User's Manual

Connecting the logger and installing it on-site
23
4.1.3 Connecting sensors
Any connection of a sensor is established via the VK86 connection cable attached to the
IN socket
.
One end of the connection cable is equipped with a box providing a terminal block the
colour-coded wires coming from the logger are connected to. Each colour represents a
specific input / output of the logger.
The VK86 connection cable has the following pin assignment:
Wire Description
white
Channel 1 +
Channels up to 4 sensors can be connected to.
brown
Channel 1 –
green
Channel 2 +
yellow
Channel 2 –
grey
Channel 3 +
pink
Channel 3 –
blue
Channel 4 +
red
Channel 4 –
black
not used
purple
not used
To connect a sensor, lead its connecting wires into the box using one of the four cable
glands and connect them to the respective terminal of the terminal block.
When you connect the cable to the IN socket of the logger, make sure that the guide on
the plug fits in the groove in the socket and that you feel the plug latch in.
NOTE
When you install the measuring system please keep in mind that water and
dust can penetrate the box.
green