User's Manual

Page 11 of 40
P/N 595166-01 Rev A 1/09
DRAFT COPY
LOW BATTERY
In a low battery condition, the SEMS II distress alarm will produce a single
audible chirp from the sensor module once every two (2) seconds and the
green light on the control module will not ash.
While in low battery condition, the SEMS II distress alarm will continue to
operate for a period of time greater then the longest duration cylinder avail-
able for the respirator. However, the batteries must be replaced before the
respirator is used again. See the BATTERY REPLACEMENT section of these
instructions.
If batteries are completely discharged or have not been installed, there will
be no light or sound and the unit will not operate.
BATTERY TEST
When the SEMS II PDA distress alarm is in the off condition (cylinder valve
closed with no ashing lights, the batteries in the entire system can be checked
by depressing and holding the RESET button on the console.
• A GREEN LED will illuminate on the Control Console, if there is suf-
cient battery power remaining,
• A RED LED indicates that the batteries are low and must be replaced
before the respirator is to be used again.
If a low battery message occurs, ALL batteries must be changed before the
respirator is used again. See the BATTERY REPLACEMENT section of this
instruction for details.
WARNING
DO NOT USE A RESPIRATOR IN A LOW BATTERY
CONDITION. FAILURE TO REPLACE THE BAT-
TERIES AND/OR CONTINUING WITH MULTIPLE
USES OF THE RESPIRATOR AFTER THE LOW
BATTERY CONDITION HAS BEEN INDICATED
BY THE SEMS II DISTRESS ALARM MAY RESULT
IN FAILURE OF THE SEMS II DISTRESS ALARM
DURING USE AND POSSIBLE INJURY OR DEATH
OF THE USER.