User manual
53
Assembly 
1.  Replace the sealing disc in the diaphragm unit and 
then place the sealing disc with spring in the cover 
and press the diaphragm unit together.  Check that 
the spring ts into the recess for the sealing disc.
2
.  Set  the  inhalation unit  in  the  breathing  valve and 
screw the valve housing together with the locking 
ring.
Comment:  Only the locking ring should be turned. Otherwise there is a risk that the O-ring will be moved 
causing a leak.
3.  If the full-face mask is used, check that the O-ring in the bayonet coupling is not damaged and that it is 
placed in the O-ring groove. Information about replacing the O-ring can be found under “Replacing an 
O-ring (only full-face mask) in Section 12.4.4 
  ”
Breathing valve”.
4.   If a full-face mask is used, connect the breathing 
valve in the bayonet coupling by turning the breath-
ing valve counterclockwise. Mount the xing cover 
on the full-face mask and tighten the two screws.