APPENDIX 1: SHIPPING RETURN PROCEDURE
Cleaning of the Silaris Dock must be completed prior to the return to Sekisui Diagnostics. Cleaning must be completed according the
instructions provided. Please complete the information below and attach the completed record to the dock prior to return.
Please complete the cleaning procedure following the instructions below:
8
. Unplug the power cord from the wall outlet and Dock.
9. Close the Dock lid.
10. Using a cleaning solution (10% bleach or 70% ethanol), or a germicidal disposable wipe (such as CaviWipes™), gently wipe the
o
uter surfaces of the Dock, removing any debris. Do not press the wipe against the open vents of the back or sides of the Dock.
11. Using a new cloth, wipe the front of the Dock twice top to bottom, then twice left to right. Follow this procedure for the back, top
and bottom of the Dock.
12. Do not let liquid accumulate near any opening. Make sure no liquid enters the dock.
13. Leave the Dock “wet” for at least 3 minutes.
14. With a fresh dry cloth (such as a Kimwipe™) remove any residual moisture.
SILARIS DOCK INFORMATION
Dock Lot Number:
CONTACT & FACILILTY INFORMATION
Contact Name: Title:
Contact Phone: Email:
Facility Name:
Street Address:
City, State, Zip Code:
DECONTAMINATION INFORMATION
The Silaris Dock has been cleaned as described above. The dock surfaces have been wiped with an appropriate disinfecting
solution and left for at least 3 minutes.
Performed By:
Signature Date
Please call Sekisui Diagnostics Technical Support at (800) 332-1042 for further assistance.  
DOCUMENTATION FOR COMPLETION OF CLEANING PROCEDURE
Complete the DOCUMENATION FOR COMPLETION OF CLEANING PROCEDURE below, and return this document along with
the Dock, AC Power Cord, AC Adaptor using the shipping materials provided.