Eversense CGM User Guide
19
Overall, 85.4% of CGM readings fell within zone A. This indicates CGM readings were in close agreement with
reference values for the great majority of readings.
Consensus Error Grid Accuracy by Glucose Level
Zone Frequency Percent
A 8922
85.4%
B
1481
14.2%
C
39
0.4%
D
0
0.0%
E
0
0.0%
Total
10442
100.0%