9 Troubleshooting
www.turnerbiosystems.com 52
9.2 Table of Common Problems
Symptom
Possible Cause
Resolution
Injections sputter,
drip, or are
otherwise weak
Air bubbles are blocking the
line
Make sure the stainless steel tubes are
completely inserted to the bottom of
the reagent container
Reagent residue is clogging
the tip
Soak the injector tip in warm deionized
water and wipe it clean
Injector leaks.
Injector tips are damaged or
bent
See Section 1 for instructions on
replacing the injector tip
The inlet or outlet tubing is not
properly connected to the
injector syringe
See Section 1 for instructions on
removing or replacing the tubing
assembly
Injector tips do
not sit properly in
the injector tip
holder
Reagent residue has built up
inside the injector tip holder
Clean the holder with 70% ethanol and
a damp cloth
Injectors not
injecting
Air bubbles are blocking the
line
FLUSH the injectors to remove any air
bubbles
The end of the stainless steel
tube is not in the reagent
Make sure the stainless steel tube is
completely inserted to the bottom of
the reagent container
Re-prime the injectors
Reagent residue is clogging
the tubing
Replace tubing, if needed
Always flush the injector system after
use to prevent build up of reagent
residue
Reagent residue is clogging
the valves
Contact technical support
Wetness appears
on the top of the
sample tray cover
after a run
Injector tips are bent
Clean up spill and replace injector tips
Microplate has overflowed
inside instrument
Clean the optical head
Check that the total volume in each
well is less than 250ul
Thoroughly clean the interior of the
instrument
Reagent spilled inside the
instrument
Wipe up spills immediately using 70%
ethanol and a damp cloth
Standard Light
Plate reads very
low or blank
The Standard Light Plate was
not turned on before the run
Touch the START button on the
Standard Light Plate
It will automatically turn off after 5
minutes
The selected wells do not
correlate with the Standard
Light Plate
The well locations on the Standard
Light Plate are B2, D2, and F2
The Standard Light Plate was
placed in the wrong orientation
Make sure the A1 notch on the
Standard Light Plate is in the upper
right corner