Owner's manual

SS869 8/12/08 Rev B
Apply power to calibrator with AC adapter or charge battery.
Turn on power switch located in the top right corner. Power is on when green light illuminates - the pocket PC will flash an orange light.
Push Pocket PC power button in top right of the unit. It will display a “start” screen. Tap “start” and select Setra Calibrator icon from “last
programs run” or “programs” file from drop down menu.
Calibrator starts in “Real Time” mode. Tap UUT (Unit Under Test) tab and select a stored UUT ID that corresponds to the pressure sensor
you wish to test - (tap the down arrow in the top box to view all of the stored UUT test profiles). Power is supplied to test pressure sensor
when UUT ID is selected. If a needed ID is not found, see Creating and Modifying UUT ID’s below.
Plug in pressure tubing and electrical cables into marked sockets on the top panel. Connect pressure tubing to “high” and “low” pressure
fittings. Hook up connections as follows:
1. 4-20mA current loop operation, connect +EXC to red 869 banana jack and -EXC to black jack.
2. For 3-wire voltage output operation, connect +EXC to red electrical jack, -EXC common to black jack, and +Output to green
jack.
Return to “Real Time” screen to verify correct zero pressure output of test pressure sensor. Tap “tare” button to re-zero onboard pressure
references (approx 10 seconds). You can also check full-scale output by tapping “+FS” (+ full scale) button. You can also make zero and
span adjustments to test pressure sensor at this time.
Go to “Test” screen and tap “Do Test” button to initiate automatic calibration test. The calibrator will apply test pressures and record output
results until “test complete” flag is displayed. You can save the test data by tapping the “Save Test Results” button. The data is saved with
Micro-Cal
®
Model 869 Quick Start Guide
To create a new UUT ID go to the UUT screen and tap “new” button. Next, type in a new UUT ID in the upper box. Enter data by tapping
the keyboard icon in lower right of screen. Characters are entered on screen at the cursor when mini-keyboard characters are tapped. An
existing ID template must be selected in “copy from” box. It is suggested that an existing ID template be chosen that closely matches the
pressure sensor desired to test either in pressure range or output type. Tap “ok”. When screen returns to UUT, any ID values can be modi-
fied using keypad. Use dropdown boxes to change excitation, output type, or number of test points. Hit “save” button and new test profile is
ready to use.
To modify a UUT ID, select desired UUT ID from dropdown screen, make changes, and tap “save”. The newly created or modified UUT IDs
will be available from dropdown UUT ID screen.
Creating and Modifying UUT ID’s
I
II
III
IV
V
VI
VII
Setra Systems, Inc., 159 Swanson Rd, Boxborough, MA, 01719