User`s manual

Setting Measurement Conditions
3-28
3
1. Specify a measurement area smaller than the mask area and specify the binary conditions.
2. Return to the MAIN OPS MENU.
4) Measure the binary area
Measuring area
Captured image
White dots (six)
Binary mask image (registered in step 1-2)
Image after executing
a measurement
- The white dots will be converted to
black dots by applying the binary
image mask.
RECTANGLE
CIRCLE
ELLIPSE
FIXED
THRES-ADJ(VAR-DEFF)
THRES-ADJ(VAR-RATE)
NO
EXPD.CONTR.
CONTR.EXPD
NO
YES
REG COND SCREEN COND SAVE DETAIL
F C1 BRT
1
MEAS SHAPE RECTANGLE
2
MEAS AREA SET (224,208)~(287,271)
3
THRESHOLD SET [U.LM255 L.LM100(0~255) ]
4
INVERT B/W NO
5
BINARY PROCESS FIXED
6
BINARY NOISE FILT NO
SET=SELECT A MENU ESC=BACK SEL=CHNG IMGTRG=FUNC
MASK REGIST.
BINARY PROCESS
BINRY NOISE FILT
SET=CHNG ESC=BACK SEL=CHNG IMG TRG=FUNC
1
U.LM 255(0~255)
2
L.LM 100(0~255)
THRESHOLD VAL AUTO-REG RESET
F C1 BRT
RECTANGLE RESET RETURN
F C1 BRT
1
UP.L (224,208)
2
LO.R (287,271)
3
MOVE
3) Setting the binary area conditions