User manual
50
11. Close the cylinder and screw in the sealing plug by hand. The sealing plug should only be tightened by 
hand on a cylinder pack that is to be lled.
1
2. Allow the cylinder packs to dry.
13. Shake off the water from the other parts and allow them to air dry. Dry thoroughly.
14. Fit the breathing valve into the full-face mask. Turn the breathing valve counterclockwise to attach the 
breathing valve bayonet coupling. Attach the xing cover.
1
5. Remove the cylinder pack for lling. Inform the air lling site if the cylinder pack has been subjected 
to exceptional stress or if the pressure in the cylinder falls below 30 bar (435 psi). On a lled cylinder 
pack, the sealing plug should be tightened with an open-ended wrench (23 mm).
Comment:  The cylinder pack should not be completely emptied. There must always be a residual pressure 
of at least 10 bar (145 psi) in the air cylinder to prevent moisture from nding its way through the valve.
Check the check valve's tightness.
The tightness of the check valve must be checked during the monthly cleaning routine. 
1
.  Connect the breathing valve to the intermediate pressure hose on the regulator.
2.  Remove the breathing valve from the full-face mask, alternatively, remove the mouthpiece from the 
octopus breathing valve (see “Replace mouthpiece” in Section 12.4.4 “Breathing valve”).
3
.  Open the cylinder valve.
4.  Fill the space at the check valve disc with water. 
5.  Ensure that water is not leaking into the breathing valve. The water level over the check valve must not 
decrease. If there is a leak, the check valve must be cleaned or replaced.
6
.  Blow the valve dry using the vent button. Check that the check valve disc is even after dry-blowing.
7.  Close  the  cylinder  valve  and  insert  the  breathing  valve in  the  full-face  mask  or  replace  the  mouth-
piece.
Disinfecting the full-face mask/breathing valve
The full-face mask/breathing valve must be disinfected during the monthly cleaning procedure or when 
the need arises. 
Interspiro recommends that full-face masks and breathing valves used by several persons be cleaned and 
disinfected after each use.