User manual
52
4.  When  the  exhalation  unit  has  been  removed, 
press the threads to remove the diaphragm unit 
from the exhalation unit. Place your hand over the 
diaphragm unit to prevent it from getting lost.
5
.  Clean  the  spring,  cover,  diaphragm  unit  and 
mouthpiece  in  water.  Carefully  pull  the  dia-
phragm unit so that the diaphragm is stretched 
and the dirt on the surface can be removed. Make 
sure that the diaphragm unit and the loose parts 
are not damaged.
6
.  There is a white plastic sealing disc on the diaphragm 
unit. The sealing disc must be removed and cleaned. 
Clean the inner parts of the diaphragm unit. Clean 
the inside of the diaphragm unit's black sealing ange 
that is in contact with the sealing disc. Make sure that 
the diaphragm does not have any holes or damage 
from wear. If the diaphragm needs to be replaced, 
this must be done by a certied service technician.
7
.  After washing and rinsing, the parts are disinfected by submerging them in a solution consisting of 0.1 % 
chlorhexidine and the rest water for one hour.
8
.  Clean the parts carefully in clean water and allow them to air dry.










