2014 PRODUCT CATALOG ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
DIFFERENTIAL PRESSURE INTRODUCTION S etra is a leading manufacturer of a broad portfolio of pressure transducers, humidity transmitters, current switches and current transducers. The company was founded in 1967 by Dr. S.Y. Lee and Dr. Y.T. Li, former Professors of Engineering at the Massachusetts Insititute of Technology.
Setra’s capacitive pressure transducers are expertly designed adaptations of a simple, durable and fundamentally stable device...the electrical capacitor. In a typical Setra configuration, a compact housing contains two closely spaced, parallel, electrically isolated metallic surfaces, one of which is essentially a diaphragm capable of slight flexing under pressure. The diaphragm is constructed of a low-hysteresis material such as 17-4 PH SS or a proprietary compound of fused glass and ceramic (Setraceram).
DIFFERENTIAL PRESSURE Absolute Pressure — Pressure measured relative to full vacuum. Referred to as pounds per square inch absolute (PSIA). Atmospheric Pressure — Pressure of the atmosphere at the earth’s surface NIST standard atmospheric pressure = 1.01325 bar. BAR — Unit of pressure (or stress). 1 bar = 750.07 mm of mercury at 0°C, at 45°. Barometric Pressure — Atmospheric pressure, often measured in millibars, in Hg (inches of mercury), or hectopascals.
2 TECHNOLOGIES 3 TERMINOLOGY & DEFINITIONS PRODUCT SECTION 1.1 4 Differential Pressure Transducers Model 260 8 Model 264 10 Model 265 12 Model 267 / 267MR 14 Model 269 18 Model 230 20 Model 231 24 Model 231RS 28 Model 239 32 PRODUCT SECTION 2.1 Room Pressure Monitors Model MRMS 36 Model SRPM 38 Model SRCM 42 Model SRIM 46 Model SRMD 48 PRODUCT SECTION 3.
DIFFERENTIAL PRESSURE 6 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
DIFFERENTIAL PRESSURE PRODUCT SECTION 1.1 DIFFERENTIAL PRESSURE MODELS: 260 264 267/267MR 269 231 231RS 265 230 239 ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
DIFFERENTIAL PRESSURE Model 260 Very Low Differential Multi-Configurable Pressure Transducer FEATURES ■ Optional 4 Digit LCD ■ Field Selectable Multi-Range ■ Field Selectable Multi-Output ■ Simple 5-Step Setup ■ Field Accessible Push-Button Zero and Span ■ Hinged Cover ■ External Mounting Tabs ■ Unregulated AC/DC Operation ■ Microprocessor-Based Electronics NOTE: Setra quality standards are based on ANSI-Z540-1. The calibration of this product is NIST traceable. U.S. Patent nos.
DIFFERENTIAL PRESSURE Model 260 Very Low Differential Multi-Configurable Pressure Transducer DIMENSIONS 4.75 121 Optional Static Probe Diagram 4.25 108 4.15 105 1500 9.3 237 O0.862 22 1.25 32 0.59 15 1.49 38 7.7 196 1.50 38 0.63 16 0.51 13 IN MM 1.3 32 1.03 26 0.
DIFFERENTIAL PRESSURE Model 264 Very Low Differential Pressure Transducer FEATURES ■ Up to 10 PSI Overpressure (Range Dependent) ■ Installation Time Minimized with Snap Track Mounting and Easy-To-Access Pressure Ports and Electrical Connections ■ 0 to 5 VDC or 2-wire 4 to 20 mA Analog Outputs Are Compatible with Energy Management Systems ■ Reverse Wiring Protection ■ Internal Regulation Permits Use with Unregu- Model 264 w/ Conduit Cover Option lated DC Power Supplies ■ Fire Retardant Case NOTE: Setr
DIFFERENTIAL PRESSURE Model 264 Very Low Differential Pressure Transducer DIMENSIONS Code T1 Electrical Termination Dimensions #6 SCREW WIRE CLAMP 3 PLACES 2.750 69.85 Optional 1/2” Conduit Electrical Enclosure Dimensions 0.078 1.98 0.312 7.92 1.662 42.21 PRESSURE PORTS 0.385 9.78 1.187 30.15 1.19 30.15 1.19 30.15 4.562 115.87 5.062 128.57 PRESSURE PORTS 0.594 15.08 0.49 12.55 4.56 115.87 IN MM IN MM 0.99 1.66 1.91 42.21 48.51 3.00 76.20 25.17 1.624 1.811 41.25 45.99 1.50 38.10 5.
DIFFERENTIAL PRESSURE Model 265 Very Low Differential Pressure Transducer FEATURES ■ Up to 10 PSI Overpressure (Range Dependent) ■ 24 VDC or 24 VAC Excitation ■ High Level 0 to 5 VDC, 0 to 10 VDC or 2-wire 4 to 20 mA Analog Outputs Are Compatible with All Energy Management Systems ■ Full Protected Against Reverse Wiring ■ 1% Accuracy Improves VAV Performance ■ Optional Accuracies up to 0.
DIFFERENTIAL PRESSURE Model 265 Very Low Differential Pressure Transducer DIMENSIONS Code T1 Electrical Termination Dimensions Optional A1 Conduit Electrical Enclosure Dimensions 6-32 SCREW W/TERMINAL WASHERS 3 PLACES 1.64 42 1.89 3.16 48 80 φ 0.156 φ 4 1.94 49.3 MOUNTING HOLES 2 PLACES 0.385 10 1.67 42 φ .875 1.00 25 (FOR 1/2” CONDUIT FITTING) 1.56 40 IN MM 1.41 36 0.80 20 1.10 28 3.34 85 2.19 56 2.74 70 0.61 16 4.
DIFFERENTIAL PRESSURE Model 267/267MR Very Low Differential Pressure Transducer FEATURES ■ Model 267MR Offers Multi-Range Capability, 6 Model 267MR - Multi-Range Field Selectable Ranges via Dip Switches, and Field Selectable 0-5 or 0-10 VDC Output ■ Model 267 Offers an Optional 3 1/2 Digit LCD Display with a 0.5% FS Standard Accuracy Model 267 w/ Display Option ■ NEMA 4/IP65 Rated Housing ■ Optional Accuracies as High as 0.25% FS NOTE: Setra quality standards are based on ANSI-Z540-1.
Very Low Differential Pressure Transducer SPECIFICATIONS Performance Data Standard Optional Accuracy RSS1 (at constant temp) ±1.0% FS ±0.4% FS ±0.25% FS Non-Linearity, BFSL ±0.98% FS ±0.38% FS ±0.22% FS Hysteresis ±0.10% FS ±0.10% FS ±0.10% FS Non-Repeatability ±0.5% FS ±0.5% FS ±0.
DIFFERENTIAL PRESSURE Model 267/267MR Very Low Differential Pressure Transducer D-SUB ELECTRICAL TERMINATION PIN 1 + Output PIN 4 + EXC CONNECTION PINS 8, 9* Common + Excitation 4 + Output 1 Common 8, 9 Excitation 9 to 30 VAC/ 11.
Very Low Differential Pressure Transducer ORDERING INFORMATION (Model 267MR) 2 6 7 1 Model Range Code 2671 = 267 See Table 1 Below Output Pressure Fitting/Elec. Termination Accuracy (Full Scale) Display 11 3/16” Barbed Brass Fitting Std. 2D 0-5 VDC Std. G1 PG-13.5 Strain Relief Opt. 2E Std. G2 PG9 Strain Relief Std. D9 9 pin D-Sub Conn. Std. A1 1/2” Conduit Opening 4-20 mA 0-10 VDC 1/4”NPTF Brass Fitting Opt. 1K PG-9 Strain Relief Opt. 2K PG-13.5 Strain Relief Opt.
DIFFERENTIAL PRESSURE Model 269 Very Low Differential Pressure Transducer FEATURES ■ Installation Time Minimized with DIN Rail Mounting and Easy- To-Access Pressure Ports and Electrical Connections ■ Removable Process Head Eliminates the Need to Cut Tubes for Easy Installation ■ Detachable Terminal Block so Field Wiring Can Remain In-Situ During Calibration ■ Secure Calibration Key for Making Zero and Span Adjustments ■ 2-wire 4 to 20 mA Analog Outputs Compat- NOTE: Setra quality standards are based
DIFFERENTIAL PRESSURE Model 269 Very Low Differential Pressure Transducer DIMENSIONS Optional Display Location Din Rail Version Base Mount Version 0.48 12 Optional Display Location 0.75 19 Security Cal Key Detachable Process Head Detachable Terminal Block Cal Cal Button Encoder 0.48 12 0.75 19 Security Key Accessory PN 269425-02 Needed to Calibrate Transducer 5.63 143 3.90 99 For 35mm Din Rail (Meeting EN-50022 Specifications) IN MM 2.90 73 3.40 86 1.29 32 2.
DIFFERENTIAL PRESSURE Model 230 Wet-to-Wet Pressure Transducer FEATURES ■ Ideal for Applications with Line Pressure up to 350 psig ■ NEMA 4/IP65 Rating ■ No Liquid Fill Diaphragm ■ Available with 3-Valve or 5-Valve Manifold As- sembly Option ■ Low Line Pressure Effect NOTE: Setra quality standards are based on ANSIZ540-1. The calibration of this product is NIST traceable.
Wet-to-Wet Pressure Transducer SPECIFICATIONS Performance Data Accuracy RSS (at constant temp) 1 ±0.25% FS Non-Linearity, BFSL ±0.20% FS Hysteresis 0.10% FS Non-Repeatability 0.05% FS Physical Description (Model 230) Electrical Data (Voltage) Case Stainless Steel/Aluminum Circuit 3-Wire (Exc, Out, Com) Electrical Connection Barrier strip terminal block with conduit enclosure & 0.875 Excitation 9 to 30 VDC for 0-5 VDC Output Output7 0 to 5 VDC8, 0 to 10 VDC8 100 ohms DIA conduit opening.
DIFFERENTIAL PRESSURE Model 230 Wet-to-Wet Pressure Transducer WIRING Voltage Transducer • Power Supply + — Current Transmitter — Power Supply + Output Signal (Voltage) + Output Signal (4 to 20 mA) — + — DIMENSIONS (3-Valve Manifold Assembly) 3-Valve Manifold Assembly Description 7.05 179 (Order as Pressure Code Fitting “3V”.) 4,00 102 2.33 59 MODEL 230 DIFFERENTIAL PRESSURE TRANSDUCER V3 SHUNT VALVES 7.83 199 SHUT OFF VALVES V3 5.
Wet-to-Wet Pressure Transducer DIMENSIONS (5-Valve Manifold Assembly) 7.05 179 5-Valve Manifold Assembly Description 2.33 59 4,00 102 (Order as Pressure Code Fitting “5V”.) DIFFERENTIAL PRESSURE Model 230 7.83 199 V3 V4 5.80 143 5.63 179 V5 V2 V6 V7 Model 230 Differential Pressure Transducer SHUNT VALVE 0.38 10 V3 V4 V1 3.25 83 V5 Low Process/Commission 1/4”NPT Connection V2 SHUT OFF VALVES High Process Connection 1/4”NPT 0.156 4 0.
DIFFERENTIAL PRESSURE Multi-Sense® Model 231 Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer FEATURES ■ Field Selectable Output - True 4 to 20 mA, 0 to 5, 1 to 5, and 0 to 10 VDC ■ Field Selectable Pressure Ranges ■ Field Accessible Push-Button Zero and Re- mote Zero ■ Dual Sensors ■ Optional 3- or 5-Valve Manifold ■ Hinged Cover NOTE: Setra quality standards are based on ANSIZ540-1. The calibration of this product is NIST traceable. U.S. Patent nos.
Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer DIMENSIONS WIRING 6.0 152 5.6 141 R 0.1 2 0.3 9 DIFFERENTIAL PRESSURE Multi-Sense® Model 231 2.0 51 1.2 31 1.7 42 + Controller Voltage + Relay 5.2 131 3-Wire - Voltage Output 0 to 5 VDC 0 to 10 VDC 1 to 5 VDC Remote Zero 1/8” NPTF 1.6 40 High DC/AC Power Supply 1.0 24 Low 1/2” Conduit Opening 2.0 51 DC Power Supply + Controller Current + (4-20 mA) Relay 4.0 102 4.6 117 4.
DIFFERENTIAL PRESSURE Multi-Sense® Model 231 Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer DIMENSIONS (3-Valve Manifold Assembly) 2.00 51 3-Valve Manifold Assembly Description (Order as Pressure Code Fitting "3V".) Manifold Block Valves (3) Valve type Brass V1 for connection to +port V2 for connection to -port V3 for equalizing pressure 90 Degree On/Off Process Connections 1/4” -18 NPT Internal Thread MODEL 230 DIFFERENTIAL PRESSURE TRANSDUCER 7.
DIFFERENTIAL PRESSURE Multi-Sense® Model 231 Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer INSTALLATION Model 231 High Pressure Remote Sensor Low Pressure Remote Sensor Valve A1 High Valve B1 Low Valve A = High Side Valve Valve B = Low Side Valve Pipe Pipe 1. Valves not included. Pressure Range Code Selector (IMportant: READ BEFORE ORDERING) Examine the pressure application and determine what is the Highest System Line Pressure.
DIFFERENTIAL PRESSURE Multi-Sense® Model 231RS Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer 231RS Armored Cable Version 231RS Cable Version FEATURES ■ Wet-to-Wet Transducer w/ Remote Sensors ■ Conduit and Cable Versions ■ Field Selectable Output - True 4 to 20 mA, 0 to 5, 1 to 5, and 0 to 10 VDC ■ Each Unit Provides 4 Unidirectional and 4 Bi- directional Switch Selectable Pressure Ranges ■ Field Accessible Push-Button Zero and Re231RS Conduit Version mote Zero ■ Jumper Selectable
Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer DIMENSIONS WIRING 6.00 152.4 1.98 50.3 5.56 141.2 R 0.1 4.00 2.5 102.0 0.35 8.8 DIFFERENTIAL PRESSURE Multi-Sense® Model 231RS 2.00 50.8 Side View Front View 2x 0.86 21.9 1.56 39.6 4.0 102 5.16 131.1 IN MM 1.22 30.9 1.24 31.5 0.95 24.1 1.24 31.5 DC/AC Power Supply + Controller Voltage + Relay 3-Wire - Voltage Output 0 to 5 VDC 0 to 10 VDC 1 to 5 VDC Remote Zero Bottom View Conduit Version IN MM 6.0 152.4 1.98 50.3 5.
DIFFERENTIAL PRESSURE Multi-Sense® Model 231RS Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer DIMENSIONS WIRING 1/2 inch Conduit Opening (CN8 & CN11) Conduit Adapter IN MM 1.40 35.48 Transducer w/Conduit 3-Pin Packard Connector (CN8 & CN11) 3.61 91.7 3/4” HEX 1.68 42.5 Transducer w/Packard Connector 1.62 41.
DIFFERENTIAL PRESSURE Multi-Sense® Model 231RS Wet-to-Wet, Differential, Multi-Configurable Pressure Transducer INSTALLATION Model 231RS High Pressure Remote Sensor Low Pressure Remote Sensor Valve A1 High Valve B1 Low Valve A = High Side Valve Valve B = Low Side Valve 1. Valves not included. Pipe Pipe Pressure Range Code Selector (IMportant: READ BEFORE ORDERING) Range Code A B C D Max. Line Pressure RS1 50 25 10 5 50 RS2 75 37.5 15 7.
DIFFERENTIAL PRESSURE Model 239 High Accuracy, Low-Differential Pressure Transducer FEATURES ±0.14% FS Accuracy Fast Warm-Up Low Thermal Effects Fast Response Time (<10ms) Withstands High Overpressure RoHS Compliant Meets CE Conformance Standards DESCRIPTION The Model 239 Series pressure transducers are designed for very low pressure applications that require high accuracy.
High Accuracy, Low-Differential Pressure Transducer ORDERING INFORMATION 2 3 9 1 Model 2391 1 F Pressure Ranges 239 Pressure Fitting Output Unidirectional Bidirectional 1F 1/8” NPT Female Termination Accuracy Options4 N 11 4 to 20 mA 02 2’ Cable 22 GA W ±0.14% FS 1 9 ±0.073% FS 0R5WD 0 to 0.5 in. W.C. R25WB ±0.25 in. W.C. 2S ±2.5 VDC 10 10’ Cable 22 GA 001WD 0 to 1 in. W.C. 0R5WB ±0.5 in. W.C. 2B 0 to 5 VDC2 25 25’ Cable 22 GA 2R5WD 0 to 2.5 in. W.C. 001WB ±1 in. W.C.
DIFFERENTIAL PRESSURE 34 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
ROOM PRESSURE MONITORS MODELS: MRMS SRIM SRCM SRMD DIFFERENTIAL PRESSURE PRODUCT SECTION 2.1 SRPM ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
Room DIFFERENTIAL PressurePRESSURE Monitors Model MRMS Multi-Room Monitoring Station FEATURES ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ DESCRIPTION The MRMS (Multi-Room Monitoring Station) is designed for installation in a central location, such as a nurses station or main control room. It is designed to be flush mounted to provide remote viewing and alarm monitoring for up to 8 rooms or critical spaces equipped with Setra’s Pressure and Room Condition Monitors, such as the Model SRPM or SRCM.
Room Pressure Monitors DIFFERENTIAL PRESSURE Model MRMS Multi-Room Monitoring Station DIMENSIONS mm in. SSP-MRMS Rev. A 2/21/2012 WIRING ORDERING INFORMATION M R M S Model Face Plate Logo MRMS = MRMS Std. S Setra Opt. B Blank/No Logo Ordering Example: MRMSS = Model MRMS with Setra logo on Face Plate. "BACnet"isaregisteredtrademarkofASHRAE" ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Room DIFFERENTIAL PressurePRESSURE Monitors Model SRPM Room Pressure Monitor FEATURES ■ Touch Screen Display ■ BACnet®Option ■ Password Enabled ■ Local Audible Alarm ■ Visual Red and Green Room Status Displays ■ SPST Alarm Relay ■ Door Status Monitor ■ Variable Alarm Delay NOTE: Setra quality standards are based on ANSI-Z540-1. The calibration of this product is NIST traceable. ■ Positive and Negative Pressure Monitoring U.S. Patent nos.
Room Pressure Monitor SPECIFICATIONS Performance Data Standard Optional Environmental Data Electrical Data (Voltage) Operating Temperature3 °F (°C) 32 to +120 (0 to +50) Circuit Storage Temperature °F (°C) -20 to +160 (-30 to +170) Output 5 to 95% RH (Non-Condensing) Excitation Accuracy RSS (at constant temp) ±0.25% ±0.5% Non-Linearity, BFSL ±0.24% ±0.49% Operating Humidity Hysteresis ±0.05% ±0.05% Physical Description Non-Repeatability ±0.05% ±0.05% Zero Setting Tolerance ±0.
Room DIFFERENTIAL Pressure PRESSURE Monitors Model SRPM Room Pressure Monitor SRPM MENU TREE Touch-Sense Screen The touch sensitive screen allows easy navigation of the menu tree to configure the SRPM for negative or positive room pressure control. If pressure is Normal, the screen is Green 40 If pressure is Normal, and Door is open, the screen is Yellow If pressure falls outside of preset limits (Alarmed State), the screen is Red Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
Room Pressure Monitor ORDERING INFORMATION S R P M Model Range Code Excitation/Output SRPM = SRPM See Table 1 Below A1 24 VAC/4-20 mA or 0-5 and 0-10 VDC E ±0.5% FS V1 120/240 VAC/4-20 mA or 0-5 and 0-10 VDC V ±0.25% FS A2 24 VAC w/ BACnet® V2 120/240VAC BACnet® Table 1. Range Specification RANGE CODE INCHES W.C. 005WB 2R5WB 001WB 0R5WB R25WB 0R1WB R05WB ±5 ±2.5 ±1.0 ±0.5 ±0.25 ±0.1 ±0.
Room PressurePRESSURE Monitors DIFFERENTIAL Model SRCM Room Pressure and Condition Monitor FEATURES ■ True Pressure Measurement • High accuracy Setra low differential technology • Dead ended solution —no contamination or clogging • Standard on-board sensor and optional remote sensor ■ Display 4 Ambient Parameters • Pressure, Temperature, Humidity, User-Defined (ex.
Room Pressure and Condition Monitor Attractive, Intuitive, and Configurable Design Simultaneously Monitor and Display 4 Parameters Per Room (up to 2 rooms) Selectable Room Condition Banners 1/3 Screen Banners Display 4 Ambient Parameters (slide bar off ) Pressure, Temperature, Humidity, and User-Defined Parameter Room I.D.
Room DIFFERENTIAL PressurePRESSURE Monitors Model SRCM Room Pressure and Condition Monitor SPECIFICATIONS Performance Data Code F Environmental Data Electrical Data (Voltage) Operating Temperature °F (°C) 32 to + 120 (0 to +50) Circuit 3 Code H 3-Wire (Exc, Out, Com), Accuracy RSS1 (at constant temp) ±0.25% FS ±0.5% Storage Temperature °F (°C) -20 to +160 (-30 to +70) Output4 0 to 5 VDC, 0 to 10 VDC Non-Linearity (BFSL based) ±0.24% ±0.
Room Pressure and Condition Monitor ORDERING INFORMATION S R C M Model Range Code Excitation/Output Accuracy Pressure Snubber* SRCM = SRCM See Table 1 Below A1 24 VAC/4-20 mA or 0-5 and 0-10 VDC H ±0.5% FS N 0 A2 24 VAC w/ BACnet® F ±0.25% FS 1 1 2 2 Room Pressure Monitors DIFFERENTIAL PRESSURE Model SRCM Table 1. Range Specification RANGE CODE INCHES W.C. RANGE CODE PASCALS R05WB 0R1WB R25WB 0R5WB 001WB 2R5WB 005WB ±0.05 ±0.10 ±0.25 ±0.50 ±1.00 ±2.50 ±5.
Room PressurePRESSURE Monitors DIFFERENTIAL Model SRIM Room Isolation Monitor FEATURES Large 2-Line LCD Display Wall Mount Membrane Keypad for Ease of Configuration & Menu Navigation Upper/Lower Alarm Limit Setting Wall & Duct Mount Versions Mounting Thickness <50mm Easy Calibration without Removing the Wiring or Plumbing True 2-Wire 4 to 20 mA, 0 to 5 VDC & 0 to 10 VDC, Field Selectable Duct Mount Unidirectional or Bidirectional Ranges Accuracy Options: ±1.0%, ±0.5%, ±0.
DIFFERENTIAL Room PressurePRESSURE Monitors Model SRIM Room Isolation Monitor ORDERING INFORMATION S R I M Model N Pressure Ranges Model SRIM Type in. W.C. Output Differential 11 4 to 20 mA Mounting/Logo Accuracy Pressure Snubber Option WL Wall Mount, w/ Logo C ±1.0% FS N Quantity 0 N R05W* 0 to 0.05 ±0.05 D Unidirectional 2B 0 to 5 VDC DL Duct Mount, w/ Logo F ±0.25% FS w/ cal. cert 1 Quantity 1 0R1W 0 to 0.1 ±0.
Room Pressure PRESSURE Monitors DIFFERENTIAL Model SRMD Setra Room Monitoring Display FEATURES Dual Display Highly Visible 1” LCD Display Single LCD Display or Dual LCD Display Model Flush Mount Design Wipe Down Capable Available in Red, Green or Blue LCD Backlight Mount in Standard 4-11/16” sq.
DIFFERENTIAL PRESSURE Room Pressure Monitors Model SRMD Setra Room Monitoring Display DIMENSIONS .25 256.35 6.35 5.81 5.81 147.62 147.62 .99 0.99 25.15 25.15 3.68 3.68 93.52 93.52 5.81 5.81 147.62 147.62 25.25 6.35 6.35 .99 0.99 25.15 25.15 5.81 5.81 147.62 147.62 1.68 1.68 42.72 42.72 5.81 5.81 147.62 147.62 2.06 2.06 52.45 52.45 1.07 1.07 27.05 27.08 Dual Display in.
DIFFERENTIAL PRESSURE 50 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
POWER MONITORING DIFFERENTIAL PRESSURE PRODUCT SECTION 3.1 MODELS: Power Patrol Patrol Flex CT Split Core Standard CT Split Core Performance CT Power Squad 24 ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
Power Monitoring DIFFERENTIAL PRESSURE Power Patrol Advanced Power Meter The Setra Power Patrol is every electrical contractor’s dream. The networked 3-phase power meter works with Rogowski Coils and has a small enough form factor to be mounted inside or outside of the panel using either mounting tabs or the DINrail clip making it the easiest installation in the industry. Rogowski and CT Compatible The Power Patrol works with either Rogowski Coil “flex” CTs or conventional split-core CTs.
Power Patrol Advanced Power Meter SPECIFICATIONS Technical Communications Service Type Single Phase, Three Phase-Four Wire (WYE), Three Phase-Three Wire (Delta) Direct Power From L1 Phase to L2 Phase. 80-600VAC CAT III 50/60Hz, 70 mA Max. Nonuser replaceable .5 Amp internal fuse protection Max Distance 1200 meters with data rate of 100K bits.
Power DIFFERENTIAL Monitoring PRESSURE Power Squad 24 Multi-Circuit Power Meter POWER SQUAD 24 The Power Squad 24 is a versatile, multi-channel (CT) instrument. The modular design allows it to be configured for monitoring multiple electrical circuits (sharing a common voltage source) or for current-only monitoring of branch circuits.
Multi-Circuit Power Meter SPECIFICATIONS Technical Communications Service Type Single Phase, Three Phase-Four Wire (WYE), Three Phase-Three Wire (Delta) Direct BACnet TP, BACnet MS/TP, Modbus TCP, Modbus RTU Power From L1 Phase to L2 Phase. 80-600VAC CAT III 50/60Hz, 70 mA Max. Nonuser replaceable .5 Amp internal fuse protection Max Distance 1200 meters with data rate of 100K bits.
power monitoring DIFFERENTIAL PRESSURE Patrol Flex Rogowski Coil DESCRIPTION The Setra Patrol Flex is an AC current probe utilizing the Rogowski principle. The flexible and lightweight measuring head allows quick and easy installation in hard to reach areas, without batteries or an external power source. Setra Patrol Flex is available as a 3-pack, perfect for use with 3-phase power applications. The Flex clamp fits around bus bars and large or hard-toreach conductors.
Current Transformers DESCRIPTION Split Core Standard CTs provide linear output voltage that is directly proportional to the input current. These current transformers are safely and easily installed over existing electrical power lines without disconnecting the lines or interrupting service.
power monitoring DIFFERENTIAL PRESSURE Split Core Performance CT Current Transformers DESCRIPTION The hinged Split Core Performance CTs are small, low cost current transformers with high accuracy over a wide dynamic range with excellent phase shift. These CTs are ideal where space is limited such as when metering multiple loads within a panel board. Use for current measurement, energy metering, load surveys, demand metering, energy research, and submetering.
DIFFERENTIAL PRESSURE PRODUCT SECTION 4.1 CURRENT SENSORS MODELS: CSS Series CTC Series CCM Mini CSC Series Sure Set ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
DIFFERENTIAL PRESSURE Current Sensors Model CSS Series Solid Core Current Switches FEATURES ■ Solid Core Design ■ Adjustable Switch Setpoints ■ Switch LED Indication ■ Relay LED Indication ■ Over/Under Current Sensing ■ Snap-On Power Relay DESCRIPTION The CSS models are ideal for new installations and provide the greatest savings opportunity.
Solid Core Current Switches SPECIFICATIONS Model CSSGFN025NN CSSGA2100NN CSSGA2100R1 w/ snap-on relay Amperage Range 0.25 to 200 A 1.00 to 135 A 1.00 to 135 A Continuous Operating Current 200 A, 600 VAC 125 A, 600 VAC 135 A, 600 VAC Switch Setpoint Fixed Adjustable Adjustable Output Relay No No SPST,NO. 10 A @ 260 VAC, 5 A @ 30 VDC Actuation Coil No No 24VAC/DC Switch LED Indication No Yes Yes Relay LED Indication No No Yes Trip Setpoint o.25 A 1.00 to 135 A 1.
DIFFERENTIAL PRESSURE Current Sensors Model CSC Series Split Core Current Switches FEATURES ■ Clamped/Split Core Design ■ Adjustable Switch Setpoints ■ Switch LED Indication ■ Relay LED Indication ■ Over/Under Current Sensing ■ Snap-On Power Relay DESCRIPTION A significant increase or decrease in operating current may result in motor belt loss, slippage, or mechanical failure, which could jeopardize the user’s process.
Split Core Current Switches SPECIFICATIONS MODEL CSCGFN015NN CSCGFN150NN CSCGA2125NN CSCGFN150R1 w/snap-on relay CSCGA2125R1 w/snap-on relay Amperage Range 0.15 to 200 A/ 1.5 to 200 A 1.25 to 135 A 1.5 to 200 A 1.25 to 135 A Continuous Operating Current 200 A, 600 V AC/ 200 A, 600 V AC 135 A, 600 V AC 200 A, 600 V AC 135 A, 600 V AC Switch Setpoint Fixed Adjustable Fixed Adjustable Output Relay No No SPST. NO 10 A @ 260 V AC, 5 A @ 30 VDC SPST. NO.
DIFFERENTIAL Current Sensors PRESSURE Model CTC Series Split Core Current Transducers FEATURES ■ Clamped/Split Core Design ■ Slide Switch, Selectable Amperage Ranges ■ Snap-On Power Relay ■ Relay LED Indication on CTC when Used with Optional CCR-24 or CCR-12 Command Relay ■ Low Cost Solution ■ 24 VDC Loop Power or Self-Powered ■ Simple Installation DESCRIPTION CTC Series Split Core Current Transducers combine accurate magnetic current sensing with signal conditioning elec- tronics.
Split Core Current Transducers SPECIFICATIONS MODEL CTCG420NN CTCGV05NN CTCGV10NN Multi-Range 30/60/120 A 30/60/120 A 20/100/150 A Continuous Operating Current 120 A Max. 120 A Max. 150 A Max.
DIFFERENTIAL PRESSURE Current Sensors Sure-Set Split Core Current Switch FEATURES ■ Sure-Set Scaled in Motor Hp, Allowing User to Preset Unit Prior to Installation ■ 9 Motor HP Settings Per Model ■ De-rating Built-In from Full Load Amps to Detect Belt Loss or other Mechanical Load ■ Snap-on Power Relay Option BENEFITS DESCRIPTION Eliminates Exposure to Arc Flash - The Sure-Set Model SSC Split Core Current Switch provides a unique approach to calibration and installing current sensors that eliminate
Split Core Current Switch SPECIFICATIONS MODEL Motor Hp Range SSC-2H SSC-4H 5, 7.5, 10, 15, 20, 25, 30, 40, 50 15, 20, 25, 30, 40, 50, 60, 75, 100 Continuous Operating Current 135A, 600V AC Switch Setpoint Adjustable, 9 position selector switch Output Relay Contacts (option) Optional.
DIFFERENTIAL PRESSURE Current Sensors CCM MINI Current Clamped Mini FEATURES ■ Clamped/Split Core Design ■ Under Current Sensing ■ Integral Mounting Flange with DIN-Rail Capability BENEFITS ■ Low Cost Solution ■ Self-Powered ■ Simple Installation ■ Accurate Fixed Setpoint, DESCRIPTION The CCM Mini is a cost effective solution to monitoring light to medium current loads.
Current Clamped Mini SPECIFICATIONS MODEL CCM015NN Amperage Range 0.15 to 60 A Continuous Operating Current 60A, 300V AC Current Set Point Fixed Switch LED Indication No Relay LED Indication Yes Trip Point Set Value 0.15A Current Switching Mode DIFFERENTIAL PRESSURE Current Sensors CCM MINI Under Current Sensing Dimensions 1.57 H X 1.66 L X 1.52 W in. (39.9) x 42.2 L x 38.6 W mm) Aperture Size 0.3 in. (7.
DIFFERENTIAL PRESSURE 70 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
DIFFERENTIAL PRESSURE PRODUCT SECTION 5.1 GAUGE PRESSURE MODELS: 206 3100 209 3200 256 ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
GAUGE PRESSURE DIFFERENTIAL PRESSURE Model 206 Pressure Transducers FEATURES ■ Solid Stability for Confident Installations ■ Exceptional EMI/RFI Performance Prevents False System Shutdown ■ NEMA-4/IP-65 Certified (206) for Use in Harsh Environments ■ Reverse Wiring Protection ■ Rugged Design Withstands High Shock/ NOTE: Setra quality standards are based on ANSI-Z540-1. The calibration of this product is NIST traceable. U.S. Patent nos.
SPECIFICATIONS Performance Data Environmental Data Electrical Data (Voltage) ±0.13% FS Operating3 Temperature °F (°C) 32 to + 120 (0 to +50) Circuit Non-Linearity, BFSL ±0.1% FS Storage Temperature °F (°C) -20 to +160 (-30 to +70) Output 25 psig Range2 ±0.2% Operating Humidity 5 to 95% RH (non-condensing) External Load 0 to 800 ohms Acceleration 10 g Maximum5 Minimum Supply Voltage (VDC) 9 + 0.
Gauge Pressure DIFFERENTIAL PRESSURE Model 206 Pressure Transducers DIMENSIONS Cable with Conduit Version IN MM CONDUIT FITTING 1/2" NPT 1.16 29.5 7/8" HEX C206 Cable Version φ 2.00 φ 50.8 ZERO 2.00 50.8 φ 0.44 φ 11.1 SPAN 2.13 54.0 0.78 19.8 19/32 ACROSS FLATS 7/16-20 SAE PRESSURE FITTING SPAN ZERO 206 0.36 9.1 φ 2.00 φ 50.8 2.00 50.8 Hirschmann Connector 0.75 0.50 19.1 12.7 HIRSCHMANN CONNECTOR ZERO 0.82 20.8 SPAN 3/4" HEX 1/4"-18 NPT PRESSURE PORT 206 2.00 50.8 1.00 25.4 0.
ORDERING INFORMATION 2 0 6 1 Model Range Code 2061 = 206 See Table 1 Below Pressure Type Fitting Output Termination Accuracy Options2 8 NN None G Gauge 1M 1/4” NPT Male 11 4 to 20 mA XX Cable Length1 C Compound 2M 1/8” NPT Male 22 0.1 to 5.1 VDC H1 Hirschmann A Cleaning for Oxygen Service A Absolute 1F 1/8” NPT Female 27 1 to 5 VDC A1 1/2” Conduit B Mating Bayonet Connector 2F 1/4” NPT Female 28 1 to 6 VDC T1 Terminal Block C Cal Cert J7 7/16” SAE 2T 0.
DIFFERENTIAL PRESSURE Gauge Pressure Model 209 Pressure Transducers FEATURES ■ High Over Pressure Option Available on Selected Ranges ■ Rugged Design Withstands Harsh Environments ■ Operates Over a Wide Temperature Band ■ Compatible w/ Wide Range of Gases & Liquids DESCRIPTION ■ Operates on Low Cost Unregulated DC Power NOTE: Setra quality standards are based on ANSI-Z540-1. The calibration of this product is NIST traceable. U.S. Patent nos.
SPECIFICATIONS Performance Data Environmental Data Accuracy RSS1 (at constant temp) ±0.25% FS Operating Temperature °F (°C) -40 to + 185 (-40 to +85) Circuit Non-Linearity, BFSL ±0.22% FS Storage Temperature °F (°C) -40 to + 185 (-40 to +85) Excitation 9 to 30 VDC Hysteresis 0.10% FS 3 Shock 200g operating Output 0.5 to 5.5 VDC7 Non-Repeatability 0.
Pressure Transducers DIMENSIONS CONDUIT VERSION CABLE ANCHOR 1.40 35.0 φ φ φ φ 1.62 41.6 2.40 60.1 φ 1.62 41.0 φ 3.61 92.0 0.50 12.7 φ 1.58 40.0 φ φ DIFFERENTIAL PRESSURE Gauge Pressure Model 209 φ 2.00 50.8 3/4" HEX 3/4" HEX 1.68 43.0 1/8" NPT 1/4" NPT φ OPTIONAL 3-Pin PACKARD CONNECTOR Type: P2S Series 150 1.37 35.1 φ in. mm 0.33 8.4 φ OPTIONAL HIRSCHMANN CONNECTOR Type: G4A1M #931807-106 0.29 1.38 7.4 35.1 φ φ 1.62 41.2 3.0 76 φ 2.24 56.9 3/4" HEX 3/4" HEX 1/4" NPT 1.
Pressure Transducers ORDERING INFORMATION 2 0 9 1 Model Range Code 2091 = 209 See Table 1 Below Table 1. Range Specification Pressure Type Pressure Fitting Output Elec. Termination Options G Gauge 2M 1/4” NPT Male 11 4-20 mA XX C Compound J7 7/16” SAE Male 24 0.5 to 5.5 VDC P1 Packard (3-Pin) S Sealed 1M 1/8” NPT Male 28 1 to 6 VDC P3 Packard (4-Pin)3 V Vacuum L4 1/4 Female SAE 45 0.5 to 4.
DIFFERENTIAL Gauge Pressure PRESSURE Model 256 Pressure Transducers BENEFITS ■ Low Cost ■ High Accuracy ■ NEMA-4/IP-65 ■ Wide Operating Temperature Range ■ Compatible with a Wide Range of Gases or Liquids ■ Corrosive Resistant All Stainless Steel Wetted Parts NOTE: Setra quality standards are based on ANSI-Z540-1. The calibration of this product is NIST traceable. DESCRIPTION U.S. Patent nos.
Gauge Pressure DIFFERENTIAL PRESSURE Model 256 Pressure Transducers DIMENSIONS Wiring 4.02 102.1 2.50 61 Removable Terminal Block Connector Removable Zero/Span 0.38 9 Screw Terminal Designations + EXC EXC GND OUT + OUT Access Plugs 2.00 50.8 254 64.5 1.46 37.1 + EXC EXC GND in. mm Voltage Connections 3.56 90.4 1/4” NPT Current Connections ORDERING INFORMATION 2 5 6 1 Range Code Pressure Type 2561 = 256 See Table 1 Below G Gauge Table 1.
DIFFERENTIAL PRESSURE Gauge Pressure Model 3100/3200 Standard & Heavy Duty OEM Pressure Transducers FEATURES ■ Low Cost for High Volume OEM Installations ■ Thin Film Tech. Assures Long-Term Stability ■ Wide Choice of Pressure Ranges from 50 PSI M12 x 1P up to 32,000 PSI Din 9.4 mm ■ Long-Term Stability Better Than ±0.1% FS/Yr AMP Superseal 1.5 DESCRIPTION Packard Metri Pack ■ 0.
Standard & Heavy Duty OEM Pressure Transducers SPECIFICATIONS Performance Data Thermal Effects2 Accuracy RSS1 Compensated Range °F (°C) Electrical Data (Voltage)6 -40 to +221 (-40 to +105) Model 3100 ±0.25% FS Zero/Span Shift %FS/100°F (%FS/100°C) Model 3200 ±0.25% FS Model 3100 0.83 (1.5) Model 3200 0.94 (2.0) Environmental Data Operating Temperature °F (°C) -40 to +221 (-40 to +105) Storage Temperature °F (°C) -40 to +221 (-40 to +105) Approvals Zero Tolerance Model 3100 Model 3200 ±0.
DIFFERENTIAL PRESSURE Gauge Pressure Model 3100/3200 Standard & Heavy Duty OEM Pressure Transducers ELECTRICAL FITTINGS Din 9.4 mm M12 x 1P Amp Supeseal 1.5 1 2 Deutsch DT4-4P Packard Metri Pack 3-Pin Deutsch A 3 1.02 (25.86) 3 1 2 4 3 3 4 2 2 1 4 0.76 (19) Code B B 1.50 (38) 1.53 (39) 1.46 (37) 0.71 0.72(18) (18) 1.63 (41.38) 0.75 (19) 0.75 (19) C B 1.02 (26) 0.38 (9.7) (10) 0.38 0.28 (7) A C 1 2 0.75 (19) Code E Code 6 0.
DIFFERENTIAL PRESSURE Gauge Pressure Model 3100/3200 Standard & Heavy Duty OEM Pressure Transducers ORDERING INFORMATION Output See Table 1 Range Code B 4-20 mA C Compound C 1-6 VDC G Gauge H 1-5 VDC S N 0.5-4.5 VDC Sealed Gauge2 R 0-5 VDC S 0-10 VDC T 0.5-5.5 V Ratiometric Table 1. Model Specification DESCRIPTION 3100 3200 Std. 3100 Std. 3200 3101 Temp. Output Range: -40°C to +105°C 31021 Temp. Output Range: -0°C to +100°C Temp. Output Range: -0°C to +80°C 32011 Temp.
DIFFERENTIAL PRESSURE 86 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
INDOOR AIR QUALITY DIFFERENTIAL PRESSURE PRODUCT SECTION 6.1 MODEL SRH: Wall Mount Duct Mount Outside Air ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
Humidity DIFFERENTIAL Transmitters PRESSURE Model SRH Relative Humidity Sensor FEATURES ■ Available in Wall, Duct Mount or Outdoor Air ■ Key Component of Comprehensive HVAC/R System ■ Active Temperature with Jumper Selectable ■ Tspan Ranges of 40oC, 50oC, and 60oC ■ Excellent Reliability through ASIC Technology ■ Robust Capacitive Sensor Design ■ Low Cost of Ownership ■ Three Accuracy Options: 2%, 3%, & 5% ■ Replaceable Sensor Tip ■ Quick Mount, 2 Screw Install with Plug-In Terminal Wiring ■ 5 Year Warran
Relative Humidity Sensor SPECIFICATIONS RH Performance Data Temperature Sensing Options (Passive) Physical Description T1: Thermistor NTC 10K Ω 77°F/25°C (Direct Connect) Type II Enclosure Materials Sensing Element Capacitive Polymer Humidity Operating Range 0 to 99% RH (non-condensing) T2: RTD Output 1000 Ω 32°F/0°C (Direct Connect) Wall Mount VA 94-V0 Accuracy @ 68°F (20°C) 2%, 3%, 5%1 T6: Thermistor NTC 10K Ω 77oF/25oC Type III Duct & Outside Air Polycarbonate 94-V0 Non-Repeatability
DIFFERENTIAL PRESSURE Model SRH Relative Humidity Sensor DIMENSIONS Outside Air Wall Mount [71.50] [71.5] 102.6 4.04 154.9 6.10 2.81 2.81 68 2.68 60.4 2.38 114.4 4.50 [119] [119] 4.69 4.69 [102.7] 102.7 4.05 4.05 [318] 318 12.52 12.52 SECURITY LOCK Security Lock UNIVERSAL UNIVERSAL WALL/PIPE WALL/PIPE MOUNTING FLANGE MOUNTING FLANGE [24.7] [24.7] .98 0.98 POROUS POROUS WEATHER SHIELD WEATHER SHIELD SECURITY LATCH Security Latch REVISION RECORD REV.
Relative Humidity Sensor ORDERING INFORMATION S R H 1 N Model Accuracy Configuration Outputs SRH1 = SRH 2P 2% W Wall 11 4 - 20 mA 3P 3% D Duct 2C 5P 5% O Outside Air 0 -5 or 0-10 VDC (user-selectable) 1 Temperature Outputs Display3 Options T0 N C None (RH only) T1 10K Ω Type II Thermistor ( Passive) T2 1000 Ω RTD (Passive) T3 -58 to 140°F (-50 to 60°C [Active])2,3 T5 +14 to 140°F (-10 to 60°C [Active])2,3 T6 10K Ω Type III Thermistor [Passive] None NIST Certificate
DIFFERENTIAL PRESSURE 92 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Very Low Pressure Calibrators MODELS: 869 DIFFERENTIAL PRESSURE PRODUCT SECTION 7.1 869XP ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
Calibrators DIFFERENTIAL PRESSURE Model 869/869XP Ultra-Low Pressure Generating and Documenting Calibrator FEATURES ■ Easy Step-by-Step User Interface Process ■ Designed with Built-in Leak Test Function ■ Provides Accuracy and Stability Plots ■ Handles Multiple Engineering Units Model 869 DESCRIPTION ■ Both Pressure Generation and Monitoring Model 869XP The Model 869 and 869XP are designed for use in critical environments that require portability, high accuracy, and periodic low pressure sensor calib
Ultra-Low Pressure Generating and Documenting Calibrator SPECIFICATIONS Measurement General Specifications Pressure Transducer Accuracy ±0.04% FS Pressure Units (Selectable) n.W.C., PA, kPa, mbar, cm W.C. Pressure Fittings Barbed, Plug-in O-ring Quick Connects Precision 0.0002” W.C. Warmup 1 Hour Electrical Banana Plug Jacks Reading Rate 20 Readings/Seconds, Typical Voltage Meter ±0.005% FSO at ±10.5 VDC Gravity/Orientation Negligible Current Meter ±0.
DIFFERENTIAL PRESSURE 96 Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Accessories DIFFERENTIAL PRESSURE PRODUCT SECTION 8.1 Power Supplies Room Pressure Status Pressure Tips and Tubing 299 Dri-Sense PT ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.
DIFFERENTIAL PRESSURE ACCESSORIES Models 867 • 86730V • 868 • 874 • 890 24 VDC Power Supplies FEATURES Model 867 & Model 867 30V ■ Small Size ■ Light Weight ■ Integral Barrier Strip Terminal for Input and Output Wiring ■ Convenient Mounting Holes ■ Withstands Momentary Short Circuit without Models 867/867 30 V and Model 864 The Models 867 and 874 are low cost power supplies that have the advantage of being able to withstand a momentary short circuit without failure.
24 VDC Power Supplies Specifications Models 867/867 30V/ 874 Output Model 867 24 VDC unregulated filtered <29 VDC with no load, >21 VDC at 100 mA No more than 0,7 pk - pk ripple Model 867 30V 30 VDC unregulated filtered Model 874 24 VDC @ 80 mA <29 VDC with no Load Input Model 867 & 867 30V 120 VAC, 60Hx\z Model 874 220-240 VAC, 50/60 Hz DIFFERENTIAL PRESSURE ACCESSORIES Models 867 • 86730V • 868 • 874 • 890 Specifications Model 868 Input Voltages 105 to 125 VAC Input Frequency 50 to 440 Hz
DIFFERENTIAL PRESSURE ACCESSORIES Room Pressure Status Applications ■ Hospital Patient Isolation Wards ■ Pharmaceutical ■ Semiconductor Fabs ■ Cleanrooms ■ Research Laboratories ■ Animal Resource Facilities Model SRAN - Remote Annunciator Setra’s Remote Annunciator (SRAN) allows remote indication of room pressure status at monitoring/nurses station. A Green LED indicates Normal room condition, a Red LED and Audible Alarm signal a breach in room pressure status.
The Stainless Steel Static Pressure Tips are used to measure static pressure in ducts or rooms. They are to be connected to differential pressure switches and transmitters. Two static pressure sensors are used in applications where differential pressure is required across a filter or coil. These sensors include a mounting flange with integral rubber gasket and two screws for simplifying mounting on a duct.
DIFFERENTIAL ACCESSORIES PRESSURE Model 299 Dri-Sense FEATURES ■ Visible Desiccant Status ■ Easily Replaceable ■ Replaceable Terminal Interface Circuit Board ■ Surge Suppression ■ NEMA 4X Industrial Housing Description The NEMA 4X rated Model 299 Dri-Sense pressure transducer enclosure is designed for field termination of pressure transducers.
Order Using Setra’s Configurable Part Number Our products feature configurable part numbers. Configurable part numbers are designed to simplify and expedite the ordering process as well as provide you with a convenient reference number for inventory control. Individual part numbers identify the product and its unique specifications. The following is an example of how to order using Setra’s configurable part numbers: Example: Order a Model 264 (2641), with a range of 0.25 in.
DIFFERENTIAL PRESSURE 104 HVAC 2013 ONLINE CATALOG Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.