5 Injector System
www.turnerbiosystems.com 38
2. Place the container of 70% ethanol on the bottle holder. The stainless steel inlet tube should
be inserted to reach the bottom of the bottle.
3. Place an empty Waste Collection Tray on the Microplate Sample Tray.
4. From the INSTRUMENT CONTROL screen, go to the READ screen. Touch the SETUP
button to run the PRIME protocol.
5. Allow the solution to sit in the fluid path for 30 minutes before flushing the injector(s).
6. Run the CUSTOM FLUSH protocol nine times with deionized water.
7. Empty and clean the Waste Collection Tray and place it on the Microplate Sample Tray.
8. Place a container of deionized water on the bottle holder with the stainless steel inlet tube
inserted to reach the bottom of the bottle.
9. From the INSTRUMENT CONTROL screen, go to the READ screen. Touch the SETUP
button and run the FLUSH protocol.
10. Select the option for CUSTOM FLUSH.
11. Select nine FLUSH cycles.
12. When all FLUSH cycles are complete, empty and clean the Waste Collection Tray and place
it on the Microplate Sample Tray.
13. FLUSH the injector(s) three times with air. A small volume of water will remain in the injector
after the air purge.
5.5 Cleaning the Waste Collection Tray
The Waste Collection Tray prevents liquids from splashing or spilling inside the instrument during
the PRIME and FLUSH protocols. The maximum volume capacity of the Waste Collection Tray is
approximately 50ml. Use care when removing the Waste Collection Tray from the Microplate
Sample Tray so that no liquid spills in the inside the instrument.
After each PRIME and FLUSH sequence, discard any fluid in the Waste Collection Tray and rinse
with deionized water.
5.6 Replacing Injector Tips
Injector tip replacements are available from Turner BioSystems. Only injector tips from Turner
BioSystems are fully tested and compatible with the ModulusMicroplate Reader. See Appendix
C for ordering information.
1. Manually open the instrument door and hold it down with one hand.
2. Gently remove the injector tip assembly from the injector tip holder located
on the Luminescence Module.
3. Twist the white tip valve counterclockwise to release it from the gray tube
fitting. See Figure 18-15 for a picture of the white tip valve.
4. Untwist the white tip fitting from the white valve to release the tip.
5. To remove the tip, hold the tip at the base and twist off.
6. Discard tip.
7. Insert a new tip onto the valve.
8. Twist the fitting clockwise until tight to complete the injector tip assembly. A
small gap of approximately 1mm between the two fittings is normal.
9. Twist the assembled fitting with the new tip clockwise onto the tube fitting.
10. Insert the tip into the injector tip holder on the Luminescence Module.
5.7 Inserting Injector Tip Assembly
1. Hold the injector tip assembly by the valve and fitting. Pull upward.
2. Gently and continuously push the injector tip assembly into the injector tip holder until the
injector drops down into place.
Figure 5-5: Injector Tip
Assembly