Specifications

42
OxyProbe
®
D.O. Sensor with 1.5Sanitary Flange
for Sanitary Pipe Tee and Novaseptic Side Ports
Model D 210 Series OxyProbe
®
D.O. Senso rs
These sensors are unique in their construction. The front
end of the sensor is the s ame as the 19 mm sensor, with t he
smaller diameter and no crevices or o -rings. It fits directly
into a 1.5” sanitary pipe tee, eliminating the ne ed for a hous-
ing, reducing contamination risks. Designed with an inte-
gral 1.5 sanitary flan ge, a gasket seal at the flange holds the
sensor securely in place.
It is importan t to note, th ese sensors are not int
ended for
use with the stand ard 25 mm sanitary flange port s found o n
page 18. However, they can be used with the NovAseptic
ports shown on page 43.
The interior design and replacement parts of this sensor are
the same as other 25 mm Oxy
Probes. The modular design
allows quick replacement of any part, making it e asy to
maintain and repair the sensor. OxyProbe dissolved oxygen
sensors carry a 1 -year warranty.
For peformance specifications see page 35.
Features:
316L Stainless Steel
Serialized for traceabil ity
RA32 finish with electropolish on wetted parts
Additional Features Include:
Fits in both sanitary flang e ports a nd NovAseptic ports.
Can also be us ed in-line in a sanitary pipe tee.
It is important to maintain the sensors by inspecting and calibrating
them regul arly. The membrane cartridge should be checked and
refilled prior to every use.
How to choose a sensor:
(1) Confirm that the pip e tee has the proper c learance.
(2) Decide whether a right a ngle connector is preferred
over a strai
ght connector.
(3) Choose the optimum length cable and a connector. All
D.O. sensors require a separate cabl e connector assem-
bly. See pages 74 and 75.
Model D215
70 mm length
Model D210
70 mm length
nsors Dissolved Oxygen Sensors Dissolved Oxygen Sensors Disso
o r d e r i n g i n f o r m a t i o n
Sensor Part Number
Sensor Typical Insertion
Length Length
19 mm OxyProbe® with 1.5 Sanitary Flange
Straight Connector:
D210-B070-PT-D9 70 mm 70 mm 2.8”
Right Angle Connector:
D215-B070-PT-D9 70 mm 70 mm 2.8”
All OxyProbe dissolved oxygen sensors
are available with a right angle connector.
This low profile design provides:
Protection against cable fatigue
Improved grip for sensor extraction
Reduced random sensor mobility on flat
surfaces (wont roll off the table)
Broadley-James Corporation (U.S. & Canada) Toll-Free: 800-288-2833 E-mail: sales@broadleyjames.com
C
M
Y
CM
MY
CY
CMY
K
42-novaseptic_2013.pdf 1 8/11/14 10:29 AM