User's Manual

Page 11 of 28
P/N 595277-01 Rev E 10/17
WARNING
FAILURE TO REPLACE THE BATTERIES AND/
OR CONTINUING WITH MULTIPLE USES OF
THE RESPIRATOR AFTER THE LOW BATTERY
CONDITION HAS BEEN INDICATED BY THE
PAK-ALERT DISTRESS ALARM MAY RESULT IN
FAILURE OF THE PAK-ALERT DISTRESS ALARM
DURING USE AND POSSIBLE INJURY OR DEATH
OF THE USER.
LOW BATTERY
In a low battery condition, the PAK-ALERT distress alarm will produce a single
audible chirp from the Sensor Module once every two (2) seconds and the
green light on the control module will not ash.
In low battery condition, the PAK-ALERT distress alarm will not emit the 3
beeps when cylinder valve is activated.
While in low battery condition, the PAK-ALERT distress alarm will continue
to operate for a period of time greater then the longest duration cylinder
available for the respirator. However, the batteries must be replaced before
the respirator is used again. See THE BATTERY REPLACEMENT section
of these instructions.
If batteries are completely discharged or have not been installed, there will
be no light or sound and the unit will not operate.
BATTERY TEST
To test the batteries, verify that the PAK-ALERT distress alarm is in the off
condition (cylinder valve closed with no ashing green LED on the control
console). Press and hold the reset button on the console.
A green light on the console will illuminate to indicate sufcient battery power
remaining. If a red light appears, the batteries must be replaced before the
respirator is used again.
TO TURN OFF THE PAK-ALERT DISTRESS ALARM
When use of the respirator with the PAK-ALERT distress alarm is no longer
required, close the cylinder valve on the respirator and vent the residual air
from the respirator system by opening the regulator purge valve. After all the air
ow stops, close the regulator purge valve and press the reset button twice
to turn off the PAK-ALERT distress alarm. If there is air pressure left in the
system, the green ashing light will continue to ash while a fteen second
beep sequence is heard from the Sensor Module as the residual air bleeds
from the system. As soon as the air has completely bled from system, the unit
will sound a quick two tone chirp and the PASS DEVICE distress alarm will
be inactive. If there is no pressure in the system when the RESET button is
pressed twice, there will be no fteen second beep sequence. When the unit
sounds a quick two tone chirp, the PAK-ALERT distress alarm is inactive. If
there is any air pressure left in the system, the PASS DEVICE distress alarm
will return to the active mode.
If the respirator cylinder valve is open and/or pressure remains in the respira-
tor, the PAK-ALERT distress alarm can not be turned off. Pressing the reset
button when the respirator is pressurized will only reset an alarm condition
and return the PAK-ALERT distress alarm to automatic mode.
If the respirator cylinder is turned off and depressurized without pressing the
reset button twice, the PAK-ALERT distress alarm will continue to monitor
motion in automatic mode. This means that the PAK-ALERT distress alarm
may be used to monitor motion after the respirator is turned off and depres-
surized. Resetting the full alarm after the respirator has been depressurized
will not turn off the PAK-ALERT distress alarm. Press the reset switch twice
with no alarm condition to turn off the PAK-ALERT distress alarm (there will
be no fteen (15) second beep sequence and two tone chirp will be heard).
DRAFT
NOT FOR DISTRIBUTION