Product specifications

Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
DIFFERENTIAL PRESSURE
46
Model SRIM
Room Isolation Monitor
FEATURES
Large 2-Line LCD Display
Membrane Keypad for Ease of Conguration &
Menu Navigation
Upper/Lower Alarm Limit Setting
Wall & Duct Mount Versions
Mounting Thickness <50mm
Easy Calibration without Removing the Wiring or
Plumbing
True 2-Wire 4 to 20 mA, 0 to 5 VDC & 0 to 10 VDC,
Field Selectable
Unidirectional or Bidirectional Ranges
Accuracy Options: ±1.0%, ±0.5%, ±0.25% FS
Ranges as Low as ±0.05 in. W.C.
CE Compliant
APPLICATIONS
Pharmaceutical Manufacturing
Hospitals
Laboratories
Vivariums
DESCRIPTION
In a pharmaceutical, hospital, laboratory, or animal research
facility, the integrity of the ventilation control system is at the

be maintained at a positive pressure to prevent product con-
tamination or a negative pressure to prevent contaminants
from escaping into adjacent areas, the proper pressurization of
the room is essential.
To assure proper pressurization is maintained in these critical
environments, a room pressure monitor is employed to measure
and alert sta and personnel of any change in pressure no mat-
ter how small. A fail-safe solution to monitoring these very low
pressure changes is Setras Model SRIM room pressure monitor,
which utilizes highly accurate capacitance sensing technology
to measure and display true low pressure dierential.
SPECIFICATIONS
Performance Data Physical Description Electrical Data
Code F Code H Code C/G Electrical Connection Screw Terminal Circuit 3-Wire (Exc, Out, Com)
Accuracy RSS*
±0.25% FS ±0.5% FS ±1.0% FS
Dimensions See reverse side Output 0 to 5 VDC, 0 to 10 VDC, 4 to 20 mA (Loop Powered)
Non-Linearity (BFSL)
±0.22% FS ±0.49% FS ±0.98% FS
Weight 8.9oz (Duct), 9.8oz (Wall) Excitation 18 to 32 VDC
Hysteresis
±0.1% FS ±0.1% FS ±0.1% FS
Display Custom 2-Line Character LCD Current Consumption 5 mA (voltage output mode)
Non-Repeatability
±0.05% FS ±0.05% FS ±0.05% FS
Pressure Fittings Barbed Fittings for 1/4” Tubing
*Calibrated into a 50K ohm load, operable into a 10K ohm load or greater
Zero Setting Tol.
±0.5% FS ±0.5% FS ±1.0% FS
Case Fire Retardant Plastic UL94V-0
Pressure Media
Span Setting Tol.
±0.5% FS ±0.5% FS ±1.0% FS
Air or non-conductive, non-explosive gases
Thermal Eects
Environmental Data Certications
Compensated Range
±0.06% FS/C
Temperature CE
EN61326-1 & EN61326-2-3 BASIC Immunity & Class B Emission
Overpressure
Up to 10 PSI
Operating 21.2 to +140°F (-6 to +60ºC) RoHS
*RSS of Non-Linearity, Non-Repeatability & Hysteresis at constant temp.
**Units calibrated at nominal 21ºC. Max thermal error computed from this datum.
Storage -4 to +185°F (-20 to +85ºC)
Specications Subject to Change without Notice
Operating Humidity 5 to 95% RH (non-condensing)

Duct Mount
ROOM PRESSURE MONITORS