Recommended isolation distance between portable or 
mobile RF communication equipment and this product 
This  STEMO500  stethoscope  is  expected  to  be  used  in  an  electromagnetic 
environment where RF radiation disturbances are controlled. According to the 
maximum output power of the communication equipment, the purchaser or 
user  can  prevent  electromagnetic  interference  by  maintaining  a  minimum 
distance  between  the  portable  or  mobile  RF  communication  equipment 
(transmitter) and the product. 
Maximum  rated 
output  power  of 
transmitter 
W 
Isolation  distance  corresponding  to  different 
transmitter frequencies / m 
150  kHz~  80 
MHz 
d = 1.2
√
𝑃 
80  MHz~ 
800MHz 
d = 1.2
√
𝑃 
800  MHz~ 
2.5GHz 
d =2.3
√
𝑃 
0.01 
0.12 
0.12 
0.23 
0.1 
0.38 
0.38 
0.73 
1 
1.2 
1.2 
2.3 
10 
3.8 
3.8 
7.3 
100 
12 
12 
23 
For the maximum rated output power of the transmitter not listed in the above 
table,  the  recommended  isolation  distance  d,  in  meters  (m),  can  be 
determined  by  the  formula  in  the  corresponding  transmitter  frequency 
column,  where  P  is  the  maximum  rated  output  power  of  the  transmitter 
provided by manufacturer, in watts (W). 
Note 1: At central frequencies of 80 MHz and 800 MHz, the formula for higher 
frequency band shall be used. 
Note 2: These guidelines may not be suitable for all situations. Electromagnetic 
propagation is affected by absorption and reflection from buildings, objects, 
and human bodies. 