accessories from bumps and impacts.
Rotary Evaporator 3. Inspection 4. Control and Display 3.1 Receiving Inspection 4.1 Control elements Unpack the equipment carefully and check for any damages which may have arisen during transport. Please contact manufacturer/supplier for technical support. Motor Condenser Note: If there is any apparent damage to the system, please do not plug it into the power line. Evaporating flask Receiving flask Heating bath Operation panel 3.
Rotary Evaporator Items LED power Lift-up key Lift-down key LCD display Int key Speed key Temperature key Speed key Speed Push the key and then rotate control knob to set the rated speed in the range from 20 to 280. Temperature key Temp Push the key and then rotate control knob to set the rated temperature in the range from room temperature to 180°C. Timer key Timer Push the key and then rotate control knob to set the rated time in the range from 1 to 999min.
Rotary Evaporator the key the lift stops and remains in that position. LCD display LCD displays the real working state. Display LED power Power ON, LED power flash. When the unit running, LED power is lit. Remote Power switch Power ON or OFF. Probe Table 2 4.2 Display Remote control External probe Rotate Temperature Temperature symbol Timer Interval Timer display area Interval display area Display when using external probe. Display when the rotating function is switched ON.
Rotary Evaporator value. Figure 6 Temperature symbol Display when the heating function is switched ON. Timer display area When setting timer, displaying the rated time. When the timer function is switched ON, displaying remaining time. CAUTION! Loosen transportation lock (Figure 7). • Displaying the “Int” target value. Hold the lift with your hand on the height position and remove the thumb screw on the back of the appliance (a).
Rotary Evaporator • Then lock the locking device by turning it clockwise by 60°. • The plastic screw nut (f) helps loosen tight-fitting piston ground-in connections (Figure 9). Figure 7 • • • Remove the clamping device for the angle setting of the rotation drive on the right side of the lift by rotating the knurled screw (d) counter clockwise. - Set the drive at an angle of approx.
Rotary Evaporator device at 200 rpm for 2 min. Then tighten the cap nut firmly. g h i Figure 10 Figure 12 h Figure 11 • Figure 13 Installing condenser (g) (Figure 12 and Figure 13). - Load the cap nut (h) and locking spring (i) to condenser in turn. - • Tighten the cap nut hand-screwed (h). Start the 11 Installing the vertical glassware condenser locking device (Figure 14).
Rotary Evaporator • Connecting tubes (Figure 15). Vacuum tube Cooling water Cooling water Figure 14 Figure 15 5.2 Operation 12 • Make sure the required operating voltage and power supply voltage match. • • Ensure the socket must be properly grounded. Plug in the power cable,ensure the power is on and begin initializing. • Press Lift-up/Lift-down key, the lift goes up/down.
Rotary Evaporator Release Lift-up/Lift-down key, the lift stops. • 6. Faults Push the speed key and then rotate control knob to set the rated speed. • • Push the temperature key and then rotate control knob to set the rated temperature. Instruments can’t be power ON - Check whether the power line is unplugged - Check whether the fuse is broken or loose • Push the timer key and then rotate control knob to set the rated time. Deactivate the timer by setting the target value to “0”.
Rotary Evaporator 7. Maintenance and Cleaning • Proper maintenance can keep instruments working properly and lengthen its lifetime. • Do not spray cleanser into the instrument when cleaning. • • • Unplug the power line when cleaning.
Rotary Evaporator 8. Associated Regulations Standards interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Rotary Evaporator 9.
SCILOGEX, LLC 500 Four Rod Road Suite 122 Berlin, CT 06037 USA Tel: 1-860-828-5289 Fax:1-860-828-5389 E-mail: info@scilogex.com Http://www.scilogex.