User's Manual

85
Eversense CGM User Guide
9
Alert Descriptions and Actions (continued)
Alerts
App DisplayDescription
Low Sensor Temperature*
Appears every 20 minutes when the sensor
temperature is too low.
No glucose reading is displayed until the sensor
temperature is within normal operating conditions.
High Sensor Temperature*
Appears every 20 minutes when the sensor
temperature is too high. No glucose reading is
displayed until the sensor temperature is within
normal operating conditions.
Actions
Go to a warmer environment to increase the sensor
temperature. Keep your smart transmitter turned
on so you will start receiving glucose values when
the sensor temperature is between 26 - 40 °C
(81 - 104 °F).
Go to a cooler environment to reduce the sensor
temperature. Briefly remove the smart transmitter
while the sensor temperature cools to between
26 - 40° C (81 - 104 °F). Then put the smart
transmitter back on to start receiving glucose values
again from the sensor.