SECTION 2 POLICY AND PROCEDURES
Alarm Procedures
1. If Exit System is in alarm:
Identify the location of the alarm shown on the nurse station display or computer monitor.
Follow the hospital Policy and Procedure.
2. If one or more cutband zones are in alarm with an infant’s name and room number:
Go to the area of the alarm condition and locate the baby.
Follow the hospital Policy and Procedures.
3. If the transmitter has not been cut off the infant or the strap looks intact:
Verify that the transmitter strap is not too tight on the infant
Excess strap has been secured to transmitter case with tape.
4. If the transmitter passes the visual test and the system is still reporting a cutband alarm:
Remove the transmitter by cutting the strap.
Replace the strap and apply the transmitter to the infant.
Verify that the alarm has been silenced.
5. If the above steps do not resolve the alarm
Put the transmitter to sleep and remove it from the infant
Place a new transmitter on the infant
Verify the alarm has been silenced
Contact Secure Care Products LLC