System information

5 Injector System
www.turnerbiosystems.com 34
6. Manually open the instrument door and hold it down. Insert the
injector tip into the injector tip holder on the front of the
Luminescence Module. Note: Injector tip #1 is on the left and
tip #2 is on the right.
NOTE: During installation, use caution to prevent damage to the injector tip. Any nicks or
deformations can adversely affect the quality of the injection and mixing.
Outlet
Inj 1 Inj 2
7. Attach the outlet tubing to the inside port of the respective
injector pump. If there are two injectors, make sure the tubing
to injector #1 goes to the inside port of the left pump and the
tubing for injector #2 goes to the inside port of the right pump.
8. Insert the thicker end of the vertical support rod into the small
opening on the injector base. Press down firmly until the rod
clicks in place.
9. Slide the inlet tube holder onto the vertical support rod.
10. Following the diagram provided on the injector unit, attach the
inlet tubing to the outside port of the injector pump. If there are
two injectors, make sure the tubing goes to the correct outer
port.
Inlet Inlet
Inj 1 Inj 2