User`s manual

7-4
Area Measurement by Binary Conversion
7
How to set the register conditions
How to enter the REG-COND setting screen
There are three methods for getting to the REG-COND setting screen.
1) On the "MEA-CND" setting screen, move the
cursor to the " BIN AREA COND" and
press the SET key. The cursor will move into
the table. Move the cursor to the "" column
and press the SET key.
Note: If you press the SET key in the "X"
column, the setting will be invalid.
2) On the "MEA-CND" setting screen, press the
TRG/BRT key to move the cursor to the upper
function menu. Select the "COND" item and
then select the "REG COND" item and press
the SET key.
3) On the menu tree, move the cursor to the
"MEAS CND" item, and press the right arrow
key. The sub menu which contains the "REG
COND" item will appear. Move the cursor to
this item and press the SET key.
Set the register conditions
REGISTER NO.
The currently selected register number is
displayed.
If you want to set the measurement conditions
for a different register number, press the SET
key to highlight the register number currently
selected. Then press the up and down arrow
keys to select the register number whose
measurement conditions you want to set.
Press the SET key to confirm your selection.
Note: When the "BINARY-IMG-MASK" or the
"POLYGON" is selected at
" WINDOW" on the "MEA-CND"
screen, the " REGISTER NO." line
will not be displayed and the other item
numbers are decrement by one since
the register number for "BINARY-IMG-
MASK" and "POLYGON" is fixed.
1
WINDOW
NUM-OF-MASK
2
REGISTER NO.
00(015)
3
REGISTER EXIST
NO
REGISTER NO.
4
BIN AREA COND
15141312111009080706050403020100
× × × × × × × × × × × × × × ×
F C1 BRTMEA-CND
SCREEN COND SAVE
IMG PRE-PROC
REG COND
EVALUATION
NUM-CALC
OUT
1REGISTER NO.
2MEAS SHAPE
3MEAS AREA
4THRESHOLD
5INVERT B/W
F C1 BRTSELECT OBJECT TYPE COND
OBJECT TYPE COND
  
TYPE00
TYPE01
TYPE RUN COND
IMAGE-ADJ
MEA-CND(CAMERA1)
MEAS0
POSI-DEVIATION
MEAS01(MEAS-BIN-AREA)
IMG PRE-PROC
MEAS CND
REG-COND
EVALUATION COND
NUM-CALC
OUT-COND
MEAS(NEW)
MEA-CND(CAMERA2)
FINAL NUM. CALC
FINAL OUTPUT COND
OBJ-TYPE I/O
OBJ-TYPE SYS.
SET=TO NEXT SUB-MENU ESC=BACK SEL=CHNG IMG TRG=POPUP
+
+
1
REGISTER NO.
00(0∼15)
2
MEAS SHAPE RECTANGLE
3
MEAS AREA SET
(224,208)(287,271)
4
THRESHOLD SET [U.LM255 L.LM100 (0~255)]
5
INVERT B/W NO
F C1 BRTREG COND
SCREEN COND SAVE DETAIL
SET=VALUE INPUT ESC=BACK SEL=CHNG IMG TRG=FUNC