Operating instructions
FC Operation & Maintenance Guide (20 & 25 Amp)
13
Test #2: With one or both leads disconnected from the inductor, measure the resistance across the
inductor terminals. If the resistance is less than 0.25 ohms, the inductor is OK.
Test #3: Disconnect wire connected to the diode. Using an ohmmeter on the RX1 scale, a good
diode will cause the ohmmeter to read approximately 1/2 scale when the leads are
connected one way, and infinity when the leads are connected in reverse. A similar
reading in both directions indicates a defective diode.
Test #4: Disconnect all the leads to the SCR and its heat sink. Connect a voltmeter across the
1K? resistor to measure the voltage drop. With the battery connected as shown, V
drop
should read approximately 2.3 volts (V
source
- 0.7 volts). Remove the voltage source
to the gate, but keep it connected to the 1K? resistor and cathode. V
drop
should equal
zero. Reconnect the gate and reverse the battery polarity. V
drop
should equal zero.
Readings other than these indicate a defective SCR. See Figure 6.3
Test #5: Due to the modest cost of the control circuit, we recommend that the entire unit be
replaced rather than attempting to repair it. If the troubleshooting guide has not revealed
any defective components (tests #1-4), the control circuit should be replaced as a unit.
Test #6: Remove all wires from the current shunt. Place a milliohmmeter across the two terminals.
The following formula should be used to determine the correct resistance:
Resistance = 0.5/Output current rating of the charger
If the resistance is more than 20% too low, the current shunt should be replaced.
FIGURE 6.3
SCR test setup
25 ž
approx.
3 volt source(2 "D"
batteries in series)
Cathode
Gate
SCR
Anode
-
+
+
-
V
drop
1 Kž










