Product Manual

Instructions for use – Additional Information • 1
English
17-10-07-659-002_02-2020S
INSTRUCTIONS FOR USE ADDITIONAL INFORMATION
This document contains error corrections and additional troubleshooting
information for the devices listed below:
seca 651/650
seca 655/654
seca mBCA 555/554, seca mBCA 552
Factory settings
Functions/device mode (seca mBCA 555/554, seca mBCA 552 only)
TROUBLESHOOTING
In case of any error messages not mentionend in the instructions for use or
in the list below, please contact
seca Service.
Measuring weight
Bioimpedance measurement (seca mBCA 555/554, seca mBCA 552 only)
Function Factory setting
Filter
Low Medium
Function Device mode
Basic Advanced Expert Service
Measure bioimpedance
Fault/error code Cause Remedy
016-272XX-XXX
Maximum load exceeded Ask patient to step off the scale
024-272XX-XXX
No GAL value set Contact seca Service
Fault/error code Cause Remedy
007-276XX-XXX
BIA board needs readjustment Contact seca Service
013-276XX-XXX
to
018-276XX-XXX
Invalid measurement due to
interference
Remove or switch off nearby electronic
devices and repeat measurement
If error reoccurs, contact seca Service
BIA board defective Contact seca Service
020-276XX-XXX
Invalid measurement detected
Restart measurement
If error reoccurs, contact seca Service
021-276XX-XXX
Impedance of electrodes too high
Spray patien‘s skin with electrode spray at
the contact points
If error reoccurs, contact seca Service
030-276XX-XXX
Patient's hands or feet not positioned
correctly.
Make sure that the patient's hands or feet
are positioned correctly
If error reoccurs ,contact seca Service
034-276XX-XXX
Self-test not successful
Restart system
Disconnect the plug-in power supply unit
from the socket
Wait approx. 1 minute
Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
If error reoccurs, contact seca Service
036-276XX-XXX
Temperature sensor out of range Contact seca Service
039-276XX-XXX
Patient's hands or feet not positioned
correctly.
Impedance of electrodes too high
Make sure that the patient's hands or feet
are positioned correctly
Spray patien‘s skin with electrode spray at
the contact points
Electrodes defective Contact seca Service